Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

被引:2
|
作者
Fenoglietto, P. [1 ]
Khodri, M. [2 ]
Nguyen, D. [3 ]
Josserand-Pietri, F. [4 ]
Ailleres, N. [1 ]
机构
[1] Inst Reg Canc Montpellier ICM, Dept Oncol Radiotherapie, F-34298 Montpellier 5, France
[2] ICL, Dept Oncol Radiotherapie, St Etienne, France
[3] Grp ORLAM, Serv Radiotherapie, Macon, France
[4] Grp ORLAM, Serv Radiotherapie, Villeurbanne, France
关键词
VOLUMETRIC-MODULATED ARC; QUALITY-ASSURANCE; CLINICAL IMPLEMENTATION; PROSTATE-CANCER; DYNAMIC IMRT; RADIOTHERAPY; DELIVERY; THERAPY; RAPIDARC; VERIFICATION;
D O I
10.1186/s13014-015-0577-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc (TM) device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. Patients and Methods: Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc (TM) technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm. Results: The number of points passing the gamma criteria between the measured and computed doses was 94.79 +/- 2.57 % for linac A and 94.61 +/- 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 +/- 1.26 % and 98.59 +/- 1.20 % points passing the threshold were obtained for linacs A and B, respectively. The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 +/- 0.06 % of the points passing the criteria. Conclusion: For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] IN-VIVO DETERMINATION OF THE ACCURACY OF FIELD MATCHING IN BREAST-CANCER IRRADIATION USING AN ELECTRONIC PORTAL IMAGING DEVICE
    HOLMBERG, O
    HUIZENGA, H
    IDZES, MHM
    LEBESQUE, JV
    VIJLBRIEF, RE
    MIJNHEER, BJ
    RADIOTHERAPY AND ONCOLOGY, 1994, 33 (02) : 157 - 166
  • [42] Development of Artificial Intelligence as a Conversion Tool for Cine Electronic Portal Imaging Device Images to Radiotherapy Dosimetry: Preliminary Study
    Ramadhan M.M.
    Wibowo W.E.
    Pawiro S.A.
    Iranian Journal of Medical Physics, 2022, 19 (05): : 296 - 304
  • [43] Electronic portal imaging device detection of radioopaque markers for the evaluation of prostate position during megavoltage irradiation: A clinical study
    Vigneault, E
    Pouliot, J
    Laverdiere, J
    Roy, J
    Dorion, M
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (01): : 205 - 212
  • [44] Two-dimensional measurement of photon beam attenuation by the treatment couch and immobilization devices using an electronic portal imaging device
    Vieira, SC
    Kaatee, RSJP
    Dirkx, MLP
    Heijmen, BJM
    MEDICAL PHYSICS, 2003, 30 (11) : 2981 - 2987
  • [45] Technical aspects of daily online positioning of the prostate for three-dimensional conformal radiotherapy using an electronic portal imaging device
    Herman, MG
    Pisansky, TM
    Kruse, JJ
    Prisciandaro, JI
    Davis, BJ
    King, BF
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2003, 57 (04): : 1131 - 1140
  • [46] Assessment of Multi-leaf Collimator Positional Accuracy Using Radiochromic EBT3 Film and an Electronic Portal Imaging Device
    Forough Nikeghbali
    Abolfazl Nickfarjam
    Fateme Shirani
    Nouraddin Abdi Goushbolagh
    Nasim Namiranian
    Bagher Farhood
    Masoud Shabani
    Kazem Vakili
    Journal of the Korean Physical Society, 2020, 76 : 795 - 801
  • [47] Quality assurance of MLC leaf position accuracy and relative dose effect at the MLC abutment region using an electronic portal imaging device
    Sumida, Iori
    Yamaguchi, Hajime
    Kizaki, Hisao
    Koizumi, Masahiko
    Ogata, Toshiyuki
    Takahashi, Yutaka
    Yoshioka, Yasuo
    JOURNAL OF RADIATION RESEARCH, 2012, 53 (05) : 798 - 806
  • [48] Feasibility study using a Ni-Ti stent and electronic portal imaging to localize the prostate during radiotherapy
    Carl, J
    Lund, B
    Larsen, EH
    Nielsen, J
    RADIOTHERAPY AND ONCOLOGY, 2006, 78 (02) : 199 - 206
  • [49] Daily organ tracking in intensity-modulated radiotherapy of prostate cancer using an electronic portal imaging device with a dose saving acquisition mode
    Vetterli, Daniel
    Thalmann, Sandrine
    Behrensmeier, Frank
    Kemmerling, Ludger
    Born, Ernst J.
    Mini, Roberto
    Greiner, Richard H.
    Aebersold, Daniel M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 79 (01) : 101 - 108
  • [50] Quick, efficient and effective patient-specific intensity-modulated radiation therapy quality assurance using log file and electronic portal imaging device
    Kumar, Rajesh
    Amols, H. I.
    Lovelock, M.
    Sharma, S. D.
    Datta, D.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2017, 13 (02) : 297 - 303