Rapid scanning optical delay line based on a diffraction grating pair for a low-coherence reflectometer

被引:1
|
作者
Kolokoltsev, Y. [1 ]
Gomez-Arista, I. [2 ]
Garduno-Mejia, J. [2 ]
Avendano-Alejo, M. [2 ]
Trevino-Palacios, C. G. [3 ]
Kolokoltsev, O. [2 ]
机构
[1] Univ Nacl Autonoma Mexico, Fac Ciencias Fis, Mexico City, DF, Mexico
[2] Univ Nacl Autonoma Mexico, Ctr Ciencias Aplicadas & Desarrollo Tecnol, Mexico City, DF, Mexico
[3] INAOE, Puebla, Mexico
关键词
PULSE COMPRESSOR; PHASE MODULATOR; SWEPT-SOURCE; TOMOGRAPHY; DISPERSION; DESIGN; SYSTEM;
D O I
10.1364/AO.57.004542
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We present a simple low-coherence time-domain interferometric reflectometer with a rapidly scanning optical delay line (RSODL) based on a non-parallel diffraction grating (DG) pair. The novelty of the solution is that the lightwave in the reference channel is focused on a galvo-mirror in a sub-mm static spot, which allows implementation of fast microelectromechanical systems scan optics. It is shown that the DG pair can be operated as a non-dispersive element that provides dynamic group delay of a reference lightwave. The DG pair system is also capable of tuning the RSODL dispersion from negative to positive values. The experimental depth range in air was obtained as large as 2.5 mm for axial resolution of 20 mu m. (C) 2018 Optical Society of America
引用
收藏
页码:4542 / 4550
页数:9
相关论文
共 50 条
  • [1] Robust and rapid optical low-coherence reflectometer using a polygon mirror
    Delachenal, N.
    Wälti, R.
    Gianotti, R.
    Christov, S.
    Wagner, P.
    Salathé, R.P.
    Dürr, U.
    Ulbers, G.
    Optics Communications, 1999, 162 (04): : 195 - 199
  • [2] Robust and rapid optical low-coherence reflectometer using a polygon mirror
    Delachenal, N
    Wälti, R
    Gianotti, R
    Christov, S
    Wagner, P
    Salathé, RP
    Dürr, U
    Ulbers, G
    OPTICS COMMUNICATIONS, 1999, 162 (4-6) : 195 - 199
  • [3] Noninvasive monitoring of glucose concentration using differential absorption low-coherence interferometry based on rapid scanning optical delay line
    Zhou, Yong
    Zeng, Nan
    He, Yonghong
    9TH INTERNATIONAL CONFERENCE ON PHOTONICS AND IMAGING IN BIOLOGY AND MEDICINE (PIBM 2010), 2011, 277
  • [4] DESIGN OF A PRECISION OPTICAL LOW-COHERENCE REFLECTOMETER
    BOOSTER, DH
    CHOU, H
    HART, MG
    MIFSUD, SJ
    RAWSON, RF
    HEWLETT-PACKARD JOURNAL, 1993, 44 (01): : 39 - &
  • [5] Integrated-optic variable delay line and its application to a low-coherence reflectometer
    Takiguchi, K
    Itoh, M
    Takahashi, H
    OPTICS LETTERS, 2005, 30 (20) : 2739 - 2741
  • [6] Optical low-coherence reflectometer for differential phase measurement
    Davé, DP
    Milner, TE
    OPTICS LETTERS, 2000, 25 (04) : 227 - 229
  • [7] Optical low-coherence reflectometer for measuring WDM components
    NTT Opto-Electronics Laboratories, Tokai, Ibaraki-ken 319-1193, Japan
    不详
    IEEE Photonics Technol Lett, 7 (857-859):
  • [8] Optical low-coherence reflectometer for measuring WDM components
    Mechels, S
    Takada, K
    Okamoto, K
    IEEE PHOTONICS TECHNOLOGY LETTERS, 1999, 11 (07) : 857 - 859
  • [9] Scanning delay line with a rotating-parallelogram prism for low-coherence interferometry
    Giniunas, L
    Danielius, R
    Karkockas, R
    APPLIED OPTICS, 1999, 38 (34) : 7076 - 7079
  • [10] Spectral shaping in rapid scanning optical delay line of optical coherence tomography
    Wu, JG
    Xue, P
    Sun, S
    Guo, JH
    CHINESE PHYSICS LETTERS, 2003, 20 (12) : 2172 - 2174