Effectiveness of various decontamination methods on CAD/CAM camera mirror sleeves

被引:0
作者
Kruze, Jonathan P. [1 ]
Henrichs, Lori E. [2 ]
Mack, Kelli [3 ]
Vandewalle, Kraig S. [4 ]
机构
[1] Joint Base San Antonio, Randolph, TX USA
[2] Joint Base San Antonio, Clin Res Div, Lackland AFB, TX USA
[3] USAF, Infect Control, DECS, Ft Sam Houston, TX USA
[4] Joint Base San Antonio, Dent Res, Adv Educ Gen Dent Residency, 1615 Truemper St, Lackland AFB, TX 78236 USA
关键词
bacteria; decontamination; CAD/CAM; mirror sleeves; PSEUDOMONAS-AERUGINOSA; STAPHYLOCOCCUS-AUREUS; ORAL-CAVITY; DISINFECTION; ENDOSCOPES; SUSCEPTIBILITY; STERILIZATION; INSTRUMENT; EFFICACY;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Aim: The purpose of this study was to evaluate the effectiveness of several methods of disinfection and sterilization of computer-aided design/computer-aided manufacturing (CAD/CAM) camera mirror sleeves (Omnicam; Dentsply Sirona). Materials and methods: The outer surface of seven groups of mirror sleeves were inoculated by submerging them in suspensions of Staphylococcus aureus and Pseudomonas aeruginosa. Post inoculation, the groups were decontaminated as follows: Group A: no decontamination or sterilization following inoculation (positive control); Group B: surface cleaning with a neutral soap (Dawn Dish Soap, Procter & Gamble) and water only; Group C: surface disinfection with 17% isopropanol (CaviWipes; Metrex). Groups D to F received a different high-level disinfection (HLD) solution in an HLD container (Dentsply Sirona) as follows: Group D: 0.55% ortho-phthalaldehyde (OPA) (Cidex OPA; Johnson & Johnson); Group E: 7.5% hydrogen peroxide (H2O2) (Sporox II; Sultan); Group F: 7.35% H2O2 and 0.23% peracetic acid (PAA) (Compliance; Metrex). Group G received dry-heat sterilization (Rapid Heat Sterilizer; Cox). Also, dry-heat sterilized mirror sleeves that were not exposed to bacteria and not disinfected served as a negative control. The presence of bacteria was tested on the inside and outside of the sleeves by plating samples on TSA II. A percent reduction in CFU/ml from the positive control group was determined per group. Results: All methods of disinfection except Dawn Dish Soap resulted in greater than 99.99% reduction in CFU/ml compared with the positive control group. Conclusions: Both HLD or dry-heat sterilization resulted in no growth of microorganisms in cultures taken from both the outside and inside surfaces of the bacteria-contaminated mirror sleeves.
引用
收藏
页码:177 / 185
页数:9
相关论文
共 31 条
[1]  
Almeida Camilla Machado Feitosa de, 2012, Dental Press J. Orthod., V17, P105
[2]   Evaluation of the efficacy of glutaraldehyde and peroxygen for disinfection of dental instruments [J].
Angelillo, IF ;
Bianco, A ;
Nobile, CGA ;
Pavia, M .
LETTERS IN APPLIED MICROBIOLOGY, 1998, 27 (05) :292-296
[3]   Pseudomonas aeruginosa and Periodontal Pathogens in the Oral Cavity and Lungs of Cystic Fibrosis Patients: a Case-Control Study [J].
Caldas, Rocio Rivas ;
Le Gall, Florence ;
Revert, Krista ;
Rault, Gilles ;
Virmaux, Michele ;
Gouriou, Stephanie ;
Hery-Arnaud, Genevieve ;
Barbier, Georges ;
Boisrame, Sylvie .
JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (06) :1898-1907
[4]  
Carvalho Maria Reggiani Azevedo, 2015, J Contemp Dent Pract, V16, P619
[5]  
Centers for Disease Control and Prevention, 2016, Summary of Infection Prevention Practices in Dental Settings: Basic Expectations for Safe Care
[6]  
Centers for Disease Control and Prevention (CDC), GUIDELINE DISINFECTI, V52, P1
[7]  
Chu N S, 2000, Gastrointest Endosc Clin N Am, V10, P233
[8]   Synergistic Interactions of Pseudomonas aeruginosa and Staphylococcus aureus in an In Vitro Wound Model [J].
DeLeon, Stephanie ;
Clinton, Allie ;
Fowler, Haley ;
Everett, Jake ;
Horswill, Alexander R. ;
Rumbaugh, Kendra P. .
INFECTION AND IMMUNITY, 2014, 82 (11) :4718-4728
[9]  
Dentsply Sirona Dental Systems, TEST REP STER MIRR S, P1
[10]  
Dentsply Sirona Dental Systems, 2017, INSTR CEREC OMN CERE, P1