A FPGA-Based Broadband EIT System for Complex Bioimpedance Measurements-Design and Performance Estimation

被引:27
|
作者
Kusche, Roman [1 ,2 ]
Malhotra, Ankit [1 ,3 ]
Ryschka, Martin [1 ]
Ardelt, Gunther [4 ]
Klimach, Paula [1 ,2 ]
Kaufmann, Steffen [1 ,3 ]
机构
[1] Lubeck Univ Appl Sci, Lab Med Elect, D-23562 Lubeck, Germany
[2] Univ Lubeck, Grad Sch Comp Med & Life Sci, D-23562 Lubeck, Germany
[3] Univ Lubeck, Inst Med Engn, D-23562 Lubeck, Germany
[4] Lubeck Univ Appl Sci, Ctr Excellence CoSA Commun Syst Applicat, D-23562 Lubeck, Germany
来源
ELECTRONICS | 2015年 / 4卷 / 03期
关键词
bioimpedance measurements; electrical impedance tomography; multi-frequency; chirp signal excitation; broadband impedance measurements; ELECTRICAL-IMPEDANCE TOMOGRAPHY; BREAST-CANCER DETECTION; SPECTROSCOPY; LUNG;
D O I
10.3390/electronics4030507
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Electrical impedance tomography (EIT) is an imaging method that is able to estimate the electrical conductivity distribution of living tissue. This work presents a field programmable gate array (FPGA)-based multi-frequency EIT system for complex, time-resolved bioimpedance measurements. The system has the capability to work with measurement setups with up to 16 current electrodes and 16 voltage electrodes. The excitation current has a range of about 10 mu A to 5 mA, whereas the sinusoidal signal used for excitation can have a frequency of up to 500 kHz. Additionally, the usage of a chirp or rectangular signal excitation is possible. Furthermore, the described system has a sample rate of up to 3480 impedance spectra per second (ISPS). The performance of the EIT system is demonstrated with a resistor-based phantom and tank phantoms. Additionally, first measurements taken from the human thorax during a breathing cycle are presented.
引用
收藏
页码:507 / 525
页数:19
相关论文
共 3 条
  • [1] An FPGA-Based Multifrequency EIT System With Reference Signal Measurement
    Xu, Yanbin
    Yan, Zhicheng
    Han, Bing
    Dong, Feng
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2021, 70
  • [2] A high accuracy broadband measurement system for time resolved complex bioimpedance measurements
    Kaufmann, S.
    Malhotra, A.
    Ardelt, G.
    Ryschka, M.
    PHYSIOLOGICAL MEASUREMENT, 2014, 35 (06) : 1163 - 1180
  • [3] A FPGA-Based Multi-Frequency Current Source for Biological EIT System
    Yang, Xianding
    Xu, Yanbin
    Dong, Feng
    2016 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE PROCEEDINGS, 2016, : 1195 - 1200