X-ray scatter characterization in dedicated breast CT with bowtie filters

被引:3
|
作者
Kontson, Kimberly [1 ]
Jennings, Robert J. [1 ]
机构
[1] US FDA, Ctr Devices & Radiol Hlth, Silver Spring, MD 20993 USA
关键词
bowtie filter; breast CT; scatter-to-primary ratio; IMAGE QUALITY; RADIATION; DESIGN;
D O I
10.1117/12.2043051
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
The scatter contamination of projection images in cone-beam computed tomography (CT) degrades image quality. The use of bowtie filters in dedicated breast CT can decrease this scatter contribution. Three bowtie filter designs that compensate for one or more aspects of the beam-modifying effects due to the differences in path length in a projection have been studied. The first produces the same beam-hardening effect as breast tissue with a single-material design. The second produces the same beam quality and intensity at the detector with a two-material design and the third eliminates the beam-hardening effect by adjusting the bowtie filter thickness such that the same effective attenuation is produced at the detector. We have selected aluminum, boron carbide/beryllium oxide, and PM MA as the materials for the previously described designs, respectively. These designs have been investigated in terms of their ability to reduce the scatter contamination in projection images acquired in a dedicated breast CT geometry. The magnitude of the scatter was measured as the scatter-to-primary ratio using experimental and Monte Carlo techniques. The distribution of the scatter was also measured at different locations in the scatter image to produce scatter distribution maps for all three bowtie filter designs. The results of this study will be useful in designing scatter correction methods and understanding the benefits of bowtie filters in dedicated breast CT.
引用
收藏
页数:12
相关论文
共 50 条
  • [1] Characterization of scatter magnitude and distribution in dedicated breast computed tomography with bowtie filters
    Kontson, Kimberly
    Jennings, Robert J.
    JOURNAL OF MEDICAL IMAGING, 2014, 1 (03)
  • [2] Deep learning for x-ray scatter correction in dedicated breast CT
    Pautasso, Juan J.
    Caballo, Marco
    Mikerov, Mikhail
    Boone, John M.
    Michielsen, Koen
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2023, 50 (04) : 2022 - 2036
  • [3] The influence of bowtie filters on x-ray CT signals
    Whiting, B
    Massoumzadeh, P
    O'Sullivan, J
    Snyder, D
    Williamson, J
    MEDICAL PHYSICS, 2005, 32 (06) : 2056 - 2056
  • [4] Bowtie filters for dedicated breast CT: Analysis of bowtie filter material selection
    Kontson, Kimberly
    Jennings, Robert J.
    MEDICAL PHYSICS, 2015, 42 (09) : 5270 - 5277
  • [5] Library based x-ray scatter correction for dedicated cone beam breast CT
    Shi, Linxi
    Vedantham, Srinivasan
    Karellas, Andrew
    Zhu, Lei
    MEDICAL PHYSICS, 2016, 43 (08) : 4529 - 4544
  • [6] Bowtie filters for dedicated breast CT: Theory and computational implementation
    Kontson, Kimberly
    Jennings, Robert J.
    MEDICAL PHYSICS, 2015, 42 (03) : 1453 - 1462
  • [7] Evaluation of x-ray scatter properties in a dedicated cone-beam breast CT scanner
    Kwan, ALC
    Boone, JM
    Shah, N
    MEDICAL PHYSICS, 2005, 32 (09) : 2967 - 2975
  • [8] X-ray scatter correction method for dedicated breast computed tomography
    Sechopoulos, Ioannis
    MEDICAL PHYSICS, 2012, 39 (05) : 2896 - 2903
  • [9] An X-Ray Scatter Correction Method for Dedicated Breast Computed Tomography
    Sechopoulos, I.
    MEDICAL PHYSICS, 2012, 39 (06) : 3914 - 3914
  • [10] X-ray scatter correction for dedicated cone beam breast CT using a forward-projection model
    Shi, Linxi
    Vedantham, Srinivasan
    Karellas, Andrew
    Zhu, Lei
    MEDICAL PHYSICS, 2017, 44 (06) : 2312 - 2320