Capacitor Distortion in Magnetic Particle Imaging

被引:0
作者
Zheng, Bo [1 ]
Goodwill, Patrick [1 ]
Yang, Wisely [1 ]
Conolly, Steven [1 ]
机构
[1] Univ Calif Berkeley, Dept Bioengn, 340 Hearst Mem Min Bldg, Berkeley, CA 94720 USA
来源
MAGNETIC PARTICLE IMAGING: A NOVEL SPIO NANOPARTICLE IMAGING TECHNIQUE, 1ST EDITION | 2012年 / 140卷
基金
美国国家科学基金会;
关键词
D O I
暂无
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The signal-to-noise ratio in magnetic particle imaging can be limited by distortion interference arising in the imaging system. In this work, we investigate the contribution of resonant transmit capacitors to system interference. Feedthrough interference spectra obtained using four capacitors with varying voltage ratings in a custom MPI interference testbed show a 20dB reduction in distortion interference with higher-rated capacitors. Finally, we discuss the applicability of the interference testbed to treat other interference mechanisms in magnetic particle imaging.