Quantification of mycophenolic acid in human plasma by liquid chromatography with time-of-flight mass spectrometry for therapeutic drug monitoring

被引:1
|
作者
Antunes, Natalicia J. [1 ,2 ,3 ]
Ince, Nina [3 ,4 ]
Raymond, Jason [3 ,4 ]
Kipper, Karin [3 ,5 ]
Couchman, Lewis [3 ,6 ]
Holt, David W. [3 ]
De Nucci, Gilberto [1 ]
Johnston, Atholl [2 ,3 ]
机构
[1] Univ Estadual Campinas, Dept Pharmacol, Fac Med Sci, Campinas, SP, Brazil
[2] Queen Mary Univ London, William Harvey Res Inst, Barts & London Sch Med & Dent, Clin Pharmacol, London, England
[3] St Georges Univ London, Analyt Serv Int Ltd, London, England
[4] Univ Surrey, Dept Chem, Guildford, Surrey, England
[5] Univ Tartu, Inst Chem, Tartu, Estonia
[6] Kings Coll London, Pharmaceut Sci Clin Acad Grp, London, England
基金
巴西圣保罗研究基金会;
关键词
immunosuppressant; therapeutic drug monitoring; LC– TOF– MS; mycophenolic acid; TRANSPLANT RECIPIENTS; CLINICAL PHARMACOKINETICS; GLUCURONIDE; IDENTIFICATION; MOFETIL;
D O I
10.1002/bmc.5011
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
This study presents, for the first time, the development and validation of a liquid chromatography and time-of-flight mass-spectrometry (LC-TOF-MS) based assay to quantify mycophenolic acid (MPA) in patient samples as part of a routine therapeutic drug monitoring service. MPA was extracted from 50 mu l human plasma by protein precipitation, using sulindac as internal standard (IS). Separation was obtained on a Luna (TM) Omega polar C-18 column kept at 40 degrees C. The mobile phase consisted of a mixture of acetonitrile-deionized water (50:50, v/v) with 0.1% formic acid at a flow rate of 350 mu l/min. Analyte and IS were monitored on a TOF-MS using a Jet-Stream (TM) (electrospray) interface running in positive mode. Assay performance was evaluated by analysing patient plasma (N = 69) and external quality assessment (N = 6) samples. The retention times were 2.66 and 2.18 min for MPA and IS, respectively. The lower limit of quantification of MPA was 0.1 mu g/ml. The within- and between-assay reproducibility results ranged from 1.81 to 10.72%. Patient and external quality assessment sample results were comparable with those obtained previously by an in-house validated LC-MS/MS method. This method showed satisfactory analytical performance for the determination of MPA in plasma over the calibration range of 0.1-15.0 mu g/ml.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Quantification of levetiracetam in human plasma by liquid chromatography-tandem mass spectrometry: Application to therapeutic drug monitoring
    Matar, Kamal M.
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2008, 48 (03) : 822 - 828
  • [2] Therapeutic drug monitoring of immunosuppressants by liquid chromatography-mass spectrometry
    McShane, Adam J.
    Bunch, Dustin R.
    Wang, Sihe
    CLINICA CHIMICA ACTA, 2016, 454 : 1 - 5
  • [3] Quantification of 11 Therapeutic Kinase Inhibitors in Human Plasma for Therapeutic Drug Monitoring Using Liquid Chromatography Coupled With Tandem Mass Spectrometry
    Herbrink, Maikel
    de Vries, Niels
    Rosing, Hilde
    Huitema, Alwin D. R.
    Nuijen, Bastiaan
    Schellens, Jan H. M.
    Beijnen, Jos H.
    THERAPEUTIC DRUG MONITORING, 2016, 38 (06) : 649 - 656
  • [4] Simultaneous quantification of olanzapine and its metabolite N-desmethylolanzapine in human plasma by liquid chromatography tandem mass spectrometry for therapeutic drug monitoring
    Lou, Hong-gang
    Ruan, Zou-rong
    Jiang, Bo
    Chen, Jin-liang
    BIOMEDICAL CHROMATOGRAPHY, 2015, 29 (05) : 671 - 678
  • [5] Simultaneous quantification of selective serotonin reuptake inhibitors and metabolites in human plasma by liquid chromatography-electrospray mass spectrometry for therapeutic drug monitoring
    Ansermot, Nicolas
    Brawand-Amey, Marlyse
    Eap, Chin B.
    JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2012, 885 : 117 - 130
  • [6] Free mycophenolic acid determination in human plasma ultrafiltrate by a validated liquid chromatography-tandem mass spectrometry method
    Luszczynska, Paulina
    Pawinski, Tomasz
    Kunicki, Pawel K.
    Sikorska, Katarzyna
    Marszalek, Ryszard
    BIOMEDICAL CHROMATOGRAPHY, 2017, 31 (10)
  • [7] Characterization of metabolites in rat plasma after intravenous administration of salvianolic acid A by liquid chromatography/time-of-flight mass spectrometry and liquid chromatography/ion trap mass spectrometry
    Shen, Yi
    Wang, Xueyan
    Xu, Lihua
    Liu, Xiaowei
    Chao, Ruobing
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2009, 23 (12) : 1810 - 1816
  • [8] Quantification of typical antipsychotics in human plasma by ultra-high performance liquid chromatography tandem mass spectrometry for therapeutic drug monitoring
    Gradinaru, Julieta
    Vullioud, Astrid
    Eap, Chin B.
    Ansermot, Nicolas
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2014, 88 : 36 - 44
  • [9] Determination of mycophenolic acid in human plasma by ultra performance liquid chromatography tandem mass spectrometry
    Vivek Upadhyay
    Vikas Trivedi
    Gaurang Shah
    Manish Yadav
    Pranav S.Shrivastav
    JournalofPharmaceuticalAnalysis, 2014, 4 (03) : 205 - 216
  • [10] Liquid chromatography/mass spectrometry for therapeutic drug monitoring of immunosuppressants
    Deters, M
    Kaever, V
    Kirchner, GI
    ANALYTICA CHIMICA ACTA, 2003, 492 (1-2) : 133 - 145