Central and Midperipheral Corneal Thickness Measured with Scheimpflug Imaging and Optical Coherence Tomography

被引:13
作者
Huang, Jinhai [1 ,2 ,3 ]
Ding, Xixia [1 ,2 ,3 ]
Savini, Giacomo [1 ,2 ,4 ]
Jiang, Zhengxuan [5 ]
Pan, Chao [1 ,2 ]
Hua, Yanjun [1 ,2 ]
Wu, Fan [1 ,2 ]
Feng, Yifan [1 ,2 ]
Yu, Ye [1 ,2 ]
Wang, Qinmei [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Sch Optometry & Ophthalmol, Wenzhou, Zhejiang, Peoples R China
[2] Wenzhou Med Univ, Hosp Eye, Wenzhou, Zhejiang, Peoples R China
[3] Minist Hlth PR China, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[4] Studio Oculist DAzeglio, Bologna, Italy
[5] Anhui Med Univ, Affiliated Hosp 2, Hefei, Anhui, Peoples R China
关键词
ANTERIOR SEGMENT MEASUREMENTS; PACHYMETRY MEASUREMENTS; ULTRASONIC PACHYMETRY; DIURNAL-VARIATION; CONTACT-LENS; REPEATABILITY; REPRODUCIBILITY; AGREEMENT; GALILEI; INTRAOBSERVER;
D O I
10.1371/journal.pone.0098316
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose: To compare corneal thickness measurements using Pentacam (Oculus, Germany), Sirius (CSO, Italy), Galilei (Ziemer, Switzerland), and RTVue-100 OCT (Optovue Inc., USA). Methods: Sixty-six eyes of 66 healthy volunteers were enrolled. Three consecutive measurements were performed with each device. The mean value of the three measurements was used for subsequent analysis. Central corneal thickness (CCT), thinnest corneal thickness (TCT), and midperipheral corneal thickness (MPCT; measured at superior, inferior, nasal, and temporal locations with a distance of 1 mm (CT2mm) or 2.5 mm (CT5mm) from the corneal apex) were analyzed. Differences and agreement between measurements were assessed using the repeated-measures analysis of variance (ANOVA) and Bland-Altman analyses, respectively. Results: Statistically significant differences (p<0.001) among the four devices were revealed in CCT, TCT and CT2mm measurements. The CCT, TCT, and CT2mm values were ranked from the thickest to the thinnest as follows: Galilei>Sirius>Pentacam>RTVue OCT. For these measurements, agreement between measurements by Sirius and Pentacam was good, whereas Galilei overestimated and RTVue underestimated corneal thickness compared to Sirius and Pentacam. As regards CT5mm measurements, Pentacam provided the largest values, whereas RTVue OCT yielded the smallest values. Agreement of the CT5mm measurements was good between the Pentacam, Sirius, moderate between Galilei and the other two Scheimpflug systems, and poor between the RTVue OCT and the remaining devices. Conclusions: The Pentacam and Sirius can be used interchangeably for CCT measurements, while the Galilei and RTVue systematically over- and underestimate CCT, respectively. The three Scheimpflug cameras, but not the RTVue, may be used interchangeably for MPCT measurements.
引用
收藏
页数:7
相关论文
共 34 条
[1]   Repeatability and agreement of 2 Scheimpflug analyzers in measuring the central corneal thickness and anterior chamber angle, volume, and depth [J].
Bedei, Andrea ;
Appolloni, Ilaria ;
Madesani, Antonio ;
Pietrelli, Alessia ;
Franceschi, Sara ;
Barabesi, Lucio .
EUROPEAN JOURNAL OF OPHTHALMOLOGY, 2012, 22 :S29-S32
[2]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[3]   Central and Peripheral Corneal Thinning in Keratoconus [J].
Brautaset, Rune L. ;
Nilsson, Maria ;
Miller, William L. ;
Leach, Norman E. ;
Tukler, Johanna H. ;
Bergmanson, Jan P. G. .
CORNEA, 2013, 32 (03) :257-261
[4]   Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry [J].
Chen, Shihao ;
Huang, Jinhai ;
Wen, Daizong ;
Chen, Wuhe ;
Huang, David ;
Wang, Qinmei .
ACTA OPHTHALMOLOGICA, 2012, 90 (05) :449-455
[5]   Central and peripheral corneal thickness measured with the TOPCON specular microscope SP-2000P [J].
Cho, P ;
Cheung, SW .
CURRENT EYE RESEARCH, 2000, 21 (04) :799-807
[6]   Corneal thickness: measurement and implications [J].
Ehlers, N ;
Hjortdal, J .
EXPERIMENTAL EYE RESEARCH, 2004, 78 (03) :543-548
[7]  
Gordon MO, 2002, ARCH OPHTHALMOL-CHIC, V120, P714
[8]   Comparison of Corneal Thickness Measurements Using Galilei, HR Pentacam, and Ultrasound [J].
Hosseini, Hamid Reza Jahadi ;
Katbab, Asadollah ;
Khalili, Mohammad Reza ;
Abtahi, Mohammad Bagher .
CORNEA, 2010, 29 (10) :1091-1095
[9]   A Comparison between Scheimpflug Imaging and Optical Coherence Tomography in Measuring Corneal Thickness [J].
Huang, Jinhai ;
Ding, Xixia ;
Savini, Giacomo ;
Pan, Chao ;
Feng, Yifan ;
Cheng, Dan ;
Hua, Yanjun ;
Hu, Xiuli ;
Wang, Qinmei .
OPHTHALMOLOGY, 2013, 120 (10) :1951-1958
[10]   A Comprehensive Comparison of Central Corneal Thickness Measurement [J].
Huang, Jinhai ;
Pesudovs, Konrad ;
Yu, Ayong ;
Wright, Thomas ;
Wen, Daizong ;
Li, Ming ;
Yu, Ye ;
Wang, Qinmei .
OPTOMETRY AND VISION SCIENCE, 2011, 88 (08) :940-949