Comparison of four methods for assessing patient effective dose from radiological examinations

被引:27
作者
Theocharopoulos, N
Perisinakis, K
Damilakis, J
Varveris, H
Gourtsoyiannis, N
机构
[1] Univ Crete, Fac Med, Dept Phys Med, GR-71409 Iraklion, Greece
[2] Univ Crete, Fac Med, Dept Radiotherapy, GR-71409 Iraklion, Greece
[3] Univ Crete, Fac Med, Dept Radiol, GR-71409 Iraklion, Greece
关键词
effective dose; thermoluminescence dosimetry; DAP; ESD; energy imparted;
D O I
10.1118/1.1500769
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Three methods of indirect effective dose estimation were reviewed and compared to a direct effective dose determination method. An anthropomorphic phantom and thermoluminescence dosimetry were used to obtain dosimetric data associated with anterior-posterior (AP) abdominal radiography, posterior-anterior (PA) chest radiography, PA head radiography, and AP heart fluoroscopy. Effective dose was determined using: (i) organ specific dose values directly determined by thermoluminescence dosimeters, (ii) data published by National Radiological Protection Board (NRPB) and entrance surface dose (ESD), (iii) NRPB data and dose area product (DAP), (iv) energy imparted derived from DAR The effective dose values estimated from the Rando phantom measurements were 161, 32.3, and 8.4 muSv/projection for the abdomen, chest, and head radiographs, respectively. Cardiac fluoroscopy yielded an effective dose value of 111 muSv/min. The effective dose values obtained indirectly using NRPB data and DAP were in good agreement with directly assessed values in all simulated exposures (difference <8%). The effective doses using NRPB data and ESD values differed from directly assessed values by less than 15% for the radiographic exposures and 60% for heart fluoroscopy. The energy imparted method yielded 136, 31, and 6.6 μSv/projection for the abdomen, chest, and head radiographs, respectively, and 111 μSv/min for heart fluoroscopy. Indirect patient effective dose determination using the NRPB dosimetric data and the measured value of incident radiation allows for reliable patient effective dose estimates. The use of DAP rather than ESD is recommended because it yields accurate results even for complex radiologic exposures involving fluoroscopy. The value of energy imparted may be used for the accurate determination of patient effective dose, especially when specific organ dose values are not of interest. The calculation of energy imparted with the use of EAP provides a reliable starting point for estimation of effective dose from radiologic examinations for which dosimetric data are not provided by NRPB. (C) 2002 American Association of Physicists in Medicine.
引用
收藏
页码:2070 / 2079
页数:10
相关论文
共 37 条
[1]  
BO WJ, 1998, BASIC ATLAS SECTIONA
[2]   Radiation doses to patients undergoing scoliosis radiography [J].
Chamberlain, CC ;
Huda, W ;
Hojnowski, LS ;
Perkins, A ;
Scaramuzzino, A .
BRITISH JOURNAL OF RADIOLOGY, 2000, 73 (872) :847-853
[3]   AN INVESTIGATION INTO THE PERFORMANCE OF AN AUTOMATED QUALITY ASSURANCE AND DOSIMETRY SYSTEM IN DIAGNOSTIC-RADIOLOGY [J].
CHAPPLE, CL ;
FAULKNER, K ;
HARRISON, RM .
BRITISH JOURNAL OF RADIOLOGY, 1990, 63 (752) :635-639
[4]   Conceptus radiation dose and risk from cardiac catheter ablation procedures [J].
Damilakis, J ;
Theocharopoulos, N ;
Perisinakis, K ;
Manios, E ;
Dimitriou, P ;
Vardas, P ;
Gourtsoyiannis, N .
CIRCULATION, 2001, 104 (08) :893-897
[5]   Accidental embryo irradiation during barium enema examinations - An estimation of absorbed dose [J].
Damilakis, J ;
Perisinakis, K ;
Grammatikakis, J ;
Panayiotakis, G ;
Gourtsoyiannis, N .
INVESTIGATIVE RADIOLOGY, 1996, 31 (04) :242-245
[6]   The effect of beam tube potential variation on gonad dose to patients during chest radiography investigated using high sensitivity LiF:Mg,Cu,P thermoluminescent dosemeters [J].
Fung, KKL ;
Gilboy, WB .
BRITISH JOURNAL OF RADIOLOGY, 2001, 74 (880) :358-367
[7]   COMPARISON OF 2 METHODS FOR ASSESSING PATIENT DOSE FROM COMPUTED-TOMOGRAPHY [J].
GELEIJNS, J ;
VANUNNIK, JG ;
ZOETELIEF, J ;
ZWEERS, D ;
BROERSE, JJ .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (796) :360-365
[8]   Computation of energy imparted in diagnostic radiology [J].
Gkanatsios, NA ;
Huda, W .
MEDICAL PHYSICS, 1997, 24 (04) :571-579
[9]   NORMALIZED ORGAN DOSES FOR VARIOUS DIAGNOSTIC RADIOLOGIC PROCEDURES [J].
GRAY, JE ;
RAGOZZINO, MW ;
VANLYSEL, MS ;
BURKE, TM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (03) :463-470
[10]  
HART D, 1994, NRPBR262