Diffuse reflectance measurement using gas absorption spectroscopy

被引:4
作者
Gao, Qiang [1 ]
Yu, Jia [1 ]
Zhang, Yungang [1 ]
Zhang, Zhiguo [1 ]
Cao, Wenwu [1 ,2 ]
机构
[1] Harbin Inst Technol, Condensed Matter Sci & Technol Inst, Harbin 150080, Peoples R China
[2] Penn State Univ, Mat Res Inst, University Pk, PA 16802 USA
基金
中国博士后科学基金;
关键词
Diffuse reflectance; Diode laser; Gas absorption spectroscopy; Integrating sphere; INTEGRATING-SPHERE;
D O I
10.1016/j.snb.2014.02.010
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A precise method for the measurement of diffuse reflectance by using gas absorption spectroscopy technique with an integrating sphere is demonstrated. A quantitative relationship between the diffuse reflectance and the gas absorption spectrum is formulated, which has been further validated by experiments. The precision of the reflectivity measurement depends on the diameter and port fraction of the integrating sphere as well as gas concentration, and it increases linearly with the magnitude of the reflectivity. A high precision of 0.005% was achieved at the reflectivity of 0.98844(5). (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:147 / 150
页数:4
相关论文
共 50 条
[21]   Assessing human skin with diffuse reflectance spectroscopy and colorimetry [J].
Seo, InSeok ;
Liu, Yang ;
Bargo, Paulo R. ;
Kollias, Nikiforos .
PHOTONIC THERAPEUTICS AND DIAGNOSTICS VIII, PTS 1 AND 2, 2012, 8207
[22]   Characterization and Optical Early Diagnosis by Diffuse Reflectance Spectroscopy [J].
Yable, Didier K. ;
Canale, Laurent ;
Dupuis, Pascal ;
Haba, Theodore Cisse ;
Zoueu, Jeremie T. ;
Zissis, Georges .
2020 20TH IEEE INTERNATIONAL CONFERENCE ON ENVIRONMENT AND ELECTRICAL ENGINEERING AND 2020 4TH IEEE INDUSTRIAL AND COMMERCIAL POWER SYSTEMS EUROPE (EEEIC/I&CPS EUROPE), 2020,
[23]   Determination of tissue oxygen saturation by diffuse reflectance spectroscopy [J].
Sanchez-Ramos, Laura Lucia ;
Morales-Cruzado, Beatriz ;
Perez-Gutierrez, Francisco Gerardo .
JOURNAL OF BIOMEDICAL OPTICS, 2023, 28 (09)
[24]   Determination of Moisture Content in 5-Fluorouracil using Diffuse Reflectance Infrared Spectroscopy [J].
Singh, Parul ;
Jangir, Deepak Kumar ;
Mehrotra, Ranjana ;
Kandpal, H. C. .
PERSPECTIVES IN VIBRATIONAL SPECTROSCOPY, 2008, 1075 :172-174
[25]   Feasibility of quantitative diffuse reflectance spectroscopy for targeted measurement of renal ischemia during laparoscopic partial nephrectomy [J].
Goel, Utsav O. ;
Maddox, Michael M. ;
Elfer, Katherine N. ;
Dorsey, Philip J. ;
Wang, Mei ;
McCaslin, Ian Ross ;
Brown, J. Quincy ;
Lee, Benjamin R. .
JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (10)
[26]   In vivo assessment of liver fibrosis using diffuse reflectance and fluorescence spectroscopy: A proof of concept [J].
Fabila, Diego ;
Manuel de la Rosa, Jose ;
Stolik, Suren ;
Moreno, Edgard ;
Suarez-Alvarez, Karina ;
Lopez-Navarrete, Giuliana ;
Guzman, Carolina ;
Aguirre-Garcia, Jesus ;
Acevedo-Garcia, Christian ;
Kershenobich, David ;
Escobedo, Galileo .
PHOTODIAGNOSIS AND PHOTODYNAMIC THERAPY, 2012, 9 (04) :376-382
[27]   Thermal stability studies of 5-fluorouracil using diffuse reflectance infrared spectroscopy [J].
Singh, Parul ;
Tyagi, Gunjan ;
Mehrotra, Ranjana ;
Bakhshi, A. K. .
DRUG TESTING AND ANALYSIS, 2009, 1 (5-6) :240-244
[28]   Inexpensive diffuse reflectance spectroscopy system for measuring changes in tissue optical properties [J].
Glennie, Diana L. ;
Hayward, Joseph E. ;
McKee, Daniel E. ;
Farrell, Thomas J. .
JOURNAL OF BIOMEDICAL OPTICS, 2014, 19 (10)
[29]   Infrared diffuse reflectance instrumentation and standards at NIST [J].
Hanssen, LM ;
Kaplan, S .
ANALYTICA CHIMICA ACTA, 1999, 380 (2-3) :289-302
[30]   The effect of instrument design on diffuse reflectance measurements [J].
White, H ;
Taylor, J .
AIC: 9TH CONGRESS OF THE INTERNATIONAL COLOUR ASSOCIATION, 2002, 4421 :761-764