Design and Implementation of a New Wireless Carotid Neckband Doppler System with Wearable Ultrasound Sensors: Preliminary Results

被引:20
作者
Song, Ilseob [1 ]
Yoon, Jongmin [1 ]
Kang, Jinbum [1 ]
Kim, Min [1 ]
Jang, Won Seuk [2 ]
Shin, Na-Young [3 ]
Yoo, Yangmo [1 ,4 ]
机构
[1] Sogang Univ, Dept Elect Engn, Seoul 04107, South Korea
[2] Yonsei Univ, Dept Med Engn, Seoul 06229, South Korea
[3] Catholic Univ Korea, Dept Radiol, Seoul 06591, South Korea
[4] Sogang Univ, Dept Biomed Engn, Seoul 04107, South Korea
来源
APPLIED SCIENCES-BASEL | 2019年 / 9卷 / 11期
基金
新加坡国家研究基金会;
关键词
carotid ultrasonography; neckband; wearable ultrasound; point-of-care; SPECTRAL ESTIMATION TECHNIQUES; ARTERY-DISEASE; RESISTIVE INDEX; DIAGNOSIS; THROMBOSIS; VELOCITY;
D O I
10.3390/app9112202
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
Featured Application Continuous monitoring of cerebrovascular and cardiovascular diseases. Abstract Noninvasive monitoring of blood flow in the carotid artery is important for evaluating not only cerebrovascular but also cardiovascular diseases. In this paper, a wireless neckband ultrasound Doppler system, in which two 2.5-MHz ultrasonic sensors are utilized for acquiring Doppler signals from both carotid arteries, is presented for continuously evaluating blood flow dynamics. In the developed wireless neckband Doppler system, the acquired Doppler signals are quantized by 14-bit analog-to-digital-converters running at 40 MHz, and pre-processing operations (i.e., demodulation and clutter filtering) are performed in an embedded field programmable gate array chip. Then, these data are transferred to an external smartphone (i.e., Galaxy S7, Samsung Electronics Co., Suwon, Korea) via Bluetooth 2.0. Post-processing (i.e., Fourier transform and image processing) is performed using an embedded application processor in the smartphone. The developed carotid neckband Doppler system was evaluated with phantom and in vivo studies. In a phantom study, the neckband Doppler system showed comparable results with a commercial ultrasound machine in terms of peak systolic velocity and resistive index, i.e., 131.49 +/- 3.97 and 0.75 +/- 0.02 vs. 131.89 +/- 2.06 and 0.74 +/- 0.02, respectively. In addition, in the in vivo study, the neckband Doppler system successfully demonstrated its capability to continuously evaluate hemodynamics in both common carotid arteries. These results indicate that the developed wireless neckband Doppler system can be used for continuous monitoring of blood flow dynamics in the common carotid arteries in point-of-care settings.
引用
收藏
页数:12
相关论文
共 29 条
[1]  
Awad E., 1999, Proceedings of the First Joint BMES/EMBS Conference. 1999 IEEE Engineering in Medicine and Biology 21st Annual Conference and the 1999 Annual Fall Meeting of the Biomedical Engineering Society (Cat. No.99CH37015), DOI 10.1109/IEMBS.1999.803950
[2]  
Awad E., 1999, THESIS
[3]   TRANSCRANIAL DOPPLER MEASUREMENT OF MIDDLE CEREBRAL-ARTERY BLOOD-FLOW VELOCITY - A VALIDATION-STUDY [J].
BISHOP, CCR ;
POWELL, S ;
RUTT, D ;
BROWSE, NL .
STROKE, 1986, 17 (05) :913-915
[4]   Cerebral hemodynamic changes gauged by transcranial Doppler ultrasonooraphy in patients with posttraumatic brain swelling treated by surgical decompression [J].
Bor-Seng-Shu, E ;
Hirsch, R ;
Teixeira, MJ ;
De Andrade, AF ;
Marino, R .
JOURNAL OF NEUROSURGERY, 2006, 104 (01) :93-100
[5]   A CRITICAL-STUDY OF ULTRASOUND DOPPLER SPECTRAL-ANALYSIS FOR DETECTING CAROTID DISEASE [J].
BROWN, PM ;
JOHNSTON, KW ;
KASSAM, M ;
COBBOLD, RSC .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1982, 8 (05) :515-&
[6]   Relationship between the resistive index and vascular compliance and resistance [J].
Bude, RO ;
Rubin, JM .
RADIOLOGY, 1999, 211 (02) :411-417
[7]   A Generalized Unsharp Masking Algorithm [J].
Deng, Guang .
IEEE TRANSACTIONS ON IMAGE PROCESSING, 2011, 20 (05) :1249-1261
[8]   Comparison of carotid arterial resistive indices with intima-media thickness as sonographic markers of atherosclerosis [J].
Frauchiger, B ;
Schmid, HP ;
Roedel, C ;
Moosmann, P ;
Staub, D .
STROKE, 2001, 32 (04) :836-841
[9]   DUPLEX ULTRASOUND SCANNING OF THE CAROTID ARTERIES WITH VELOCITY SPECTRUM ANALYSIS [J].
GARTH, KE ;
CARROLL, BA ;
SOMMER, FG ;
OPPENHEIMER, DA .
RADIOLOGY, 1983, 147 (03) :823-827
[10]   Design and Implementation of a Smartphone-Based Portable Ultrasound Pulsed-Wave Doppler Device for Blood Flow Measurement [J].
Huang, Chih-Chung ;
Lee, Po-Yang ;
Chen, Pay-Yu ;
Liu, Ting-Yu .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2012, 59 (01) :182-188