Calibration of LiBaF3:Ce scintillator for fission spectrum neutrons

被引:4
作者
Reeder, PL [1 ]
Bowyer, SM [1 ]
机构
[1] Pacific NW Natl Lab, Richland, WA 99352 USA
关键词
LiBaF3 : Ce scintillator; neutron calibration; time-tagged Cf-252;
D O I
10.1016/S0168-9002(01)02007-1
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The scintillator LiBaF3 doped with small amounts of Ce+3 has the ability to distinguish heavy charged particles (p, d, t, or alpha) from beta and/or gamma radiation based on the presence or absence of nanosecond components in the scintillation light output, Since the neutron capture reaction on Li-6 produces recoil alphas and tritons, this scintillator also discriminates between neutron induced events and beta or gamma interactions. An experimental technique using a time-tagged Cf-252 source has been used to measure the efficiency of this scintillator for neutron capture, the calibration of neutron capture pulse height, and the pulse height resolution-all as a function of incident neutron energy. (C) 2002 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:469 / 485
页数:17
相关论文
共 14 条
  • [11] REEDER PL, 2000, C P IEEE NSS MIC 200
  • [12] REEDER PL, 2000, P METH APPL RAD CHEM
  • [13] REEDER PL, 2000, P C APPL ACC RES IND
  • [14] REILLY D, 1991, PASSIVE NONDESTRUCTI, pCH11