Comparison of Anterior Segment Biometric Measurements between Pentacam HR and IOLMaster in Normal and High Myopic Eyes

被引:18
作者
Dong, Jing [1 ]
Tang, Maolong [2 ]
Zhang, Yaqin [3 ]
Jia, Yading [3 ]
Zhang, Haining [3 ]
Jia, Zhijie [3 ]
Wang, Xiaogang [3 ]
机构
[1] Shanxi Med Univ, Hosp 1, Taiyuan, Shanxi, Peoples R China
[2] Oregon Hlth & Sci Univ, Casey Eye Inst, Portland, OR 97201 USA
[3] Shanxi Eye Hosp, Taiyuan, Shanxi, Peoples R China
关键词
CHAMBER DEPTH MEASUREMENTS; LENS POWER CALCULATION; CORNEAL TOPOGRAPHY; REFRACTIVE ERROR; AXIAL LENGTH; REPEATABILITY; KERATOMETRY; THICKNESS; ORBSCAN; REPRODUCIBILITY;
D O I
10.1371/journal.pone.0143110
中图分类号
O [数理科学和化学]; P [天文学、地球科学]; Q [生物科学]; N [自然科学总论];
学科分类号
07 ; 0710 ; 09 ;
摘要
Purpose To compare the anterior chamber depth (ACD), keratometry (K) and astigmatism measurements taken by IOLMaster and Pentacam HR in normal and high myopic (HM) eyes. Design A prospective observational case series. Methods Sixty-six normal eyes and 59 HM eyes underwent ACD, keratometry and astigmatism measurements with both devices. Axial length (AL) was measured on IOLMaster. The interdevice agreement was evaluated using the Bland-Altman analysis and paired t-test. The correlations between age and AL & ACD were analyzed. Vector analysis was used to compare astigmatism measurements. Results The ACD from IOLMaster and Pentacam HR was different for the normal group (P = 0.003) but not for the HM group (P = 0.280). IOLMaster demonstrated higher steep K and mean K values than Pentacam HR for both normal and HM groups (P<0.001 for all). IOLMaster also have higher flat K values for the HM groups (P<0.001) but were statistically equivalent with Pentacam HR for the normal group (P = 0.119) IOLMaster and Pentacam HR were different in astigmatism measurements for the normal group but were statistically equivalent for the HM group. For the normal group, age was negatively correlated with AL, IOLMaster ACD and Pentacam HR ACD (r = -0.395, P = 0.001; r = -0.715, P < 0.001; r = -0.643, P < 0.001). For the HM group, age was positively correlated with AL but negatively correlated with IOLMaster ACD and Pentacam HR ACD (r = 0.377, P = 0.003; r = -0.392, P = 0.002; r = -0.616, P < 0.001). Conclusions The IOLMaster and Pentacam HR have significant difference in corneal power measurements for both normal and HM groups. The two instruments also differ in ACD and astigmatism measurement for the normal group. Therefore, a single instrument is recommended for studying longitudinal changes in anterior segment biometric measurements. Age should be considered as an influencing factor for both AL and ACD values in the normal and HM group.
引用
收藏
页数:10
相关论文
共 30 条
[1]   Repeatability of IOLMaster biometry in children [J].
Carkeet, A ;
Saw, SM ;
Gazzard, G ;
Tang, W ;
Tan, DTH .
OPTOMETRY AND VISION SCIENCE, 2004, 81 (11) :829-834
[2]   Agreement Between Placido Topography and Scheimpflug Tomography for Corneal Astigmatism Assessment [J].
Delrivo, Marianella ;
Ruisenor Vazquez, Pablo R. ;
David Galletti, Jonatan ;
Garibotto, Martin ;
Fuentes Bonthoux, Fernando ;
Pfoertner, Tomas ;
Gaston Galletti, Jeremias .
JOURNAL OF REFRACTIVE SURGERY, 2014, 30 (01) :49-53
[3]   Comparison of different techniques of anterior chamber depth and keratometric measurements [J].
Elbaz, Uri ;
Barkana, Yaniv ;
Gerber, Yariv ;
Avni, Isaac ;
Zadok, David .
AMERICAN JOURNAL OF OPHTHALMOLOGY, 2007, 143 (01) :48-53
[4]   Corneal thickness, curvature, and elevation readings in normal corneas: Combined Placido-Scheimpflug system versus combined Placido-scanning-slit system [J].
Guilbert, Emmanuel ;
Saad, Alain ;
Grise-Dulac, Alice ;
Gatinel, Damien .
JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (07) :1198-1206
[5]  
Huang D, 1999, J REFRACT SURG, V15, P538
[7]   Repeatability and Reproducibility of Corneal Curvature Measurements Using the Pentacam and Keratron Topography Systems [J].
Kawamorita, Takushi ;
Nakayama, Nanami ;
Uozato, Hiroshi .
JOURNAL OF REFRACTIVE SURGERY, 2009, 25 (06) :539-544
[8]  
Klein BEK, 1998, INVEST OPHTH VIS SCI, V39, P1507
[9]   Validity and Repeatability of Anterior Chamber Depth Measurements With Pentacam and Orbscan [J].
Lackner, Birgit ;
Schmidinger, Gerald ;
Skorpik, Christian .
OPTOMETRY AND VISION SCIENCE, 2005, 82 (09) :858-861
[10]   The repeatability and accuracy of axial length and anterior chamber depth measurements from the IOLMaster™ [J].
Lam, AKC ;
Chan, R ;
Pang, PCK .
OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2001, 21 (06) :477-483