Characterization of a scintillating fiber-optic dosimeter for photon beam therapy

被引:12
作者
Jang, Kyoung Won [1 ]
Cho, Dong Hyun [1 ]
Shin, Sang Hun [1 ]
Yoo, Wook Jae [1 ]
Seo, Jeong Ki [1 ]
Lee, Bongsoo [1 ]
Kim, Sin [2 ]
Moon, Joo Hyun [3 ]
Cho, Young-Ho [4 ]
Park, Byung Gi [5 ]
机构
[1] Konkuk Univ, Sch Biomed Engn, Res Inst Biomed Engn, Chungju 380701, South Korea
[2] Jeju Natl Univ, Dept Nucl & Energy Engn, Appl Radiol Sci Res Inst, Cheju 690756, South Korea
[3] Dongguk Univ, Dept Energy & Environm Syst Engn, Gyeongju 780714, South Korea
[4] Catholic Univ, Dept Radiol Sci, Taegu 712702, South Korea
[5] Soonchunhyang Univ, Dept Energy & Environm Engn, Asan 336745, South Korea
关键词
organic scintillator; fiber-optic dosimeter; photon beam; radiotherapy; IONIZATION-CHAMBER; DOSE DISTRIBUTIONS; RADIATION; SENSOR;
D O I
10.1007/s10043-009-0072-x
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
A scintillating fiber-optic dosimeter has many advantages such as real-time readout, high-resolution measurement, water-equivalence and no corrections for temperature, pressure and humidity. Organic scintillator which has water or tissue equivalent characteristics is very important to measure absorbed dose, dose rate and dose distributions exactly without any corrections and due to its small size, the sensitive volume enables accurate dose measurements in regions of high dose gradients with high spatial resolution. In this study, a scintillating fiber-optic dosimeter with an organic scintillator is fabricated to measure high-energy photon beam from a clinical linear accelerator. And we have measured linear responses of a fiber-optic dosimeter according to dose rates and monitor units of a clinical linear accelerator. Also, a percent depth dose curve for 6MV photon beam with different field sizes are obtained.
引用
收藏
页码:383 / 386
页数:4
相关论文
共 16 条
[1]  
Aoyama T, 1999, PATRAS MEDICAL PHYSICS 99, P117
[2]   Water-equivalent dosimeter array for small-field external beam radiotherapy [J].
Archambault, Louis ;
Beddar, A. Sam ;
Gingras, Luc ;
Lacroix, Frederic ;
Roy, Rene ;
Beaulieu, Luc .
MEDICAL PHYSICS, 2007, 34 (05) :1583-1592
[3]   Plastic scintillation dosimetry and its application to radiotherapy [J].
Beddar, A. S. .
RADIATION MEASUREMENTS, 2006, 41 :S124-S133
[4]   SILICON DIODE DOSIMETRY [J].
DIXON, RL ;
EKSTRAND, KE .
INTERNATIONAL JOURNAL OF APPLIED RADIATION AND ISOTOPES, 1982, 33 (11) :1171-1176
[5]   First tests of a liquid ionization chamber to monitor intensity modulated radiation beams [J].
Eberle, K ;
Engler, J ;
Hartmann, G ;
Hofmann, R ;
Hörandel, JR .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (21) :3555-3564
[6]   Spectral discrimination of Cerenkov radiation in scintillating dosimeters [J].
Frelin, AM ;
Fontbonne, JM ;
Ban, G ;
Colin, J ;
Labalme, M ;
Batalla, A ;
Isambert, A ;
Vela, A ;
Leroux, T .
MEDICAL PHYSICS, 2005, 32 (09) :3000-3006
[7]  
HENDEE WR, 2005, RAD THERAPY PHYS, P93
[8]   Measurements of high energy X-ray dose distributions using multi-dimensional fiber-optic radiation detectors [J].
Jang, Kyoung Won ;
Cho, Dong Hyun ;
Shin, Sang Hun ;
Lee, Bongsoo ;
Chung, Soon-Cheol ;
Tack, Gye-Rae ;
Yi, Jeong Han ;
Kim, Sin ;
Cho, Hyosung .
MODERN PHYSICS LETTERS B, 2008, 22 (11) :797-802
[9]  
KHAN FM, 1994, PHYS RAD THERAPY, P183
[10]   Development of one-dimensional fiber-optic radiation sensor for measuring dose distributions of high energy photon beams [J].
Lee, Bongsoo ;
Jang, Kyoung Won ;
Cho, Dong Hyun ;
Yoo, Wook Jae ;
Kim, Hyung Shik ;
Chung, Soon-Cheol ;
Yi, Jeong Han .
OPTICAL REVIEW, 2007, 14 (05) :351-354