Exoseal for puncture site closure after antegrade procedures in peripheral arterial disease patients

被引:16
作者
Hackl, Gerald [1 ]
Gary, Thomas [1 ]
Belaj, Klara [1 ]
Hafner, Franz [1 ]
Rief, Peter [1 ]
Deutschmann, Hannes [2 ]
Brodmann, Marianne [1 ]
机构
[1] Med Univ Graz, Dept Internal Med, Div Angiol, Styria, Austria
[2] Med Univ Graz, Dept Radiol, Div Intervent Radiol, Styria, Austria
关键词
GUIDED THROMBIN INJECTION; CRITICAL LIMB ISCHEMIA; VASCULAR CLOSURE; ENDOVASCULAR INTERVENTIONS; DEVICE; ANGIOSEAL; EFFICACY; SAFETY; COMPLICATIONS; HEMOSTASIS;
D O I
10.5152/dir.2014.14002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE Exoseal is a vascular closure device consisting of a plug applier and a bio-absorbent polyglycolic acid plug available in sizes 5 F, 6 F, and 7 F. In this study, we aimed to evaluate the effectiveness and safety of the Exoseal vascular closure device (Cordis Corporation, Bridgewater, New Jersey, USA) for puncture site closure after antegrade endovascular procedures in peripheral arterial occlusive disease (PAOD) patients. MATERIALS AND METHODS In this retrospective study, a total of 168 consecutive patients who underwent an interventional procedure due to PAOD, were included. In each case, an antegrade peripheral endovascular procedure was performed via the common femoral artery using the Seldinger technique, and Exoseal 5 F, 6 F, or 7 F was used for access site closure. The primary endpoint was a technically successful application of Exoseal. All complications at the access site within 24 hours were registered as a secondary endpoint. RESULTS In a group of 168 patients (64.9% men, average age 71.9 +/- 11.9 years), the technical application of Exoseal was successful in 166 patients (98.8%). Within the first 24 hours after the procedure, 12 complications (7.2%) were recorded including, three pseudoaneurysms (1.8%) and nine hematomas (5.4%). None of the complications required surgical intervention. CONCLUSION Exoseal is a safe and effective device with high technical success and acceptable complication rates for access site closure after antegrade peripheral endovascular procedures.
引用
收藏
页码:426 / 431
页数:6
相关论文
共 27 条
[1]   Meta-analysis of randomized trials on the efficacy of vascular closure devices after diagnostic angiography and angioplasty [J].
Biancari, Fausto ;
D'Andrea, Vito ;
Di Marco, Carlo ;
Savino, Grazia ;
Tiozzo, Valentina ;
Catania, Antonio .
AMERICAN HEART JOURNAL, 2010, 159 (04) :518-531
[2]   Angioseal use after antegrade femoral arteriotomy in patients undergoing percutaneous revascularization for critical limb ischemia: A case series [J].
Biondi-Zoccai, Giuseppe G. L. ;
Fusaro, Massimiliano ;
Tashani, Abdulkafi ;
Mollichelli, Nadia ;
Medda, Massimo ;
De Giacobbi, Graziella ;
Inglese, Luigi .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2007, 118 (03) :398-399
[3]   Retrospective Evaluation of Safety and Effectiveness of the EXOSEAL Vascular Closure Device for Single Vascular Closure and Closure after Repeat Puncture in Diagnostic and Interventional Radiology: Single-center Experience [J].
Boschewitz, Jack M. ;
Andersson, Magnus ;
Naehle, Claas P. ;
Schild, Hans H. ;
Wilhelm, Kai ;
Meyer, Carsten .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (05) :698-702
[4]   Safety and Efficacy of Periprocedural Anticoagulation With Enoxaparin in Patients Undergoing Peripheral Endovascular Revascularization [J].
Brodmann, Marianne ;
Dorr, A. ;
Hafner, F. ;
Gary, T. ;
Froehlich, H. ;
Kvas, E. ;
Deutschmann, H. ;
Pilger, E. .
CLINICAL AND APPLIED THROMBOSIS-HEMOSTASIS, 2014, 20 (05) :530-535
[5]   Endovascular and interventional procedures in obese patients: A review of procedural technique modifications and radiation management [J].
Bryk, SG ;
Censullo, ML ;
Wagner, LK ;
Rossman, LL ;
Cohen, AM .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2006, 17 (01) :27-33
[6]   Comparison of arterial closure devices in antegrade and retrograde punctures [J].
Chiu, Albert Ho Yuen ;
Wal, Ross Vander ;
Tee, Kenneth ;
Knight, Roslyn ;
Coles, Simon Richard ;
Nadkarni, Sanjay .
JOURNAL OF ENDOVASCULAR THERAPY, 2008, 15 (03) :315-321
[7]   Comparison of enoxaparin and unfractionated heparin in endovascular interventions for the treatment of peripheral arterial occlusive disease: a randomized controlled trial [J].
Duschek, N. ;
Vafaie, M. ;
Skrinjar, E. ;
Hirsch, K. ;
Waldhoer, T. ;
Huebl, W. ;
Bergmayr, W. ;
Knoebl, P. ;
Assadian, A. .
JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2011, 9 (11) :2159-2167
[8]   Treatment of iatrogenic false aneurysms [J].
Franklin, JA ;
Brigham, D ;
Bogey, WM ;
Powell, CS .
JOURNAL OF THE AMERICAN COLLEGE OF SURGEONS, 2003, 197 (02) :293-301
[9]   Arterial Thrombosis After Using Angio-Seal [J].
Georg, Yannick ;
Thaveau, Fabien ;
Lejay, Anne ;
Bajcz, Carine ;
Bakassa, Serge ;
Chakfe, Nabil ;
Kretz, Jean-Georges .
ANNALS OF VASCULAR SURGERY, 2011, 25 (08) :1078-1093
[10]  
Hussain T, 2011, Ann R Coll Surg Engl, V93, pe102, DOI 10.1308/147870811X591143