Radiation Dose for Body CT Protocols: Variability of Scanners at One Institution

被引:45
作者
Jaffe, Tracy A. [1 ]
Yoshizumi, Terry T. [1 ]
Toncheva, Greta [1 ]
Anderson-Evans, Colin [1 ]
Lowry, Carolyn [1 ]
Miller, Chad M. [1 ]
Nelson, Rendon C. [1 ]
Ravin, Carl E. [1 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Durham, NC 27710 USA
关键词
abdominal imaging; CT dose; MDCT; pulmonary angiography; TUBE CURRENT MODULATION; ROW HELICAL CT; PULMONARY-EMBOLISM; EMERGENCY-DEPARTMENT; OPTIMIZATION; ANGIOGRAPHY; REDUCTION; ABDOMEN; ENHANCEMENT; KILOVOLTAGE;
D O I
10.2214/AJR.09.2330
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
OBJECTIVE. The objective of our study was to determine, using an anthropomorphic phantom, whether patients are subject to variable radiation doses based on scanner assignment for common body CT studies. MATERIALS AND METHODS. Twenty metal oxide semiconductor field effect transistor dosimeters were placed in a medium-sized anthropomorphic phantom of a man. Pulmonary embolism and chest, abdomen, and pelvis protocols were used to scan the phantom three times with GE Healthcare scanners in four configurations and one 64-MDCT Siemens Healthcare scanner. Organ doses were averaged, and effective doses were calculated with weighting factors. RESULTS. The mean effective doses for the pulmonary embolism protocol ranged from 9.9 to 18.5 mSv and for the chest, abdomen, and pelvis protocol from 6.7 to 18.5 mSv. For the pulmonary embolism protocol, the mean effective dose from the Siemens Healthcare 64-MDCT scanner was significantly lower than that from the 16- and 64-MDCT GE Healthcare scanners (p < 0.001). The mean effective dose from the GE 4-MDCT scanner was significantly lower than that for the GE 16-MDCT scanner (p < 0.001) but not the GE 64-MDCT scanner (p = 0.02). For the chest, abdomen, and pelvis protocol, all mean effective doses from the GE scanners were significantly different from one another (p < 0.001), the lowest mean effective dose being found with use of a single-detector CT scanner and the highest with a 4-MDCT scanner. For the chest, abdomen, and pelvis protocols, the difference between the mean effective doses from the GE Healthcare and Siemens Healthcare 64-MDCT scanners was not statistically significant (p = 0.89). CONCLUSION. According to phantom data, patients are subject to different radiation exposures for similar body CT protocols depending on scanner assignment. In general, doses are lowest with use of 64-MDCT scanners.
引用
收藏
页码:1141 / 1147
页数:7
相关论文
共 37 条
[1]   CT pulmonary angiography and CT venography: Factors associated with vessel enhancement [J].
Arakawa, Hiroaki ;
Kohno, Tatsuo ;
Hiki, Taro ;
Kaji, Yasushi .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 189 (01) :156-161
[2]  
Boone John M, 2008, J Am Coll Radiol, V5, P132, DOI 10.1016/j.jacr.2007.07.008
[3]   Increasing utilization of computed tomography in the adult emergency department, 2000-2005 [J].
Broder J. ;
Warshauer D.M. .
Emergency Radiology, 2006, 13 (1) :25-30
[4]   Increased diagnosis of pulmonary embolism without a corresponding decline in mortality during the CT era [J].
Burge, A. J. ;
Freeman, K. D. ;
Klapper, P. J. ;
Haramati, L. B. .
CLINICAL RADIOLOGY, 2008, 63 (04) :381-386
[5]   Use of computed tomography scanning before and after siting in an emergency department [J].
Chen, JL ;
Dorfman, GS ;
Li, MC ;
Cronan, JJ .
ACADEMIC RADIOLOGY, 1996, 3 (08) :678-682
[6]   Utilization patterns and diagnostic yield of 3421 consecutive multidetector row computed tomography pulmonary angiograms in a busy emergency department [J].
Donohoo, Jay H. ;
Mayo-Smith, William W. ;
Pezzullo, John A. ;
Egglin, Thomas K. .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2008, 32 (03) :421-425
[7]   Automatic tube current modulation technique for multidetector CT: is it effective with a 64-detector CT? [J].
Funama Y. ;
Awai K. ;
Hatemura M. ;
Shimamura M. ;
Yanaga Y. ;
Oda S. ;
Yamashita Y. .
Radiological Physics and Technology, 2008, 1 (1) :33-37
[8]   Optimization of eight-element multi-detector row helical CT technology for evaluation of the abdomen [J].
Gupta, AK ;
Nelson, RC ;
Johnson, GA ;
Paulson, EK ;
Delong, DM ;
Yoshizumi, TT .
RADIOLOGY, 2003, 227 (03) :739-745
[9]   Estimated Radiation Dose Associated With Cardiac CT Angiography [J].
Hausleiter, Joerg ;
Meyer, Tanja ;
Hermann, Franziska ;
Hadamitzky, Martin ;
Krebs, Markus ;
Gerber, Thomas C. ;
McCollough, Cynthia ;
Martinoff, Stefan ;
Kastrati, Adnan ;
Schoemig, Albert ;
Achenbach, Stephan .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2009, 301 (05) :500-507
[10]   Technique factors and image quality as functions of patient weight at abdominal CT [J].
Huda, W ;
Scalzetti, EM ;
Levin, G .
RADIOLOGY, 2000, 217 (02) :430-435