MW-PPG Sensor: An on-Chip Spectrometer Approach

被引:24
作者
Chang, Cheng-Chun [1 ]
Wu, Chien-Ta [1 ]
Choi, Byung Il [2 ]
Fang, Tong-Jing [3 ]
机构
[1] Natl Taipei Univ Technol, Dept Elect Engn, Taipei 10608, Taiwan
[2] NanoLambda Co Ltd, Daejeon 34141, South Korea
[3] Natl Def Med Ctr, Grad Inst Physiol, Dept Physiol & Biophys, Taipei 11490, Taiwan
关键词
multi-wavelengths; MW-PPG; on-chip spectrometers; PHOTOPLETHYSMOGRAPHY;
D O I
10.3390/s19173698
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Multi-wavelength photoplethysmography (MW-PPG) sensing technology has been known to be superior to signal-wavelength photoplethysmography (SW-PPG) sensing technology. However, limited by the availability of sensing detectors, many prior studies can only use conventional bulky and pricy spectrometers as the detectors, and hence cannot bring the MW-PPG technology to daily-life applications. In this study we developed a chip-scale MW-PPG sensor using innovative on-chip spectrometers, aimed at wearable applications. Also in this paper we present signal processing methods for robustly extracting the PPG signals, in which an increase of up to 50% in the signal-to-noise ratio (S/N) was observed. Example measurements of saturation of peripheral blood oxygen (SpO(2)) and blood pressure were conducted.
引用
收藏
页数:16
相关论文
共 29 条
[1]   Similarity in bilateral photoplethysmographic peripheral pulse wave characteristics at the ears, thumbs and toes [J].
Allen, J ;
Murray, A .
PHYSIOLOGICAL MEASUREMENT, 2000, 21 (03) :369-377
[2]   THE OPTICS OF HUMAN-SKIN [J].
ANDERSON, RR ;
PARRISH, JA .
JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1981, 77 (01) :13-19
[3]  
Asare L., 2011, P SOC PHOTO-OPT INS, P1
[4]  
Asare L., 2010, P SOC PHOTO-OPT INS
[5]  
Asare L., 2011, P SOC PHOTO-OPT INS
[6]   Different patterns of parasympathetic activation in uni- and bilateral migraineurs [J].
Avnon, Y ;
Nitzan, M ;
Sprecher, E ;
Rogowski, Z ;
Yarnitsky, D .
BRAIN, 2003, 126 :1660-1670
[7]   Riaht-left correlation of the sympathetically induced fluctuations of photoplethysmographic signal in diabetic and non-diabetic subjects [J].
Buchs, A ;
Slovik, Y ;
Rapoport, M ;
Rosenfeld, C ;
Khanokh, B ;
Nitzan, M .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2005, 43 (02) :252-257
[8]  
Carni D.L., 2016, P 2016 IEEE INT S ME
[9]   Spectrum Reconstruction for On-Chip Spectrum Sensor Array Using a Novel Blind Nonuniformity Correction Method [J].
Chang, Cheng-Chun ;
Lin, Hsuan-Yu .
IEEE SENSORS JOURNAL, 2012, 12 (08) :2586-2592
[10]  
Fung P., 2004, P 26 ANN INT C IEEE, P1