Reproducibility of Scheimpflug Tomography Measurements Regarding Corneal Front and Back Surface Power

被引:4
作者
Stavridis, E. [1 ]
Eppig, T. [2 ]
Szentmary, N. [1 ,3 ]
Seitz, B. [1 ]
Langenbucher, A. [2 ]
机构
[1] Univ Klinikum Saarlandes, Klin Augenheilkunde, Homburg, Germany
[2] Univ Saarland, Inst Expt Ophthalmol, D-66424 Homburg, Germany
[3] Semmelweis Univ, Klin Augenheilkunde, H-1085 Budapest, Hungary
关键词
cornea; physiological optics; refractive surgery; PENETRATING KERATOPLASTY; TOPOGRAPHY SYSTEMS; REPEATABILITY; KERATOCONUS; AGREEMENT; PHOTOGRAPHY; DIAGNOSTICS; ASTIGMATISM; BIOMETRY; PENTACAM;
D O I
10.1055/s-0035-1558092
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background and Purpose: The Oculus Pentacam (R) is one of the most commonly used devices in ophthalmology for assessment of the anterior eye segment. The purpose of this study was to determine the reproducibility of the corneal power as measured by the Pentacam (R) in a normal population. Methods: We enrolled 25 eyes of 25 subjects aged between 13 and 68 (46.7 +/- 21.7) years, within a spherical equivalent from - 4 D to + 4 D and refractive cylinder up to 5 D and without notable pathologies or history of surgery. A sequence of 5 measurements was performed using the Pentacam (R), the Zeiss IOLMaster (R) as well as an autorefractometer after (re-)positioning the patient's head and the measurement device. From the Pentacam (R) we collected power data for the corneal front and back surfaces, apical corneal thickness at apex and internal anterior chamber depth. From the IOLMaster (R), we extracted corneal power data, and from the autorefractometer we obtained refractive power data of both cardinal meridians. For statistical analysis, we used Cronbach's a as a measure of reproducibility and Spearman's rank correlation test for correlation of data. Results: Pentacam (R) yields highly reproducible power data for both corneal surfaces, with Cronbach's a of greater than 0.97 for primary parameters (e.g. radii of curvature) and 0.9 for secondary parameters (e.g. vector components of astigmatism). Central corneal thickness and eccentricities of both surfaces as well as anterior chamber depth yielded highly reproducible values, with alpha of >= 0.97. Corneal power derived from IOLMaster (R) and objective refraction extracted from the autorefractometer yielded a values of around 0.9. Mean corneal power of Pentacam (R) and IOLMaster (R) correlated well, but there was no correlation to spherical equivalent. Astigmatism values from Pentacam (R) and IOLMaster (R) correlated well with each other and with refractive cylinder. Conclusion: Our results indicate that measurements of anterior segment using the Pentacam (R) yield highly reproducible results in a normal population without major pathologies of the anterior eye segment.
引用
收藏
页码:1297 / 1303
页数:7
相关论文
共 29 条
[1]   Astigmatism analysis by the Alpins method [J].
Alpins, N .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2001, 27 (01) :31-49
[2]   Scheimpflug and topography systems in ophthalmologic diagnostics [J].
Auffarth G.U. ;
Borkensein A.F.M. ;
Ehmer A. ;
Mannsfeld A. ;
Rabsilber T.M. ;
Holzer M.P. .
Der Ophthalmologe, 2008, 105 (9) :810-817
[3]   Cronbach's alpha [J].
Bland, JM ;
Altman, DG .
BRITISH MEDICAL JOURNAL, 1997, 314 (7080) :572-572
[4]   Corneal topography and keratoconus diagnostics with Scheimpflug photography [J].
Buehren, J. .
OPHTHALMOLOGE, 2014, 111 (10) :920-926
[5]   Reliability and repeatability of the Pentacam on corneal curvatures [J].
Chen, Davie ;
Lam, Andrew K. C. .
CLINICAL AND EXPERIMENTAL OPTOMETRY, 2009, 92 (02) :110-118
[6]   WHAT IS COEFFICIENT ALPHA - AN EXAMINATION OF THEORY AND APPLICATIONS [J].
CORTINA, JM .
JOURNAL OF APPLIED PSYCHOLOGY, 1993, 78 (01) :98-104
[7]  
Cronbach LJ, 1951, PSYCHOMETRIKA, V16, P297
[8]   Detection of Early Forms of Keratoconus - Current Screening Methods [J].
Goebels, S. ;
Eppig, T. ;
Seitz, B. ;
Langenbucher, A. .
KLINISCHE MONATSBLATTER FUR AUGENHEILKUNDE, 2013, 230 (10) :998-1004
[9]   Reproducibility of the Optical Biometer OA-1000 (Tomey) [J].
Goebels, Susanne Christiane ;
Seitz, Berthold ;
Langenbucher, Achim .
BIOMED RESEARCH INTERNATIONAL, 2014, 2014
[10]  
Handzel DM, 2014, OPHTHALMOLOGE, V111, P927, DOI 10.1007/s00347-013-2961-4