Dosimetric and treatment planning considerations for radiotherapy of the chest wall

被引:6
作者
Aspradakis, M. M. [1 ]
McCallum, H. M. [1 ]
Wilson, N. [1 ]
机构
[1] Newcastle Gen Hosp, Reg Med Phys Dept, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
关键词
D O I
10.1259/bjr/26575438
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Radiotherapy treatment planning calculations of the chest wall are complex due to missing tissue, the thin chest wall and the presence of lung. The accuracy of calculated dose is dependent on the type of algorithm employed. This work evaluates a collapsed cone (CC) and a pencil beam (PB) convolution model for radiotherapy planning of the chest wall. Various irradiation geometries simulating the chest wall have been examined and calculations were compared with measurements with an ionization chamber in epoxy resin water substitute and in low-density lung substitute blocks. A retrospective treatment planning study comprising 6 patients was carried out to evaluate the differences in the dose distributions and monitor units predicted by the two algorithms. The calculated dose in unit density medium was within +/- 1 % for the CC model and up to +/-2% for PB. In low density medium and under full scatter conditions, CC overestimated the dose by 1 % whereas PB overestimated the dose by 9%. In the tangential irradiation geometry with water and lung media, the PB overestimated dose to the isocentre by up to 10%, whereas the dose from CC was within 3%. From the treatment planning study calculated monitor units (MU) and doses were consistent with the experimental findings. The CC model is more accurate for radiotherapy treatment planning of the chest wall and especially when there is significant involvement of lung tissue.
引用
收藏
页码:828 / 836
页数:9
相关论文
共 23 条
[1]   A PENCIL BEAM MODEL FOR PHOTON DOSE CALCULATION [J].
AHNESJO, A ;
SAXNER, M ;
TREPP, A .
MEDICAL PHYSICS, 1992, 19 (02) :263-273
[2]   APPLICATION OF THE CONVOLUTION METHOD FOR CALCULATION OF OUTPUT FACTORS FOR THERAPY PHOTON BEAMS [J].
AHNESJO, A ;
KNOOS, T ;
MONTELIUS, A .
MEDICAL PHYSICS, 1992, 19 (02) :295-301
[3]   COLLAPSED CONE CONVOLUTION OF RADIANT ENERGY FOR PHOTON DOSE CALCULATION IN HETEROGENEOUS MEDIA [J].
AHNESJO, A .
MEDICAL PHYSICS, 1989, 16 (04) :577-592
[4]   Experimental verification of convolution /superposition photon dose calculations for radiotherapy treatment planning [J].
Aspradakis, MM ;
Morrison, RH ;
Richmond, ND ;
Steele, A .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (17) :2873-2893
[5]   Comparison of dose calculation algorithms in phantoms with lung equivalent heterogeneities under conditions of lateral electronic disequilibrium [J].
Carrasco, P ;
Jornet, N ;
Duch, MA ;
Weber, L ;
Ginjaume, M ;
Eudaldo, T ;
Jurado, D ;
Ruiz, A ;
Ribas, M .
MEDICAL PHYSICS, 2004, 31 (10) :2899-2911
[6]   MATCHING OF ADJACENT RADIATION BEAMS FOR ISOCENTRIC RADIOTHERAPY [J].
CASEBOW, MP .
BRITISH JOURNAL OF RADIOLOGY, 1984, 57 (680) :735-740
[7]   A SOLID WATER PHANTOM MATERIAL FOR RADIOTHERAPY X-RAY AND GAMMA-RAY BEAM CALIBRATIONS [J].
CONSTANTINOU, C ;
ATTIX, FH ;
PALIWAL, BR .
MEDICAL PHYSICS, 1982, 9 (03) :436-441
[8]   Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours [J].
Engelsman, M ;
Damen, EMF ;
Koken, PW ;
van 't Veld, AA ;
van Ingen, KM ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 60 (03) :299-309
[9]   The clinical implications of the collapsed cone planning algorithm [J].
Irvine, C ;
Morgan, A ;
Crellin, A ;
Nisbet, A ;
Beange, I .
CLINICAL ONCOLOGY, 2004, 16 (02) :148-154
[10]  
JOHNSON S, 2002, INT HDB BREAST CANC