Modeling Solvent Evaporation During the Manufacture of Controlled Drug-Release Coatings and the Impact on Release Kinetics

被引:29
作者
Kim, Chang-Soo [1 ]
Saylor, David M. [1 ]
McDermott, Martin K. [1 ]
Patwardhan, Dinesh V. [1 ]
Warren, James A. [2 ]
机构
[1] US FDA, Div Chem & Mat Sci, Off Sci & Engn Labs, Ctr Device & Radiol Hlth, Silver Spring, MD 20903 USA
[2] NIST, Mat Sci & Engn Labs, Div Met, Gaithersburg, MD 20899 USA
关键词
drug delivery/release; stents; microstructure; computer modeling; simulation; controlled release; DIFFUSION-COEFFICIENTS; MACROPOROUS POLYMERS; ELUTING STENTS; SYSTEMS;
D O I
10.1002/jbm.b.31336
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To improve functionality and performance, controlled drug-release coatings comprised of drug and polymer are integrated with traditional medical devices, e.g., drug eluting stents. Depending on manufacturing conditions, these coatings can exhibit complex microstructures. Previously, a thermodynamically consistent model was developed for microstructure evolution in these systems to establish relationships between process variables, microstructure, and the subsequent release kinetics. Calculations based on the model were, in general, consistent with experimental findings. However, because of assumptions regarding the evaporation of solvent during fabrication, the model was unable to capture variations through the coating thickness that are observed experimentally. Here, a straightforward method is introduced to incorporate solvent evaporation explicitly into the model. Calculations are used to probe the impact of solvent evaporation rate and drug loading on the microstructure that forms during manufacturing and subsequent drug release kinetics. The predicted structures and release kinetics are found to be consistent with experimental observations. Further, the calculations demonstrate that solvent evaporation rate can be as critical to device performance as the amount of drug within the coating. For example, changes of a factor of five in the amount of drug released were observed by modifying the rate of solvent evaporation during manufacturing. (C) 2009 Wiley Periodicals. Inc.* J Biomed Mater Res Part B: Appl Biomater 90B: 688-699, 2009
引用
收藏
页码:688 / 699
页数:12
相关论文
共 26 条
[1]  
Allen SM, 1977, J PHYS, V38, pC7
[2]  
Bawa R., 1985, J CONTROLLED RELEASE, V1, P259
[3]   ON SPINODAL DECOMPOSITION [J].
CAHN, JW .
ACTA METALLURGICA, 1961, 9 (09) :795-801
[4]   FREE ENERGY OF A NONUNIFORM SYSTEM .1. INTERFACIAL FREE ENERGY [J].
CAHN, JW ;
HILLIARD, JE .
JOURNAL OF CHEMICAL PHYSICS, 1958, 28 (02) :258-267
[5]   Stent thrombosis late after implantation of first-generation drug-eluting stents - A cause for concern [J].
Camenzind, Edoardo ;
Steg, P. Gabriel ;
Wijns, William .
CIRCULATION, 2007, 115 (11) :1440-1455
[6]   EQUATION RELATING SELF-DIFFUSION AND MUTUAL DIFFUSION-COEFFICIENTS IN POLYMER-SOLVENT SYSTEMS [J].
DUDA, JL ;
NI, YC ;
VRENTAS, JS .
MACROMOLECULES, 1979, 12 (03) :459-462
[7]   Bringing reality to drug-eluting stents [J].
Faxon, DP .
CIRCULATION, 2004, 109 (02) :140-142
[8]   Thermodynamics of high polymer solutions [J].
Flory, PJ .
JOURNAL OF CHEMICAL PHYSICS, 1941, 9 (08) :660-661
[9]   Phase field theory of heterogeneous crystal nucleation [J].
Granasy, Laszlo ;
Pusztai, Tamas ;
Saylor, David ;
Warren, James A. .
PHYSICAL REVIEW LETTERS, 2007, 98 (03)
[10]  
Grassi Mario, 2005, Current Drug Delivery, V2, P97, DOI 10.2174/1567201052772906