Hydrophilic interaction liquid chromatography/electrospray mass spectrometry determination of acyclovir in pregnant rat plasma and tissues

被引:55
作者
Brown, SD [1 ]
White, CA [1 ]
Bartlett, MG [1 ]
机构
[1] Univ Georgia, Coll Pharm, Dept Pharmaceut & Biomed Sci, Athens, GA 30602 USA
关键词
D O I
10.1002/rcm.792
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Reversed-phase chromatography is the most common means of separation for small drug molecules. However, polar drugs may suffer from poor retention and peak shape in reversed-phase high-performance liquid chromatography (RP-HPLC). Hydrophilic interaction liquid chromatography (HILIC) provides a viable alternative to RP-HPLC and is an excellent way to separate polar compounds. This paper describes a HILIC/ESI-MS/MS assay for the determination of acyclovir from rat plasma, amniotic fluid, placental tissue, and fetal tissue. The isocratic separation utilizes an underivatized silica column with an acetonitrile/formate buffer mobile phase (80:20). The method is validated over a range of 50 ng/mL to 50 mug/mL with % error and % relative standard deviation of <15% over 3 days. All samples are prepared by acetonitrile protein precipitation, which yields high recovery (>84% for acyclovir). This assay can be applied to the pharmacokinetic study of the placental transfer of acyclovir. Copyright (C) 2002 John Wiley Sons, Ltd.
引用
收藏
页码:1871 / 1876
页数:6
相关论文
共 33 条
[1]  
ANDREWS EB, 1992, OBSTET GYNECOL, V79, P7
[2]  
Baker DA, 1999, INT J FERTIL WOMEN M, V44, P227
[3]   Determination of acyclovir in human plasma by high-performance liquid chromatography [J].
Boulieu, R ;
Gallant, C ;
Silberstein, N .
JOURNAL OF CHROMATOGRAPHY B, 1997, 693 (01) :233-236
[4]   Determination of acyclovir in maternal plasma, amniotic fluid, fetal and placental tissues by high-performance liquid chromatography [J].
Brown, SD ;
White, CA ;
Chu, CK ;
Bartlett, MG .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2002, 772 (02) :327-334
[5]  
BURNETTE TC, 1994, DRUG METAB DISPOS, V22, P60
[6]   DETERMINATION OF ACYCLOVIR IN HUMAN-SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY [J].
CRONQVIST, J ;
NILSSONEHLE, I .
JOURNAL OF LIQUID CHROMATOGRAPHY, 1988, 11 (12) :2593-2601
[7]   EFFECT OF AGE ON THE GASTROINTESTINAL ABSORPTION OF ACYCLOVIR IN RATS [J].
FUJIOKA, Y ;
MIZUNO, N ;
MORITA, E ;
MOTOZONO, H ;
TAKAHASHI, K ;
YAMANAKA, Y ;
SHINKUMA, D .
JOURNAL OF PHARMACY AND PHARMACOLOGY, 1991, 43 (07) :465-469
[8]   Studies of signal suppression in liquid chromatography-electrospray ionization mass spectrometry using volatile ion-pairing reagents [J].
Gustavsson, SA ;
Samskog, J ;
Markides, KE ;
Långström, B .
JOURNAL OF CHROMATOGRAPHY A, 2001, 937 (1-2) :41-47
[9]  
HENDERSON GI, 1992, J LAB CLIN MED, V120, P885
[10]   Mechanistic investigation of ionization suppression in electrospray ionization [J].
King, R ;
Bonfiglio, R ;
Fernandez-Metzler, C ;
Miller-Stein, C ;
Olah, T .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2000, 11 (11) :942-950