Effect of automatic tube voltage selection on image quality and radiation dose in abdominal CT angiography of various body sizes: A phantom study

被引:42
作者
Schindera, S. T. [1 ,2 ]
Winklehner, A. [3 ]
Alkadhi, H. [3 ]
Goetti, R. [3 ]
Fischer, M. [3 ]
Gnannt, R. [3 ]
Szucs-Farkas, Z. [1 ]
机构
[1] Univ Bern, Univ Hosp Berne, Inst Diagnost Intervent & Pediat Radiol, Bern, Switzerland
[2] Univ Basel Hosp, Dept Radiol, CH-4031 Basel, Switzerland
[3] Univ Zurich Hosp, Dept Diagnost & Intervent Radiol, CH-8091 Zurich, Switzerland
关键词
PATIENT SIZE; NOISE; ENHANCEMENT; ATTENUATION; REDUCTION; SETTINGS; 16-MDCT; ENERGY; TUMORS; PHASE;
D O I
10.1016/j.crad.2012.10.007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
AIM: To assess the effect of an automatic tube voltage selection technique on image quality and radiation dose in abdominal computed tomography (CT) angiography of various body sizes. MATERIALS AND METHODS: An abdominal aortic phantom was filled with iodinated contrast medium and placed into three different cylindrical water containers, which simulated a small, intermediate-sized, and large patient. The phantom was scanned with a standard 120 kVp abdominal CT angiography protocol and with an optimized tube voltage protocol that was modulated by an automatic tube voltage technique. The attenuation of the aorta, background, and image noise was measured, and the contrast-to-noise ratio (CNR) was calculated. Three independent readers assessed the overall image quality. RESULTS: The automatic tube voltage technique selected 70 kVp as the optimal tube voltage for the small phantom, 80 kVp for the intermediate phantom, and 100 kVp for the large phantom. Compared to the standard 120 kVp protocol, the automatic tube voltage selection yielded significantly increased CNR values in the small phantom (15.8 versus 19.4, p < 0.001), intermediate phantom (8.4 versus 8.7, p < 0.05), and large phantom (4.3 versus 4.6, p < 0.01). The automatic tube voltage selection resulted in a 55%, 49%, and 39% reduction in the volume CT dose index (CTDIvol) in the small, intermediate, and large phantoms, respectively. The subjective overall image quality of the three phantom sizes at different tube voltages ranged between poor and good. CONCLUSION: Compared to a standard 120 kVp abdominal CT angiography protocol, the automatic tube voltage selection substantially reduced the radiation dose without compromising image quality in various simulated patient sizes. (C) 2012 The Royal College of Radiologists. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:E79 / E86
页数:8
相关论文
共 18 条
[1]   Dual-energy-CT of hypervascular liver lesions in patients with HCC: investigation of image quality and sensitivity [J].
Altenbernd, Jens ;
Heusner, Till A. ;
Ringelstein, Adrian ;
Ladd, Susanne C. ;
Forsting, Michael ;
Antoch, Gerald .
EUROPEAN RADIOLOGY, 2011, 21 (04) :738-743
[2]   Multi-detector row CT angiography of the brain at various kilovoltage settings [J].
Ertl-Wagner, BB ;
Hoffmann, RT ;
Bruning, R ;
Herrmann, K ;
Snyder, B ;
Blume, JD ;
Reiser, MF .
RADIOLOGY, 2004, 231 (02) :528-535
[3]   Appropriate Patient Selection at Abdominal Dual-Energy CT Using 80 kV: Relationship between Patient Size, Image Noise, and Image Quality [J].
Guimaraes, Luis S. ;
Fletcher, Joel G. ;
Harmsen, William S. ;
Yu, Lifeng ;
Siddiki, Hassan ;
Melton, Zachary ;
Huprich, James E. ;
Hough, David ;
Hartman, Robert ;
McCollough, Cynthia H. .
RADIOLOGY, 2010, 257 (03) :732-742
[4]   Using the K-edge to improve contrast conspicuity and to lower radiation dose with a 16-MDCT: a phantom and human study [J].
Kalva, Sanjeeva P. ;
Sahani, Dushyant V. ;
Hahn, Peter F. ;
Saini, Sanjay .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2006, 30 (03) :391-397
[5]   Low kilovoltage cardiac dual-source CT: attenuation, noise, and radiation dose [J].
Leschka, Sebastian ;
Stolzmann, Paul ;
Schmid, Florian T. ;
Scheffel, Hans ;
Stinn, Bjoern ;
Marincek, Borut ;
Alkadhi, Hatem ;
Wildermuth, Simon .
EUROPEAN RADIOLOGY, 2008, 18 (09) :1809-1817
[6]   Detection of Pancreatic Tumors, Image Quality, and Radiation Dose during the Pancreatic Parenchymal Phase: Effect of a Low-Tube-Voltage, High-Tube-Current CT Technique-Preliminary Results [J].
Marin, Daniele ;
Nelson, Rendon C. ;
Barnhart, Huiman ;
Schindera, Sebastian T. ;
Ho, Lisa M. ;
Jaffe, Tracy A. ;
Yoshizumi, Terry T. ;
Youngblood, Richard ;
Samei, Ehsan .
RADIOLOGY, 2010, 256 (02) :450-459
[7]   Hypervascular Liver Tumors: Low Tube Voltage, High Tube Current Multidetector CT during Late Hepatic Arterial Phase for Detection-Initial Clinical Experience [J].
Marin, Daniele ;
Nelson, Rendon C. ;
Samei, Ehsan ;
Paulson, Erik K. ;
Ho, Lisa M. ;
Boll, Daniel T. ;
DeLong, David M. ;
Yoshizumi, Terry T. ;
Schindera, Sebastian T. .
RADIOLOGY, 2009, 251 (03) :771-779
[8]   Comparison of different body size parameters for individual dose adaptation in body CT of adults [J].
Menke, J .
RADIOLOGY, 2005, 236 (02) :565-571
[9]   Abdominal CT with low tube voltage: Preliminary observations about radiation dose, contrast enhancement, image quality, and noise [J].
Nakayama, Y ;
Awai, K ;
Funama, Y ;
Hatemura, M ;
Imuta, M ;
Nakaura, T ;
Ryu, D ;
Morishita, S ;
Sultana, S ;
Sato, N ;
Yamashita, Y .
RADIOLOGY, 2005, 237 (03) :945-951
[10]   16-MDCT angiography in living kidney donors at various tube potentials: Impact on image quality and radiation dose [J].
Sahani, Dushyant V. ;
Kalva, Sanjeeva P. ;
Hahn, Peter F. ;
Saini, Sanjay .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2007, 188 (01) :115-120