Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study

被引:2
|
作者
Fenoglietto, P. [1 ]
Khodri, M. [2 ]
Nguyen, D. [3 ]
Josserand-Pietri, F. [4 ]
Ailleres, N. [1 ]
机构
[1] Inst Reg Canc Montpellier ICM, Dept Oncol Radiotherapie, F-34298 Montpellier 5, France
[2] ICL, Dept Oncol Radiotherapie, St Etienne, France
[3] Grp ORLAM, Serv Radiotherapie, Macon, France
[4] Grp ORLAM, Serv Radiotherapie, Villeurbanne, France
关键词
VOLUMETRIC-MODULATED ARC; QUALITY-ASSURANCE; CLINICAL IMPLEMENTATION; PROSTATE-CANCER; DYNAMIC IMRT; RADIOTHERAPY; DELIVERY; THERAPY; RAPIDARC; VERIFICATION;
D O I
10.1186/s13014-015-0577-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To verify the accuracy of volumetric arc therapy (VMAT) using the RapidArc (TM) device when switching patients from one single linear accelerator (linac) to a paired energy and mechanics "twin" linac without reoptimization of the original treatment plan. Patients and Methods: Four centers using 8 linacs were involved in this study. Seventy-four patients previously treated with the 6MV photon RapidArc (TM) technique were selected for analysis, using 242 measurements. In each institution, all patients were planned on linac A, and their plans were verified both on linac A and on the twin linac B. Verifications were done using the amorphous silicium electronic portal imager (EPID) of the linacs and were analyzed with the EpiQa software (Epidos, Bratislavia, Slovakia). The gamma index formalism was used for validation with a double threshold of 3 % and 3 mm with a measurement resolution of 0.39 mm/pixel, and a smoothed resolution of approximately 2.5 mm. Results: The number of points passing the gamma criteria between the measured and computed doses was 94.79 +/- 2.57 % for linac A and 94.61 +/- 2.46 % for linac B. Concerning the smoothed measurement analysis, 98.67 +/- 1.26 % and 98.59 +/- 1.20 % points passing the threshold were obtained for linacs A and B, respectively. The difference between the 2 dose matrices acquired on the EPID was very small, with 99.92 +/- 0.06 % of the points passing the criteria. Conclusion: For linacs sharing the same mechanical and energy parameters, this study tends to indicate that patients may be safely switched from treatment with one linac to treatment with its twin linac using the same VMAT plan.
引用
收藏
页数:9
相关论文
共 50 条
  • [31] Simple Proposal for Dosimetry with an Elekta iViewGT™ Electronic Portal Imaging Device (EPID) Using Commercial Software Modules
    Liebich, Janett
    Licher, Joerg
    Scherf, Christian
    Kara, Eugen
    Koch, Nadine
    Roedel, Claus
    Ramm, Ulla
    STRAHLENTHERAPIE UND ONKOLOGIE, 2011, 187 (05) : 316 - 321
  • [32] VERIFICATION OF LUNG ATTENUATOR POSITIONING BEFORE TOTAL-BODY IRRADIATION USING AN ELECTRONIC PORTAL IMAGING DEVICE
    GLADSTONE, DJ
    VANHERK, M
    CHIN, LM
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02): : 449 - 454
  • [33] Dosimetric verification of intensity modulated beams produced with dynamic multileaf collimation using an electronic portal imaging device
    Pasma, KL
    Dirkx, MLP
    Kroonwijk, M
    Visser, AG
    Heijmen, BJM
    MEDICAL PHYSICS, 1999, 26 (11) : 2373 - 2378
  • [34] Validation of Three-dimensional Electronic Portal Imaging Device-based PerFRACTION™ Software for Patient-Specific Quality Assurance
    Sait, A. Aziz
    Figueredo, Jason
    Jones, Glenn W.
    Jones, Kai
    McGowan, Thomas
    Kapatoes, Jeff
    Ramaseshan, Ramani
    JOURNAL OF MEDICAL PHYSICS, 2019, 44 (01) : 16 - 20
  • [35] An Efficient Way to Confirm the Coincidence of SRS Collimator and Radiation Beam Axes Using Electronic Portal Imaging Device(EPID)
    Kang, D. H.
    Huang, S. H. M.
    Deng, X. W.
    Zhou, Y. J.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1841 - +
  • [36] On the accuracy of localization achievable in fiducial-based stereoscopic image registration system using an electronic portal imaging device
    Ung, N. M.
    Wee, L.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (02) : 205 - 213
  • [37] The experience of a developing country using an electronic portal imaging device for the verification of patient positioning and dosimetry in radiotherapy for prostate cancer
    Farhat, Leila
    Fourati, N.
    Mnejja, W.
    Daoud, J.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2018, 17 (03) : 297 - 301
  • [38] Monte Carlo simulation of a novel water-equivalent electronic portal imaging device using plastic scintillating fibers
    Teymurazyan, A.
    Pang, G.
    MEDICAL PHYSICS, 2012, 39 (03) : 1518 - 1529
  • [39] Fast 3D dosimetric verifications based on an electronic portal imaging device using a GPU calculation engine
    Zhu, Jinhan
    Chen, Lixin
    Chen, Along
    Luo, Guangwen
    Deng, Xiaowu
    Liu, Xiaowei
    RADIATION ONCOLOGY, 2015, 10
  • [40] In vivo dosimetry for prostate cancer patients using an electronic portal imaging device (EPID); demonstration of internal organ motion
    Kroonwijk, M
    Pasma, KL
    Quint, S
    Koper, PCM
    Visser, AG
    Heijmen, BJM
    RADIOTHERAPY AND ONCOLOGY, 1998, 49 (02) : 125 - 132