Effective Dose from Stray Radiation for a Patient Receiving Proton Therapy for Liver Cancer

被引:0
作者
Taddei, Phillip J. [1 ]
Krishnan, Sunil [1 ]
Mirkovic, Dragan [1 ]
Yepes, Pablo [2 ]
Newhauser, Wayne D. [1 ]
机构
[1] Univ Texas MD Anderson Canc Ctr, 1515 Holcombe Blvd,Unit 94, Houston, TX 77030 USA
[2] Rice Univ, Houston, TX 77005 USA
来源
APPLICATION OF ACCELERATORS IN RESEARCH AND INDUSTRY | 2009年 / 1099卷
关键词
proton radiotherapy; stray radiation; secondary neutrons; liver cancer; Monte Carlo; human phantoms; MONTE-CARLO SIMULATIONS; PROSTATE-CANCER; HEPATOCELLULAR-CARCINOMA; EQUIVALENT EXPOSURES; SECONDARY NEUTRONS; UVEAL MELANOMA; RADIOTHERAPY; FIELD; NOZZLE; MCNPX;
D O I
暂无
中图分类号
O59 [应用物理学];
学科分类号
摘要
Because of its advantageous depth-dose relationship, proton radiotherapy is an emerging treatment modality for patients with liver cancer. Although the proton dose distribution conforms to the target, healthy tissues throughout the body receive low doses of stray radiation, particularly neutrons that originate in the treatment unit or in the patient. The aim of this study was to calculate the effective dose from stray radiation and estimate the corresponding risk of second cancer fatality for a patient receiving proton beam therapy for liver cancer. Effective dose from stray radiation was calculated using detailed Monte Carlo simulations of a double-scattering proton therapy treatment unit and a voxelized human phantom. The treatment plan and phantom were based on CT images of an actual adult patient diagnosed with primary hepatocellular carcinoma. For a prescribed dose of 60 Gy to the clinical target volume, the effective dose from stray radiation was 370 mSv; 61% of this dose was from neutrons originating outside of the patient while the remaining 39% was from neutrons originating within the patient. The excess lifetime risk of fatal second cancer corresponding to the total effective dose from stray radiation was 1.2%. The results of this study establish a baseline estimate of the stray radiation dose and corresponding risk for an adult patient undergoing proton radiotherapy for liver cancer and provide new evidence to corroborate the suitability of proton beam therapy for the treatment of liver tumors.
引用
收藏
页码:445 / +
页数:2
相关论文
共 32 条
[11]   Simulation of organ-specific patient effective dose due to secondary neutrons in proton radiation treatment [J].
Jiang, HY ;
Wang, B ;
Xu, XG ;
Suit, HD ;
Paganetti, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (18) :4337-4353
[12]   Virtual commissioning of a treatment planning system for proton therapy of ocular cancers [J].
Koch, N ;
Newhauser, W .
RADIATION PROTECTION DOSIMETRY, 2005, 115 (1-4) :159-163
[13]   Monte Carlo calculations and measurements of absorbed dose per monitor unit for the treatment of uveal melanoma with proton therapy [J].
Koch, Nicholas ;
Newhauser, Wayne D. ;
Titt, Uwe ;
Gombos, Dan ;
Coombes, Kevin ;
Starkschall, George .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (06) :1581-1594
[14]   Radiotherapy for hepatocellular carcinoma: An overview [J].
Krishnan, Sunil ;
Dawson, Laura A. ;
Seong, Jinsil ;
Akine, Yasuyuki ;
Beddar, Sam ;
Briere, Tina M. ;
Crane, Christopher H. ;
Mornex, Francoise .
ANNALS OF SURGICAL ONCOLOGY, 2008, 15 (04) :1015-1024
[15]  
NCRP. National Council on Radiation Protection and Measurements, 1997, UNC FAT CANC RISK ES
[16]   Monte Carlo simulations of a nozzle for the treatment of ocular tumours with high-energy proton beams [J].
Newhauser, W ;
Koch, N ;
Hummel, S ;
Ziegler, M ;
Titt, U .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (22) :5229-5249
[17]   Monte Carlo simulations for configuring and testing an analytical proton dose-calculation algorithm [J].
Newhauser, Wayne ;
Fontenot, Jonas ;
Zheng, Yuanshui ;
Polf, Jerimy ;
Titt, Uwe ;
Koch, Nicholas ;
Zhang, Xiaodong ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (15) :4569-4584
[18]   Can megavoltage computed tomography reduce proton range uncertainties in treatment plans for patients with large metal implants? [J].
Newhauser, Wayne D. ;
Giebeler, Annelise ;
Langen, Katja M. ;
Mirkovic, Dragan ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2327-2344
[19]   Dosimetric impact of tantalum markers used in the treatment of uveal melanoma with proton beam therapy [J].
Newhauser, Wayne D. ;
Koch, Nicholas C. ;
Fontenot, Jonas D. ;
Rosenthal, Stanley J. ;
Gombos, Dan S. ;
Fitzek, Markus M. ;
Mohan, Radhe .
PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (13) :3979-3990
[20]  
PELOWITZ DB, 2005, MCNPXTM USERS MANUAL