Blood collection tube-related alterations in analyte concentrations in quality control material and serum specimens

被引:23
作者
Bowen, Raffick A. R. [1 ]
Sattayapiwat, Annie [1 ]
Gounden, Verena [2 ]
Remaley, Alan T. [2 ]
机构
[1] Stanford Univ, Dept Pathol, Stanford, CA 94305 USA
[2] NIH, Dept Lab Med, Bethesda, MD 20892 USA
关键词
Blood collection tube; Quality control material; Thyroxine; Triiodothyronine; TOTAL TRIIODOTHYRONINE; INTERFERENCE; MANAGEMENT; DEVICES;
D O I
10.1016/j.clinbiochem.2013.11.003
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: Several previous studies have described the effects of interfering substances on clinical assay results; however, the effects of exogenous substances, particularly additives from blood collection tubes on quality control (QC) specimens and serum specimens have not been well examined. This study examines the effects of blood-collection tube additives on total triiodothyronine (TT3), and thyroxine (TT4), cortisol, and routine clinical chemistry tests in QC and serum specimens from apparently healthy volunteers. Methods: QC and serum specimens were poured or collected into different blood collection tubes. TT3 and TT4, cortisol, and routine chemistry tests were analyzed from the different blood-collection tube types. Results: The findings of this study demonstrate statistically and/or clinically significant blood collection tube-related alterations in the TT3, TT4, and cortisol concentrations of QC specimens and TT4 concentrations from serum specimens. Conclusions: These findings have important implications for clinical laboratories, demonstrating that QC specimens should ideally, like patients' specimens, be poured into blood collection tubes. This strategy would reveal any adverse effects caused by blood collection tubes, which otherwise would not likely be detected by most routine QC practices. The results of this study also show the importance of producing blood collection tubes that contain additives that are truly inert and do not adversely affect clinical laboratory testing. (C) 2013 The Canadian Society of Clinical Chemists. Published by Elsevier Inc. All rights reserved.
引用
收藏
页码:150 / 157
页数:8
相关论文
共 24 条
[11]  
GIGLIELLO JF, 1975, Patent No. 3920549
[12]   Suitability of collection tubes with separator gels for collecting and storing blood samples for therapeutic drug monitoring (TDM) [J].
Karppi, J ;
Åkerman, KK ;
Parviainen, M .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2000, 38 (04) :313-320
[13]   Laboratory quality control: Using patient data to assess analytical performance [J].
Kazmierczak, SC .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2003, 41 (05) :617-627
[14]  
KESSLER SB, 1977, Patent No. 4153739
[15]   Processing controls in blood collection tubes reveals interference [J].
Kricka, LJ ;
Park, JY ;
Senior, MB ;
Fontanilla, R .
CLINICAL CHEMISTRY, 2005, 51 (12) :2422-2423
[16]   INTRODUCTION TO SAMPLE-SIZE DETERMINATION AND POWER ANALYSIS FOR CLINICAL-TRIALS [J].
LACHIN, JM .
CONTROLLED CLINICAL TRIALS, 1981, 2 (02) :93-113
[17]  
LANDT M, 1993, CLIN CHEM, V39, P1712
[18]   Preanalytical management: serum vacuum tubes validation for routine clinical chemistry [J].
Lima-Oliveira, Gabriel ;
Lippi, Giuseppe ;
Salvagno, Gian Luca ;
Montagnana, Martina ;
Picheth, Geraldo ;
Guidi, Gian Cesare .
BIOCHEMIA MEDICA, 2012, 22 (02) :180-186
[19]  
Lin FC, 2001, LAB MED, V32, P588
[20]  
Lippi G, 2012, CLIN CHEM LAB MED, V16, P1