Assay method for the perfluorooctyl bromide (perflubron) in rat blood by gas chromatography-mass spectrometry

被引:12
作者
Audran, M
Krafft, MP
De Ceaurriz, J
Mathurin, JC
Sicart, MT
Marion, B
Fabre, F
Bressolle, F [2 ]
机构
[1] Univ Montpellier I, Fac Pharm, Dept Biophys, Montpellier, France
[2] Univ Montpellier I, Fac Pharm, Lab Pharmacocinet Clin, Montpellier, France
[3] Inst Charles Sadron, F-67083 Strasbourg, France
[4] Lab Natl Depistage Dopage, Chatenay Malabry, France
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 1999年 / 734卷 / 02期
关键词
perfluorooctyl bromide; perflubron;
D O I
10.1016/S0378-4347(99)00357-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This paper describes a GC-MS method (SIM mode) for the analysis of perfluorooctyl bromide (perflubron, I) in rat blood. The chromatographic separation was performed by injection in the split mode using a CP-select 624 CB capillary column. Following destruction of the emulsion by addition of ethanol, the analytical procedure involves a liquid-liquid extraction with 1,1,2-trichlorotrifluoroethane. The bis(F-butyl)ethene (II) was used as internal standard. Observed retention times were 3.22 min for I and 2.32 min for II. Two calibration curves were used; linear detection responses were obtained for concentrations ranging from 0.009 to 0.9 mg/ml and from 0.9 to 13.5 mg/ml. The extraction efficiency averaged 50% for I and 93% for II. Precision ranged from 0.7 to 14%, and accuracy was between 91 and 109%. The limit of quantification was 9 mu g/ml. The method validation results indicate that the performance characteristics of the method fulfilled the requirements for assay method for use in pharmacokinetic studies. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:267 / 276
页数:10
相关论文
共 25 条
[1]   Effects of erythropoietin administration in training athletes and possible indirect detection in doping control [J].
Audran, M ;
Gareau, R ;
Matecki, S ;
Durand, F ;
Chenard, C ;
Sicart, MT ;
Marion, B ;
Bressolle, F .
MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1999, 31 (05) :639-645
[2]   FEMTOGRAM DETECTION OF PERFLUOROCARBON TRACERS USING CAPILLARY GAS CHROMATOGRAPHY-ELECTRON-CAPTURE NEGATIVE-ION CHEMICAL IONIZATION MASS-SPECTROMETRY [J].
BEGLEY, P ;
FOULGER, B ;
SIMMONDS, P .
JOURNAL OF CHROMATOGRAPHY, 1988, 445 (01) :119-128
[3]   EFFECT OF RECOMBINANT-HUMAN-ERYTHROPOIETIN TREATMENT ON BLOOD-PRESSURE AND SOME HEMATOLOGICAL PARAMETERS IN HEALTHY-MEN [J].
BERGLUND, B ;
EKBLOM, B .
JOURNAL OF INTERNAL MEDICINE, 1991, 229 (02) :125-130
[4]   Comparison of a direct and indirect population pharmacodynamic model: Application to recombinant human erythropoietin in athletes [J].
Bressolle, F ;
Audran, M ;
Gareau, R ;
Pham, TN ;
Gomeni, R .
JOURNAL OF PHARMACOKINETICS AND BIOPHARMACEUTICS, 1997, 25 (03) :263-275
[5]   Population pharmacodynamics for monitoring epoetin in athletes [J].
Bressolle, F ;
Audran, M ;
Gareau, R ;
Baynes, RD ;
Guidicelli, C ;
Gomeni, R .
CLINICAL DRUG INVESTIGATION, 1997, 14 (03) :233-242
[6]   Validation of liquid chromatographic and gas chromatographic methods - Applications to pharmacokinetics [J].
Bressolle, F ;
BrometPetit, M ;
Audran, M .
JOURNAL OF CHROMATOGRAPHY B-BIOMEDICAL APPLICATIONS, 1996, 686 (01) :3-10
[7]  
EKBLOM B, 1991, Scandinavian Journal of Medicine and Science in Sports, V1, P88
[8]   Erythropoietin abuse in athletes [J].
Gareau, R ;
Audran, M ;
Baynes, RD ;
Flowers, CH ;
Duvallet, A ;
Senecal, L ;
Brisson, GR .
NATURE, 1996, 380 (6570) :113-113
[9]  
GEYER RP, 1979, P 4 INT S PERFL BLOO, P3
[10]   Accuracy of fluorocrit in determination of blood perflubron concentration [J].
Henry, CJ ;
Tyler, JW ;
Fields, T ;
Brewer, WG ;
Royer, N .
ARTIFICIAL CELLS BLOOD SUBSTITUTES AND IMMOBILIZATION BIOTECHNOLOGY, 1998, 26 (03) :285-292