Computational modelling of flow through prosthetic heart valves using the entropic lattice-Boltzmann method

被引:39
作者
Yun, B. Min [1 ]
Dasi, L. P. [2 ]
Aidun, C. K. [1 ,3 ]
Yoganathan, A. P. [1 ,4 ,5 ]
机构
[1] Georgia Inst Technol, GW Woodruff Sch Mech Engn, Atlanta, GA 30318 USA
[2] Colorado State Univ, Dept Mech Engn, Ft Collins, CO 80523 USA
[3] Georgia Inst Technol, Parker H Petit Inst Bioengn & Biosci, Atlanta, GA 30332 USA
[4] Georgia Inst Technol, Wallace H Coulter Dept Biomed Engn, Atlanta, GA 30332 USA
[5] Emory Univ, Atlanta, GA 30332 USA
关键词
biomedical flows; computational methods; vortex dynamics; PARTICLE IMAGE VELOCIMETRY; BLOOD DAMAGE; HINGE; SIMULATIONS; TURBULENCE; PERFORMANCE; FIELDS; FLUID;
D O I
10.1017/jfm.2014.54
中图分类号
O3 [力学];
学科分类号
08 ; 0801 ;
摘要
Previous clinical, in vitro experimental and in silico simulation studies have shown the complex dynamics of flow through prosthetic heart valves. In the case of bileaflet mechanical heart valves (BMHVs), complex flow phenomena are observed due to the presence of two rigid leaflets. A numerical method for this type of study must be able to accurately simulate pulsatile flow through BMHVs with the inclusion of leaflet motion and high-Reynolds-number flow modelling. Consequently, this study aims at validating a numerical method that captures the flow dynamics for pulsatile flow through a BMHV. A 23 mm St. Jude Medical (SJM) Regent (TM) valve is selected for use in both the experiments and numerical simulations. The entropic lattice-Boltzmann method is used to simulate pulsatile flow through the valve with the inclusion of reverse leakage flow, while prescribing the flowrate and leaflet motion from experimental data. The numerical simulations are compared against experimental digital particle image velocimetry (DPIV) results from a previous study for validation. The numerical method is shown to match well with the experimental results quantitatively as well as qualitatively. Simulations are performed with efficient parallel processing at very high spatiotemporal resolution that can capture the finest details in the pulsatile BMHV flow field. This study validates the lattice-Boltzmann method as suitable for simulating pulsatile, high-Reynolds-number flows through prosthetic devices for use in future research.
引用
收藏
页码:170 / 201
页数:32
相关论文
共 31 条
[1]   Direct analysis of particulate suspensions with inertia using the discrete Boltzmann equation [J].
Aidun, CK ;
Lu, YN ;
Ding, EJ .
JOURNAL OF FLUID MECHANICS, 1998, 373 :287-311
[2]   LATTICE BOLTZMANN SIMULATION OF SOLID PARTICLES SUSPENDED IN FLUID [J].
AIDUN, CK ;
LU, YN .
JOURNAL OF STATISTICAL PHYSICS, 1995, 81 (1-2) :49-61
[3]   Lattice-Boltzmann Method for Complex Flows [J].
Aidun, Cyrus K. ;
Clausen, Jonathan R. .
ANNUAL REVIEW OF FLUID MECHANICS, 2010, 42 :439-472
[4]  
Black M M, 1994, Curr Top Pathol, V86, P127
[5]   Curvilinear immersed boundary method for simulating fluid structure interaction with complex 3D rigid bodies [J].
Borazjani, Iman ;
Ge, Liang ;
Sotiropoulos, Fotis .
JOURNAL OF COMPUTATIONAL PHYSICS, 2008, 227 (16) :7587-7620
[6]   The Effect of Implantation Orientation of a Bileaflet Mechanical Heart Valve on Kinematics and Hemodynamics in an Anatomic Aorta [J].
Borazjani, Iman ;
Sotiropoulos, Fotis .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2010, 132 (11)
[7]   Lattice Boltzmann method for direct numerical simulation of turbulent flows [J].
Chikatamarla, S. S. ;
Frouzakis, C. E. ;
Karlin, I. V. ;
Tomboulides, A. G. ;
Boulouchos, K. B. .
JOURNAL OF FLUID MECHANICS, 2010, 656 :298-308
[8]   Parallel performance of a lattice-Boltzmann/finite element cellular blood flow solver on the IBM Blue Gene/P architecture [J].
Clausen, Jonathan R. ;
Reasor, Daniel A., Jr. ;
Aidun, Cyrus K. .
COMPUTER PHYSICS COMMUNICATIONS, 2010, 181 (06) :1013-1020
[9]   Vorticity dynamics of a bileaflet mechanical heart valve in an axisymmetric aorta [J].
Dasi, L. P. ;
Ge, L. ;
Simon, H. A. ;
Sotiropoulos, F. ;
Yoganathan, A. P. .
PHYSICS OF FLUIDS, 2007, 19 (06)
[10]   Direct numerical simulation of the pulsatile flow through an aortic bileaflet mechanical heart valve [J].
De Tullio, M. D. ;
Cristallo, A. ;
Balaras, E. ;
Verzicco, R. .
JOURNAL OF FLUID MECHANICS, 2009, 622 :259-290