The impact on CT dose of the variability in tube current modulation technology: a theoretical investigation

被引:38
作者
Li, Xiang [1 ]
Segars, W. Paul [2 ]
Samei, Ehsan [2 ,3 ,4 ]
机构
[1] Cleveland State Univ, Dept Phys, Med Phys Grad Program, Cleveland, OH 44115 USA
[2] Duke Univ, Med Ctr, Med Phys Grad Program, Carl E Ravin Adv Imaging Labs,Dept Radiol, Durham, NC 27705 USA
[3] Duke Univ, Med Ctr, Med Phys Grad Program, Carl E Ravin Adv Imaging Labs,Dept Biomed Engn, Durham, NC 27705 USA
[4] Duke Univ, Med Ctr, Med Phys Grad Program, Carl E Ravin Adv Imaging Labs,Dept Elect & Comp E, Durham, NC 27705 USA
基金
美国国家卫生研究院;
关键词
CT; radiation; dose; Monte Carlo; tube current modulation; CANCER-RISK ESTIMATION; MONTE-CARLO; AUTOMATED EXTRACTION; PEDIATRIC CHEST; ORGAN; FEASIBILITY; INFORMATION; REDUCTION; PROGRAM; PHANTOM;
D O I
10.1088/0031-9155/59/16/4525
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Body CT scans are routinely performed using tube-current-modulation (TCM) technology. There is notable variability across CT manufacturers in terms of how TCM technology is implemented. Some manufacturers aim to provide uniform image noise across body regions and patient sizes, whereas others aim to provide lower noise for smaller patients. The purpose of this study was to conduct a theoretical investigation to understand how manufacturer-dependent TCM scheme affects organ dose, and to develop a generic approach for assessing organ dose across TCM schemes. The adult reference female extended cardiac-torso (XCAT) phantom was used for this study. A ray-tracing method was developed to calculate the attenuation of the phantom for a given projection angle based on phantom anatomy, CT system geometry, x-ray energy spectrum, and bowtie filter filtration. The tube current (mA) for a given projection angle was then calculated as a log-linear function of the attenuation along that projection. The slope of this function, termed modulation control strength, alpha, was varied from 0 to 1 to emulate the variability in TCM technology. Using a validated Monte Carlo program, organ dose was simulated for five alpha values (alpha = 0, 0.25, 0.5, 0.75, and 1) in the absence and presence of a realistic system mA limit. Organ dose was further normalized by volume-weighted CT dose index (CTDIvol) to obtain conversion factors (h factors) that are relatively independent of system specifics and scan parameters. For both chest and abdomen-pelvis scans and for 24 radiosensitive organs, organ dose conversion factors varied with alpha, following second-order polynomial equations. This result suggested the need for alpha-specific organ dose conversion factors (i.e., conversion factors specific to the modulation scheme used). On the other hand, across the full range of a values, organ dose in a TCM scan could be derived from the conversion factors established for a fixed-mA scan (h(FIXED)). This was possible by multiplying h(FIXED) by a revised definition of CTDIvol that accounts for two factors: (alpha) the tube currents at the location of an organ and (b) the variation in organ volume along the longitudinal direction. This alpha-generic approach represents an approximation. The error associated with this approximation was evaluated using the alpha-specific organ dose (i.e., the organ dose obtained by using alpha-specific mA profiles as inputs into the Monte Carlo simulation) as the reference standard. When the mA profiles were constrained by a realistic system limit, this alpha-generic approach had errors of less than similar to 20% for the full range of alpha values. This was the case for 24 radiosensitive organs in both chest and abdomen-pelvis CT scans with the exception of thyroid in the chest scan and bladder in the abdomen-pelvis scan. For these two organs, the errors were less than similar to 40%. The results of this theoretical study suggested that knowing the mA modulation profile and the fixed-mA conversion factors, organ dose may be estimated for a TCM scan independent of the specific modulation scheme applied.
引用
收藏
页码:4525 / 4548
页数:24
相关论文
共 40 条
[1]   Dose to Radiosensitive Organs During Routine Chest CT: Effects of Tube Current Modulation [J].
Angel, Erin ;
Yaghmai, Nazanin ;
Jude, Cecilia Matilda ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Goldin, Jonathan G. ;
McCollough, Cynthia H. ;
Primak, Andrew N. ;
Cody, Dianna D. ;
Stevens, Donna M. ;
McNitt-Gray, Michael F. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2009, 193 (05) :1340-1345
[2]   Monte Carlo simulations to assess the effects of tube current modulation on breast dose for multidetector CT [J].
Angel, Erin ;
Yaghmai, Nazanin ;
Jude, Cecilia Matilda ;
DeMarco, John J. ;
Cagnon, Christopher H. ;
Goldin, Jonathan G. ;
Primak, Andrew N. ;
Stevens, Donna M. ;
Cody, Dianna D. ;
McCollough, Cynthia H. ;
McNitt-Gray, Michael F. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (03) :497-511
[3]  
[Anonymous], 97 SCI ASS ANN M RAD
[4]  
[Anonymous], 2006, Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII Phase 2, V7
[5]  
[Anonymous], 2005, 05016 MHRA, DOI DOI 10.13140/RG.2.1.3128.4720
[6]  
[Anonymous], 2007, ANN ICRP
[7]  
[Anonymous], 2003, Ann ICRP, V32, P3
[8]  
Boone J.M., 2011, SIZE SPECIFIC DOSE E
[9]   Automated size-specific CT dose monitoring program: Assessing variability in CT dose [J].
Christianson, Olav ;
Li, Xiang ;
Frush, Donald ;
Samei, Ehsan .
MEDICAL PHYSICS, 2012, 39 (11) :7131-7139
[10]   Automated Extraction of Radiation Dose Information for CT Examinations [J].
Cook, Tessa S. ;
Zimmerman, Stefan ;
Maidment, Andrew D. A. ;
Kim, Woojin ;
Boonn, William W. .
JOURNAL OF THE AMERICAN COLLEGE OF RADIOLOGY, 2010, 7 (11) :871-877