Size and positioning reproducibility of an 192Ir brachytherapy stepping source

被引:4
作者
Berndt, A
Rickey, DW
Rathee, S
Bews, J
机构
[1] Cancercare Manitoba, Winnipeg, MB R3E 0V9, Canada
[2] Univ Manitoba, Dept Phys, Winnipeg, MB R3T 2N2, Canada
关键词
brachytherapy; source dimensions;
D O I
10.1118/1.598874
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this paper we describe techniques for measuring the dimensions and position reproducibility of an Ir-192 brachytherapy stepping source. Measurements were carried out using a 0.25 X 10 X 152 mm(3) collimator placed in front of a detector of our own design. The brachytherapy source was translated past the collimator in 0.025 mm increments using a stepper motor. The source was found to be 3.58 mm long and 0.69 mm wide, which is in good agreement with the manufacturer's values of 3.5x0.6 mm(2). The source position was reproducible to within 0.12 mm. (C) 2000 American Association of Physicists in Medicine.