A Fast List-Mode Reconstruction Algorithm with Dedicated Correction for Random Coincidences for the Clear-PEM System

被引:0
作者
Cao, Liji [1 ]
Bugalho, Ricardo [2 ]
Ferreira, Claudia S. [2 ,3 ]
Ortigao, Catarina [2 ]
Capote, Ricardo M. [3 ]
Varela, Joao [2 ]
Peter, Joerg [1 ]
机构
[1] German Canc Res Ctr, Div Med Phys Radiol, D-69120 Heidelberg, Germany
[2] LIP Lab Instrumentat & Expt Particles Phys, P-1000149 Lisbon, Portugal
[3] Univ Lisbon, Fac Sci, Inst Biophys & Biomed Engn, P-1149014 Lisbon, Portugal
关键词
Correction for random coincidences; image reconstruction; list mode; positron emission mammography (PEM); EMISSION MAMMOGRAPHY PEM; IMAGE-RECONSTRUCTION; EM ALGORITHM; MAXIMUM; SCATTER; PHANTOM; DESIGN;
D O I
10.1109/TNS.2014.2311171
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A fast list-mode MLEM reconstruction algorithm including low-noise random correction for the Clear-PEM imaging system is presented. In this approach, the sensitivity matrix is estimated based on planar phantom acquisitions, rather than tracing along lines of response (LORs). The algorithm is implemented by distributing the calculation of the acquired list-mode events into different threads yielding significantly increased computational speed. A list-mode random correction approach is presented in which reconstructed data are corrected according to a smooth correction image estimated from delayed coincidence data. Monte Carlo simulations based on GATE were performed to validate the proposed approach. A Derenzo phantom study shows that the 1.2-mm hot rods can be resolved using the presented algorithm. A breast-torso phantom study as well as a study based on acquired patient data demonstrate its feasibility in a clinical environment. This method reaches a significant higher calculation efficiency in which the total reconstruction time is less than 30 seconds for a typical clinical study. Results show that bias from random coincidences is largely suppressed. The presented algorithm achieves a fast and low-noise reconstructions without compromising intrinsic system resolution and sensitivity.
引用
收藏
页码:1182 / 1191
页数:10
相关论文
共 35 条
[31]  
Smith MF, 2006, IEEE NUCL SCI CONF R, P1792
[32]   POSITRON EMISSION MAMMOGRAPHY (PEM) - A PROMISING TECHNIQUE FOR DETECTING BREAST-CANCER [J].
THOMPSON, CJ ;
MURTHY, K ;
PICARD, Y ;
WEINBERG, IN ;
MAKO, R .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (04) :1012-1017
[33]  
Trindade A., 2007, THESIS U TECNICA LIS
[34]  
Yamada Y, 2007, IEEE NUCL SCI CONF R, P4397
[35]  
Yamakawa Y., 2008, NUCL SCI S 2008 NSS, P5129