A comparison of proximal femoral locking compression plates with dynamic hip screws in extracapsular femoral fractures

被引:29
|
作者
Zhong, B. [1 ]
Zhang, Y. [2 ]
Zhang, C. [1 ]
Luo, C. -F. [1 ]
机构
[1] Shanghai Jiao Tong Univ, Peoples Hosp 6, Dept Orthopaed, Shanghai, Peoples R China
[2] Shanghai Jiao Tong Univ, Tongren Hosp, Dept Orthopaed, Shanghai, Peoples R China
基金
中国国家自然科学基金;
关键词
Proximal femoral locking compression plate; Dynamic hip screw; Extracapsular femoral fracture; SUBTROCHANTERIC FRACTURES; PERTROCHANTERIC FRACTURES; FIXATION; LCP; FAILURE;
D O I
10.1016/j.otsr.2014.06.012
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: The potential value of proximal femoral locking compression plate (PFLCP) for extracapsular femoral fractures has been discussed in several case reports; however, clinical control studies are lacking. Therefore, we performed a case control study to assess: (1) If PFLCP offers better functional results and fewer complications than dynamic hip screws (DHS)? (2) Which kind of extracapsular femoral fractures would benefit from PFLCP fixation? Hypothesis: The PFLCP fixation offers better functional results and fewer complications than the DHS for the treatment of extracapsular fractures. Patients and methods: A total of 83 patients with extracapsular femoral fractures were recruited. Forty-one patients underwent PFLCP fixation, and 42 patients underwent DHS fixation. Patient information, operative time, blood loss, functional level (as assessed by Sanders' traumatic hip rating scale), bone union, and implant complications were compared for the two treatment groups. Results: Patients with stable intertrochanteric fractures who underwent PFLCP fixation demonstrated shorter bone union time than the DHS fixation group (3.3 +/- 0.2 vs. 4.3 +/- 0.1 month; P<0.0001); however, both groups had 100% bone union and good to excellent scores on Sanders' traumatic hip rating scale (P=1.000). Patients with unstable intertrochanteric fractures who underwent PFLCP fixation experienced greater blood loss (619.0 +/- 23.9 vs. 474.1 +/- 19.8 ml; P<0.0001), which was mainly due to the need for open reduction (64.3% vs. 12.5%; P=0.003),compared to the DHS fixation group. No differences were identified with respect to bony union, functional level, or complications. Patients with subtrochanteric fractures who underwent PFLCP fixation demonstrated significantly shorter operative times (82.1 +/- 4.3 vs. 102.2 +/- 2.2 minutes; P < 0.0001), less blood loss (751.8 +/- 25.4 vs. 987.6 +/- 32.0 ml; P<0.0001), shorter bone union times (5.2 +/- 0.4 vs. 8.8 +/- 1.0 month; P=0.006), more good to excellent Sanders' traumatic hip rating scale scores (92.9% vs. 55.5%; P= 0.009), and fewer complications (14.2% vs. 66.6%; P=0.005) than the DHS fixation group. Conclusion: PFLCP fixation offers better functional outcomes and fewer complications for subtrochanteric femoral fractures but not for intertrochanteric femoral fractures. Levels of evidence: Case control study, level III. (C) 2014 Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:663 / 668
页数:6
相关论文
共 50 条
  • [31] Minimally invasive technique versus conventional technique of dynamic hip screws for intertrochanteric femoral fractures
    Wang, Jin-Ping
    Yang, Tian-Fu
    Kong, Qing-Quan
    Liu, Shao-Jiang
    Xiao, Heng
    Liu, Yang
    Zhang, Hui
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2010, 130 (05) : 613 - 620
  • [32] Proximal femoral fractures
    Klopfer, T.
    Hemmann, P.
    Schreiner, A. J.
    Bahrs, C.
    TRAUMA UND BERUFSKRANKHEIT, 2019, 21 (02) : 86 - 94
  • [33] Proximal femoral nail antirotation (PFNA) for treatment of osteoporotic proximal femoral fractures
    Raviraj, A.
    Anand, Ashish
    Chakravarthy, Murali
    Pai, Sanjay
    EUROPEAN JOURNAL OF ORTHOPAEDIC SURGERY AND TRAUMATOLOGY, 2012, 22 (04) : 301 - 305
  • [34] Locking Compression Plates for the Treatment of Periprosthetic Femoral Fractures Around Well-Fixed Total Hip and Knee Implants
    Wood, Gavin C. A.
    Naudie, Doug R.
    McAuley, James
    McCalden, Richard W.
    JOURNAL OF ARTHROPLASTY, 2011, 26 (06) : 886 - 892
  • [35] Distal Locking of Proximal Femoral Nails in Intertrochanteric Fractures: A Comparison of Taloned and Static Locking Systems
    Karagoz, B.
    Keceli, O.
    Agir, I
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2024, 27 (05) : 620 - 627
  • [36] Expandable Proximal Femoral Nail versus Gamma Proximal Femoral Nail for the treatment of hip reverse oblique fractures
    Warschawski, Yaniv
    Ankori, Ran
    Rutenberg, Tal Frenkel
    Steinberg, Ely L.
    Atzmon, Ran
    Drexler, Michael
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2022, 142 (05) : 777 - 785
  • [37] Is new always better: comparison of the femoral neck system and the dynamic hip screw in the treatment of femoral neck fractures
    Schuetze, Konrad
    Burkhardt, Jakob
    Pankratz, Carlos
    Eickhoff, Alexander
    Boehringer, Alexander
    Degenhart, Christina
    Gebhard, Florian
    Cintean, Raffael
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (06) : 3155 - 3161
  • [38] Comparison of dynamic compression system versus multiple cancellous screws in the treatment of femoral neck fractures in young adults
    Aljasim, Omar
    Yener, Can
    Ozkayin, Nadir
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01):
  • [39] Biomechanical comparison of the femoral neck system and the dynamic hip screw in basicervical femoral neck fractures
    Moon, Jun-Ki
    Lee, Jung Il
    Hwang, Kyu-Tae
    Yang, Jae-Hyuk
    Park, Ye-Soo
    Park, Ki-Chul
    SCIENTIFIC REPORTS, 2022, 12 (01)
  • [40] Computational modeling in the prediction of Dynamic Hip Screw failure in proximal femoral fractures
    Hrubina, Maros
    Horak, Zdenek
    Bartoska, Radek
    Navratil, Leos
    Rosina, Jozef
    JOURNAL OF APPLIED BIOMEDICINE, 2013, 11 (03) : 143 - 151