A new device for 100 per cent humidification of inspired air

被引:14
作者
Larsson, A
Gustafsson, A
Svanborg, L
机构
[1] Univ Copenhagen, Gentofte Univ Hosp, Dept Anaesthesiol, DK-2900 Hellerup, Denmark
[2] Univ Lund Hosp, Dept Anesthesiol & Intens Care, Intens Care Unit, CCRN, Lund, Sweden
[3] Louis Gibeck AB, Res & Dev, Upplands Vasby, Sweden
来源
CRITICAL CARE | 2000年 / 4卷 / 01期
关键词
airway humidification; heated humidifier; intensive care; mechanical ventilation;
D O I
10.1186/cc651
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
A new humidifier for use during mechanical ventilation in endotracheally intubated patients is described and tested. The humidifier is based on a heat-moisture exchanger, which absorbs the expired heat and moisture and releases it into the inspired air. External heat and water are then added at the patient side of the heat-moisture exchanger, so that the inspired gas should reach 100% humidity (44 mg/l) at 37 degrees C. In bench tests using constant and decelerating inspiratory flow and minute volumes of 3-25 l the device gave an absolute humidity of 41-44 mg/l, and it reduced the amount of water consumed in eight mechanically ventilated patients compared with a conventional active humidifier. During a 24-h test period there was no water condensation in the ventilator tubing with the new device.
引用
收藏
页码:54 / 60
页数:7
相关论文
共 14 条
[1]  
[Anonymous], RESP CARE
[2]   Clinical utility of hygroscopic heat and moisture exchangers in intensive care patients [J].
Boots, RJ ;
Howe, S ;
George, N ;
Harris, FM ;
Faoagali, J .
CRITICAL CARE MEDICINE, 1997, 25 (10) :1707-1712
[3]   HUMIDIFICATION IN ANESTHESIA - REVIEW OF PRESENT SITUATION [J].
BOYS, JE ;
HOWELLS, TH .
BRITISH JOURNAL OF ANAESTHESIA, 1972, 44 (08) :879-+
[4]   HUMIDIFICATION IN THE INTENSIVE-CARE UNIT - PROSPECTIVE-STUDY OF A NEW PROTOCOL UTILIZING HEATED HUMIDIFICATION AND A HYGROSCOPIC CONDENSER HUMIDIFIER [J].
BRANSON, RD ;
DAVIS, K ;
CAMPBELL, RS ;
JOHNSON, DJ ;
POREMBKA, DT .
CHEST, 1993, 104 (06) :1800-1805
[5]   EFFECTS OF DRY ANESTHETIC GASES ON TRACHEOBRONCHIAL CILIATED EPITHELIUM [J].
CHALON, J ;
MALEBRANCHE, J ;
LOEW, DAY .
ANESTHESIOLOGY, 1972, 37 (03) :338-+
[6]   Influence of airway management on ventilator-associated pneumonia - Evidence from randomized trials [J].
Cook, D ;
De Jonghe, B ;
Brochard, L ;
Brun-Buisson, C .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1998, 279 (10) :781-787
[7]  
CRAVEN DE, 1984, AM REV RESPIR DIS, V129, P625
[8]  
DREYFUSS D, 1995, AM J RESP CRIT CARE, V151, P986
[9]   A randomized clinical trial comparing an extended-use hygroscopic condenser humidifier with heated-water humidification in mechanically ventilated patients [J].
Kollef, MH ;
Shapiro, SD ;
Boyd, V ;
Silver, P ;
Von Harz, B ;
Trovillion, E ;
Prentice, D .
CHEST, 1998, 113 (03) :759-767
[10]   EFFECT OF DRY AND HUMIDIFIED GASES ON RESPIRATORY EPITHELIUM IN RABBITS [J].
MARFATIA, S ;
DONAHOE, PK ;
HENDREN, WH .
JOURNAL OF PEDIATRIC SURGERY, 1975, 10 (05) :583-592