Kilovoltage intrafraction motion monitoring and target dose reconstruction for stereotactic volumetric modulated arc therapy of tumors in the liver

被引:50
作者
Poulsen, Per Rugaard [1 ,2 ]
Worm, Esben S. [1 ,3 ]
Petersen, Jorgen B. B. [3 ]
Grau, Cai [1 ,2 ]
Fledelius, Walther [1 ,3 ]
Hoyer, Morten [1 ,2 ]
机构
[1] Aarhus Univ Hosp, Dept Oncol, DK-8000 Aarhus C, Denmark
[2] Aarhus Univ Hosp, Inst Clin Med, DK-8000 Aarhus C, Denmark
[3] Aarhus Univ Hosp, Dept Med Phys, DK-8000 Aarhus C, Denmark
关键词
Image-guided radiotherapy; Intrafraction motion monitoring; Dose reconstruction; BODY RADIATION-THERAPY; BREATHING MOTION; FIDUCIAL MARKERS; RADIOTHERAPY; LUNG; SBRT; DISTRIBUTIONS; PROJECTIONS; MAGNITUDE; TRACKING;
D O I
10.1016/j.radonc.2014.05.007
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To use intrafraction kilovoltage (kV) imaging during liver stereotactic body radiotherapy (SBRT) delivered by volumetric modulated arc therapy (VMAT) to estimate the intra-treatment target motion and to reconstruct the delivered target dose. Methods: Six liver SBRT patients with 2-3 implanted gold markers received SBRT in three fractions of 18.75 Gy or 25 Gy. CIV-to-PTV margins of 5 mm in the axial plane and 10 mm in the cranio-caudal directions were applied. A VMAT plan was designed to give minimum target doses of 95% (CTV) and 67% (PTV). At each fraction, the 3D marker trajectory was estimated by fluoroscopic kV imaging throughout treatment delivery and used to reconstruct the actually delivered CTV dose. The reduction in D-95 (minimum dose to 95% of the CTV) relative to the planned D-95 was calculated. Results: The kV position estimation had mean root-mean-square errors of 0.36 mm and 0.47 mm parallel and perpendicular to the kV imager, respectively. Intrafraction motion caused a mean 3D target position error of 2.9 mm and a mean D-95 reduction of 6.0%. The D-95 reduction correlated with the mean 3D target position error during a fraction. Conclusions: Kilovoltage imaging for detailed motion monitoring. with dose reconstruction of VMAT-based liver SBRT was demonstrated for the first time showing large dosimetric impact of intrafraction tumor motion. (C) 2014 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:424 / 430
页数:7
相关论文
共 39 条
  • [1] Prostate intrafraction motion evaluation using kV fluoroscopy during treatment delivery: A feasibility and accuracy study
    Adamson, Justus
    Wu, Qiuwen
    [J]. MEDICAL PHYSICS, 2008, 35 (05) : 1793 - 1806
  • [2] Predictive uncertainty in infrared marker-based dynamic tumor tracking with Vero4DRT
    Akimoto, Mami
    Nakamura, Mitsuhiro
    Mukumoto, Nobutaka
    Tanabe, Hiroaki
    Yamada, Masahiro
    Matsuo, Yukinori
    Monzen, Hajime
    Mizowaki, Takashi
    Kokubo, Masaki
    Hiraoka, Masahiro
    [J]. MEDICAL PHYSICS, 2013, 40 (09)
  • [3] Development and clinical evaluation of automatic fiducial detection for tumor tracking in cine megavoltage images during volumetric modulated arc therapy
    Azcona, Juan Diego
    Li, Ruijiang
    Mok, Edward
    Hancock, Steven
    Xing, Lei
    [J]. MEDICAL PHYSICS, 2013, 40 (03)
  • [4] Monte Carlo calculation of VMAT and helical tomotherapy dose distributions for lung stereotactic treatments with intra-fraction motion
    Belec, J.
    Clark, B. G.
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (09) : 2807 - 2821
  • [5] Stereotactic body radiation therapy: The report of AAPM Task Group 101
    Benedict, Stanley H.
    Yenice, Kamil M.
    Followill, David
    Galvin, James M.
    Hinson, William
    Kavanagh, Brian
    Keall, Paul
    Lovelock, Michael
    Meeks, Sanford
    Papiez, Lech
    Purdie, Thomas
    Sadagopan, Ramaswamy
    Schell, Michael C.
    Salter, Bill
    Schlesinger, David J.
    Shiu, Almon S.
    Solberg, Timothy
    Song, Danny Y.
    Stieber, Volker
    Timmerman, Robert
    Tome, Wolfgang A.
    Verellen, Dirk
    Wang, Lu
    Yin, Fang-Fang
    [J]. MEDICAL PHYSICS, 2010, 37 (08) : 4078 - 4101
  • [6] A technique for respiratory-gated radiotherapy treatment verification with an EPID in cine mode
    Berbeco, RI
    Neicu, T
    Rietzel, E
    Chen, GTY
    Jiang, SB
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (16) : 3669 - 3679
  • [7] FIRST DEMONSTRATION OF COMBINED KV/MV IMAGE-GUIDED REAL-TIME DYNAMIC MULTILEAF-COLLIMATOR TARGET TRACKING
    Cho, Byungchul
    Poulsen, Per R.
    Sloutsky, Alex
    Sawant, Amit
    Keall, Paul J.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 74 (03): : 859 - 867
  • [8] PTV dose prescription strategies for SBRT of metastatic liver tumours
    de Pooter, Jacco A.
    Wunderink, Wouter
    Romero, Alejandra Mendez
    Storchi, Pascal R. M.
    Heijmen, Ben J. M.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 85 (02) : 260 - 266
  • [9] Real-time dynamic MLC tracking for inversely optimized arc radiotherapy
    Falk, Marianne
    af Rosenschold, Per Munck
    Keall, Paul
    Cattell, Herbert
    Cho, Byung Chul
    Poulsen, Per
    Povzner, Sergey
    Sawant, Amit
    Zimmerman, Jens
    Korreman, Stine
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 94 (02) : 218 - 223
  • [10] Motion as a perturbation: Measurement-guided dose estimates to moving patient voxels during modulated arc deliveries
    Feygelman, Vladimir
    Stambaugh, Cassandra
    Zhang, Geoffrey
    Hunt, Dylan
    Opp, Daniel
    Wolf, Theresa K.
    Nelms, Benjamin E.
    [J]. MEDICAL PHYSICS, 2013, 40 (02)