Electrochemical Immunochip Sensor for Aflatoxin M1 Detection

被引:68
作者
Parker, Charlie O. [2 ]
Lanyon, Yvonne H. [1 ]
Manning, Mary [1 ]
Arrigan, Damien W. M. [1 ]
Tothill, Ibtisam E. [2 ]
机构
[1] Univ Coll Cork, Tyndall Natl Inst, Cork, Ireland
[2] Cranfield Univ, Cranfield Hlth, Cranfield MK43 0AL, Beds, England
基金
爱尔兰科学基金会;
关键词
LINEAR SWEEP VOLTAMMETRY; MICROELECTRODE ARRAYS; MICRODISK ELECTRODES; M1; DETERMINATION; GOLD ELECTRODES; IMMUNOSENSOR; MILK; BIOSENSORS; CARBON; B-1;
D O I
10.1021/ac900511e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An investigation into the fabrication, electrochemical characterization, and development of a microelectrode array (MEA) immunosensor for aflatoxin M-1 is presented in this paper. Gold MEAs (consisting of 35 microsquare electrodes with 20 mu m x 20 mu m dimensions and edge-to-edge spacing of 200 mu m) together with on-chip reference and counter electrodes were fabricated using standard photolithographic methods. The MEAs were then characterized by cyclic voltammetry, and the behavior of the on-chip electrodes were evaluated. The microarray sensors were assessed for their applicability to the development of an immunosensor for the analysis of aflatoxin M-1 directly in milk samples. Following the sensor surface silanization, antibodies were immobilized by cross-linking with 1,4-phenylene diisothiocyanate (PDITC). Surface characterization was conducted by electrochemistry, fluorescence microscopy, scanning electron microscopy (SEM), and atomic force microscopy (AFM). A competitive enzyme linked immunosorbent assay (ELISA) assay format was developed on the microarray electrode surface using the 3,3,5',5'-tetramethylbenzidine dihyrochloride (TMB)/H2O2 electrochemical detection scheme with horseradish peroxidase (HRP) as the enzyme label. The performance of the assay and the microarray sensor were characterized in pure buffer conditions before applying to the milk samples. With the use of this approach, the detection limit for aflatoxin M-1 in milk was estimated to be 8 ng L-1, with a dynamic detection range of 10-100 ng L-1, which meets present legislative limits of 50 ng L-1. The milk interference with the sensor surface was also found to be minimal. These devices show high potential for development of a range of new applications which have previously only been detected using elaborate instrumentation.
引用
收藏
页码:5291 / 5298
页数:8
相关论文
共 50 条
[1]   DIFFUSIONAL MASS-TRANSPORT TO MICROBAND ELECTRODES OF PRACTICAL GEOMETRIES - A SIMULATION STUDY USING THE STRONGLY IMPLICIT PROCEDURE [J].
ALDEN, JA ;
BOOTH, J ;
COMPTON, RG ;
DRYFE, RAW ;
SANDERS, GHW .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1995, 389 (1-2) :45-54
[2]  
Amatore C., 1995, PHYS ELECTROCHEMISTR, P131
[3]   Electrochemical immunosensor for determination of aflatoxin B1 in barley [J].
Ammida, NHS ;
Micheli, L ;
Palleschi, G .
ANALYTICA CHIMICA ACTA, 2004, 520 (1-2) :159-164
[4]   Aflatoxin M1 determination in raw milk using a flow-injection immunoassay system [J].
Badea, M ;
Micheli, L ;
Messia, MC ;
Candigliota, T ;
Marconi, E ;
Mottram, T ;
Velasco-Garcia, M ;
Moscone, D ;
Palleschi, G .
ANALYTICA CHIMICA ACTA, 2004, 520 (1-2) :141-148
[5]   Electrochemical sensors [J].
Bakker, Eric ;
Qin, Yu .
ANALYTICAL CHEMISTRY, 2006, 78 (12) :3965-3983
[6]  
BARD AJ, 2001, FUNDAMENTALS APPL, P168
[7]   Microelectrode arrays and microfabricated devices in electrochemical stripping analysis [J].
Beni, Valerio ;
Arrigan, Damien W. M. .
CURRENT ANALYTICAL CHEMISTRY, 2008, 4 (03) :229-241
[8]   Voltammetric characterisation of silicon-based microelectrode arrays and their application to mercury-free stripping voltammetry of copper ions [J].
Berduque, Alfonso ;
Lanyon, Yvonne H. ;
Beni, Valerio ;
Herzog, Gregoire ;
Watson, Yvonne E. ;
Rodgers, Kenneth ;
Stam, Frank ;
Alderman, John ;
Arrigan, Damien W. M. .
TALANTA, 2007, 71 (03) :1022-1030
[9]   Development of a disposable amperometric immunosensor for the detection of ecstasy and its analogues using screen-printed electrodes [J].
Butler, D ;
Pravda, M ;
Guilbault, GG .
ANALYTICA CHIMICA ACTA, 2006, 556 (02) :333-339
[10]   An automated, handheld biosensor for aflatoxin [J].
Carlson, MA ;
Bargeron, CB ;
Benson, RC ;
Fraser, AB ;
Phillips, TE ;
Velky, JT ;
Groopman, JD ;
Strickland, PT ;
Ko, HW .
BIOSENSORS & BIOELECTRONICS, 2000, 14 (10-11) :841-848