Paper-Based Microfluidic Device with a Gold Nanosensor to Detect Arsenic Contamination of Groundwater in Bangladesh

被引:17
|
作者
Chowdury, Mosfera A. [1 ]
Walji, Noosheen [1 ]
Mahmud, Md. Almostasim [1 ]
MacDonald, Brendan D. [1 ]
机构
[1] Univ Ontario Inst Technol, Fac Engn & Appl Sci, 2000 Simcoe St North, Oshawa, ON L1H 7K4, Canada
来源
MICROMACHINES | 2017年 / 8卷 / 03期
关键词
microfluidic paper-based analytical devices; arsenic; gold nanosensor; nanoparticles; Bangladesh; groundwater; drinking water; DRINKING-WATER; POTABLE WATER; E; COLI; NANOPARTICLES; VARIABILITY; BIOSENSORS; PLATFORM; COMPACT; HEALTH;
D O I
10.3390/mi8030071
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this paper, we present a microfluidic paper-based analytical device (mu PAD) with a gold nanosensor functionalized with alpha-lipoic acid and thioguanine (Au-TA-TG) to detect whether the arsenic level of groundwater from hand tubewells in Bangladesh is above or below the World Health Organization (WHO) guideline level of 10 g/L. We analyzed the naturally occurring metals present in Bangladesh groundwater and assessed the interference with the gold nanosensor. A method was developed to prevent interference from alkaline metals found in Bangladesh groundwater (Ca, Mg, K and Na) by increasing the pH level on the mu PADs to 12.1. Most of the heavy metals present in the groundwater (Ni, Mn, Cd, Pb, and Fe II) did not interfere with the mu PAD arsenic tests; however, Fe III was found to interfere, which was also prevented by increasing the pH level on the mu PADs to 12.1. The mu PAD arsenic tests were tested with 24 groundwater samples collected from hand tubewells in three different districts in Bangladesh: Shirajganj, Manikganj, and Munshiganj, and the predictions for whether the arsenic levels were above or below the WHO guideline level agreed with the results obtained from laboratory testing. The mu PAD arsenic test is the first paper-based test validated using Bangladesh groundwater samples and capable of detecting whether the arsenic level in groundwater is above or below the WHO guideline level of 10 g/L, which is a step towards enabling the villagers who collect and consume the groundwater to test their own sources and make decisions about where to obtain the safest water.
引用
收藏
页数:15
相关论文
共 50 条
  • [31] Bladder cancer hunting: A microfluidic paper-based analytical device
    Jiang, Qingyun
    Han, Tingting
    Ren, Haijun
    Aziz, Aziz Ur Rehman
    Li, Na
    Zhang, Hangyu
    Zhang, Zhengyao
    Liu, Bo
    ELECTROPHORESIS, 2020, 41 (16-17) : 1509 - 1516
  • [32] Hydrogeochemistry and arsenic contamination of groundwater in the Ganges Delta Plain, Bangladesh
    Halima, M. A.
    Majumder, R. K.
    Nessa, S. A.
    Hiroshiro, Y.
    Uddin, M. J.
    Shimada, J.
    Jinno, K.
    JOURNAL OF HAZARDOUS MATERIALS, 2009, 164 (2-3) : 1335 - 1345
  • [33] Simultaneous determination of renal function biomarkers in urine using a validated paper-based microfluidic analytical device
    Rossini, Eduardo Luiz
    Milani, Maria Izabel
    Carrilho, Emanuel
    Pezza, Leonardo
    Pezza, Helena Redigolo
    ANALYTICA CHIMICA ACTA, 2018, 997 : 16 - 23
  • [34] Simultaneous colorimetric detection of nephrolithiasis biomarkers using a microfluidic paper-based analytical device
    Janrod, Manassawee
    Srisa-Art, Monpichar
    ANALYTICAL METHODS, 2023, 15 (06) : 752 - 761
  • [35] Arsenic Contamination in Groundwater of Bangladesh: Perspectives on Geochemical, Microbial and Anthropogenic Issues
    Anawar, Hossain M.
    Akai, Junji
    Mihaljevic, Martin
    Sikder, Arif M.
    Ahmed, Golam
    Tareq, Shafi M.
    Rahman, Mohammad M.
    WATER, 2011, 3 (04): : 1050 - 1076
  • [36] Validity of the latest research findings on causes of groundwater arsenic contamination in Bangladesh
    Abul Fazal, M
    Kawachi, T
    Ichion, E
    WATER INTERNATIONAL, 2001, 26 (03) : 380 - 389
  • [37] A New Paper-Based Microfluidic Device for Improved Detection of Nitrate in Water
    Charbaji, Amer
    Heidari-Bafroui, Hojat
    Anagnostopoulos, Constantine
    Faghri, Mohammad
    SENSORS, 2021, 21 (01) : 1 - 15
  • [38] GIS application for monitoring groundwater arsenic contamination in Bangladesh
    Kamal, R
    Karim, M
    Yan, KC
    GROUND WATER UPDATES, 2000, : 55 - 60
  • [39] Fabrication of paper-based microfluidic device using printed circuit technology
    Zhang, An-liang
    Zha, Yan
    AIP ADVANCES, 2012, 2 (02):
  • [40] The Development of Image Base, Portable Microfluidic Paper-Based Analytical Device
    Lu, Chuan-Pin
    Guo, Bo-Xian
    Fang, Zi-Qing
    Chung, Shu-Chiang
    PROCEEDINGS OF 2015 INTERNATIONAL CONFERENCE ON ORANGE TECHNOLOGIES (ICOT), 2015, : 144 - 148