Influence of turbulent shear stresses on the numerical blood damage prediction in a ventricular assist device

被引:22
作者
Torner, Benjamin [1 ]
Konnigk, Lucas [1 ]
Wurm, Frank-Hendrik [1 ]
机构
[1] Univ Rostock, Inst Turbomachinery, Friedrich Barnewitz Str 2, D-18119 Rostock, Germany
关键词
Computational fluid dynamics; large eddy simulation; rotary blood pump; equivalent shear stress; turbulent stress; turbulence; dissipation; PUMP DESIGN; HEMOLYSIS; TRAUMA; MODELS;
D O I
10.1177/0391398819861395
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The blood damage prediction in rotary blood pumps is an important procedure to evaluate the hemocompatibility of such systems. Blood damage is caused by shear stresses to the blood cells and their exposure times. The total impact of an equivalent shear stress can only be taken into account when turbulent stresses are included in the blood damage prediction. The aim of this article was to analyze the influence of the turbulent stresses on the damage prediction in a rotary blood pump's flow. Therefore, the flow in a research blood pump was computed using large eddy simulations. A highly turbulence-resolving setup was used in order to directly resolve most of the computed stresses. The simulations were performed at the design point and an operation point with lower flow rate. Blood damage was predicted using three damage models (volumetric analysis of exceeded stress thresholds, hemolysis transport equation, and hemolysis approximation via volume integral) and two shear stress definitions (with and without turbulent stresses). For both simulations, turbulent stresses are the dominant stresses away from the walls. Here, they act in a range between 9 and 50 Pa. Nonetheless, the mean stresses in the proximity of the walls reach levels, which are one order of magnitude higher. Due to this, the turbulent stresses have a small impact on the results of the hemolysis prediction. Yet, turbulent stresses should be included in the damage prediction, since they belong to the total equivalent stress definition and could impact the damage on proteins or platelets.
引用
收藏
页码:735 / 747
页数:13
相关论文
共 44 条
[1]  
[Anonymous], 2013, 99062012 DIN EN ISO
[2]  
ANSYS Inc, CFX THEOR GUID
[3]   Pump thrombosis-A riddle wrapped in a mystery inside an enigma [J].
Blitz, Arie .
ANNALS OF CARDIOTHORACIC SURGERY, 2014, 3 (05) :450-471
[4]   3-DIMENSIONAL NUMERICAL PREDICTION OF STRESS LOADING OF BLOOD PARTICLES IN A CENTRIFUGAL PUMP [J].
BLUDSZUWEIT, C .
ARTIFICIAL ORGANS, 1995, 19 (07) :590-596
[5]  
Breuer M., 2001, THESIS
[6]  
Buja L, 2016, Cardiovascular pathology, V4th
[7]   Shear stress and blood trauma under constant and pulse-modulated speed CF-VAD operations: CFD analysis of the HVAD [J].
Chen, Zengsheng ;
Jena, Sofen K. ;
Giridharan, Guruprasad A. ;
Sobieski, Michael A. ;
Koenig, Steven C. ;
Slaughter, Mark S. ;
Griffith, Bartley P. ;
Wu, Zhongjun J. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2019, 57 (04) :807-818
[8]   Flow features and device-induced blood trauma in CF-VADs under a pulsatile blood flow condition: A CFD comparative study [J].
Chen, Zengsheng ;
Jena, Sofen K. ;
Giridharan, Guruprasad A. ;
Koenig, Steven C. ;
Slaughter, Mark S. ;
Griffith, Bartley P. ;
Wu, Zhongjun J. .
INTERNATIONAL JOURNAL FOR NUMERICAL METHODS IN BIOMEDICAL ENGINEERING, 2018, 34 (02)
[9]   Extending the Power-Law Hemolysis Model to Complex Flows [J].
Faghih, Mohammad M. ;
Sharp, M. Keith .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2016, 138 (12)
[10]   A Quantitative Comparison of Mechanical Blood Damage Parameters in Rotary Ventricular Assist Devices: Shear Stress, Exposure Time and Hemolysis Index [J].
Fraser, Katharine H. ;
Zhang, Tao ;
Taskin, M. Ertan ;
Griffith, Bartley P. ;
Wu, Zhongjun J. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2012, 134 (08)