Comparison of Multidetector Computed Tomography and Flat-Panel Computed Tomography Regarding Visualization of Cortical Fractures, Cortical Defects, and Orthopedic Screws A Phantom Study

被引:4
作者
Neubauer, Jakob [1 ]
Benndorf, Matthias [1 ]
Lang, Hannah [1 ]
Lampert, Florian [2 ]
Kemna, Lars
Konstantinidis, Lukas
Neubauer, Claudia [1 ]
Reising, Kilian [3 ]
Zajonc, Horst [2 ]
Kotter, Elmar [1 ]
Langer, Mathias [1 ]
Goerke, Sebastian M. [2 ]
机构
[1] Univ Med Ctr Freiburg, Dept Radiol, Freiburg, Baden Wurttembe, Germany
[2] Univ Med Ctr Freiburg, Dept Plast & Hand Surg, Freiburg, Baden Wurttembe, Germany
[3] Univ Med Ctr Freiburg, Dept Orthoped & Traumatol, Freiburg, Baden Wurttembe, Germany
关键词
CONE-BEAM CT; IMAGE QUALITY; ARTIFACTS; WRIST; ARTHROGRAPHY; INJURIES; SCANS;
D O I
10.1097/MD.0000000000001231
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
To compare the visualization of cortical fractures, cortical defects, and orthopedic screws in a dedicated extremity flat-panel computed tomography (FPCT) scanner and a multidetector computed tomography (MDCT) scanner. We used feet of European roe deer as phantoms for cortical fractures, cortical defects, and implanted orthopedic screws. FPCT and MDCT scans were performed with equivalent dose settings. Six observers rated the scans according to number of fragments, size of defects, size of defects opposite orthopedic screws, and the length of different screws. The image quality regarding depiction of the cortical bone was assessed. The gold standard (real number of fragments) was evaluated by autopsy. The correlation of reader assessment of fragments, cortical defects, and screws with the gold standard was similar for FPCT and MDCT. Three readers rated the subjective image quality of the MDCT to be higher, whereas the others showed no preferences. Although the image quality was rated higher in the MDCT than in the FPCT by 3 out of 6 observers, both modalities proved to be comparable regarding the visualization of cortical fractures, cortical defects, and orthopedic screws and of use to musculoskeletal radiology regarding fracture detection and postsurgical evaluation in our experimental setting.
引用
收藏
页数:8
相关论文
共 22 条
[1]  
[Anonymous], 2011, DICOM 14
[2]   MULTIPLE SIGNIFICANCE TESTS - THE BONFERRONI METHOD .10. [J].
BLAND, JM ;
ALTMAN, DG .
BRITISH MEDICAL JOURNAL, 1995, 310 (6973) :170-170
[3]   Distal Radius in Adolescent Girls with Anorexia Nervosa: Trabecular Structure Analysis with High-Resolution Flat-Panel Volume CT [J].
Bredella, Miriam A. ;
Misra, Madhusmita ;
Miller, Karen K. ;
Madisch, Ijad ;
Sarwar, Ammar ;
Cheung, Arnold ;
Klibanski, Anne ;
Gupta, Rajiv .
RADIOLOGY, 2008, 249 (03) :938-946
[4]   Comparison of organ doses and image quality between CT and flat panel XperCT scans in wrist and inner ear examinations [J].
Damet, J. ;
Sans-Merce, M. ;
Mieville, F. ;
Becker, M. ;
Poletti, P. A. ;
Verdun, F. R. ;
Baechler, S. .
RADIATION PROTECTION DOSIMETRY, 2010, 139 (1-3) :164-168
[5]   Cone-beam computed tomography: a new low dose, high resolution imaging technique of the wrist, presentation of three cases with technique [J].
De Cock, Jens ;
Mermuys, Koen ;
Goubau, Jean ;
Van Petegem, Simon ;
Houthoofd, Brecht ;
Casselman, Jan W. .
SKELETAL RADIOLOGY, 2012, 41 (01) :93-96
[6]   Beam hardening artefacts occur in dental implant scans with the NewTom® cone beam CT but not with the dental 4-row multidetector CT [J].
Draenert, F. G. ;
Coppenrath, E. ;
Herzog, P. ;
Muller, S. ;
Mueller-Lisse, U. G. .
DENTOMAXILLOFACIAL RADIOLOGY, 2007, 36 (04) :198-203
[7]   Finger fractures imaging: accuracy of cone-beam computed tomography and multislice computed tomography [J].
Faccioli, Niccolo ;
Foti, Giovanni ;
Barillari, Marco ;
Atzei, Andrea ;
Mucelli, Roberto Pozzi .
SKELETAL RADIOLOGY, 2010, 39 (11) :1087-1095
[8]   Metallic Artifacts From Internal Scaphoid Fracture Fixation Screws Comparison Between C-Arm Flat-Panel, Cone-Beam, and Multidetector Computed Tomography [J].
Finkenstaedt, Tim ;
Morsbach, Fabian ;
Calcagni, Maurizio ;
Vich, Magdalena ;
Pfirrmann, Christian W. A. ;
Alkadhi, Hatem ;
Runge, Val M. ;
Andreisek, Gustav ;
Guggenberger, Roman .
INVESTIGATIVE RADIOLOGY, 2014, 49 (08) :532-539
[9]   CT image quality improvement using adaptive iterative dose reduction with wide-volume acquisition on 320-detector CT [J].
Gervaise, Alban ;
Osemont, Benoit ;
Lecocq, Sophie ;
Noel, Alain ;
Micard, Emilien ;
Felblinger, Jacques ;
Blum, Alain .
EUROPEAN RADIOLOGY, 2012, 22 (02) :295-301
[10]   In Vitro High-Resolution Flat-Panel Computed Tomographic Arthrography for Artificial Cartilage Defect Detection Comparison With Multidetector Computed Tomography [J].
Guggenberger, Roman ;
Winklhofer, Sebastian ;
Spiczak, Jochen V. ;
Andreisek, Gustav ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2013, 48 (08) :614-621