Indicators of the maximum radiation dose to the skin during percutaneous coronary intervention in different target vessels

被引:18
作者
Chida, Koichi
Saito, Haruo
Kagaya, Yutaka
Kohzuki, Masahiro
Takai, Yoshihiro
Takahashi, Shoki
Yamada, Shogo
Zuguchi, Masayuki
机构
[1] Tohoku Univ, Fac Med, Sch Hlth Sci, Dept Radiol Technol,Aoba Ku, Sendai, Miyagi 9808575, Japan
[2] Tohoku Univ, Grad Sch Med, Dept Radiol, Sendai, Miyagi 980, Japan
[3] Tohoku Univ, Grad Sch Med, Dept Cardiovasc Med, Sendai, Miyagi 980, Japan
[4] Tohoku Univ, Grad Sch Med, Dept Internal Med & Rehabil Sci, Sendai, Miyagi 980, Japan
关键词
catheterization; coronary intervention; angiography; patient radiation exposure; coronary diseases;
D O I
10.1002/ccd.20830
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: To evaluate whether the maximum radiation dose to the patient's skin (MSD) can be estimated during percutaneous coronary intervention (PCI) procedures, we investigated the relationship between the MSD and fluoroscopic time, dose-area product (DAP), and body weight, separately analyzing the relationships for different target vessels. Background: Many cases of skin injury caused by excessive radiation exposure during cardiac intervention procedures have been reported. However, real-time maximum-dose monitoring of the skin is unavailable for many cardiac intervention procedures. Methods: We studied 197 consecutive PCI procedures that involved a single target vessel and were conducted. The DAP was measured, and the MSD was calculated by a skin-dose mapping software program (Caregraph). The target vessels of the PCI procedures were divided into four groups based on the AHA classification system: AHA 5-10, left anterior descending artery domain (LAD), AHA 11-15, left circumflex artery domain (LCx), AHA 1-3 = R 1-3, and AHA 4 = R 4. Results: The correlation coefficient (r) between the MSD and fluoroscopic time was higher for the right coronary artery (RCA) vessels (R 1-3, 0.852; R 4, 0.715) than for the left coronary artery (LCA) vessels (LAD, 0.527; LCx, 0.646), and the r value between the MSD and DAP was higher for the RCA vessels (R 1-3, 0.871; R 4, 0.898) than for the LCA vessels (LAD, 0.628; LCx, 0.694). Similarly, the correlation coefficient between the MSD and weight x fluoroscopic time (WFP) was higher for the RCA vessels (R 1-3, 0.874; R 4, 0.807) than for the LCA vessels (LAD, 0.551; LCx, 0.735). Conclusions: The DAP and WFP can be used to estimate the MSD during PCI in the RCA but not in the LCA, especially the LAD. (C) 2006 Wiley-Liss, Inc.
引用
收藏
页码:236 / 241
页数:6
相关论文
共 27 条
  • [1] [Anonymous], 60601243 IEC
  • [2] BAIM DS, 2000, GROSSMANS CARDIAC CA, P547
  • [3] Fundamentals of X-ray imaging and radiation safety
    Bashore, T
    [J]. CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2001, 54 (01) : 126 - 135
  • [4] Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures
    Chida, K
    Saito, H
    Otani, H
    Kohzuki, M
    Takahashi, S
    Yamada, S
    Shirato, K
    Zuguchi, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 186 (03) : 774 - 778
  • [5] Does digital acquisition reduce patients' skin dose in cardiac interventional procedures? An experimental study
    Chida, K
    Saito, H
    Zuguchi, M
    Shirotori, K
    Kumagai, S
    Nakayama, H
    Matsubara, K
    Kohzuki, M
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (04) : 1111 - 1114
  • [6] Real-time measurement of radiation exposure to patients during diagnostic coronary angiography and percutaneous interventional procedures
    Cusma, JT
    Bell, MR
    Wondrow, MA
    Taubel, JP
    Holmes, DR
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1999, 33 (02) : 427 - 435
  • [7] Real-time quantification and display of skin radiation during coronary angiography and intervention
    den Boer, A
    de Feijter, PJ
    Serruys, PW
    Roelandt, JRTC
    [J]. CIRCULATION, 2001, 104 (15) : 1779 - 1784
  • [8] *FOOD DRUG ADM, 1995, IMP INF PHY OTH HEAL
  • [9] RADIOTHERAPY VERIFICATION FILM FOR ESTIMATING CUMULATIVE ENTRANCE SKIN EXPOSURE FOR FLUOROSCOPIC EXAMINATIONS
    GEISE, RA
    ANSEL, HJ
    [J]. HEALTH PHYSICS, 1990, 59 (03): : 295 - 298
  • [10] GROSMAN W, 2000, GROSSMANS CARDIAC CA, P3