Characterization of a focal plane camera fitted to a Mattauch-Herzog geometry mass spectrograph. 1. Use with a glow-discharge source

被引:44
作者
Barnes, JH
Sperline, R
Denton, MB
Barinaga, CJ
Koppenaal, D
Young, ET
Hieftje, GM [1 ]
机构
[1] Indiana Univ, Dept Chem, Bloomington, IN 47405 USA
[2] Univ Arizona, Dept Chem, Tucson, AZ 85721 USA
[3] Pacific NW Natl Lab, Richland, WA 99352 USA
[4] Univ Arizona, Steward Observ, Tucson, AZ 85721 USA
关键词
D O I
10.1021/ac020304z
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A Mattauch-Herzog geometry mass spectrograph (MHMS) has been equipped with a novel array detector, the focal plane camera (FPC). The FPC consists of an array of gold Faraday cups, each coupled to its own integrator, with interrogation of the integrators performed by a multiplexer. The initial coupling of this instrument with a pin-type glow discharge source has provided limits of detection in the single to hundreds of nanograms per gram regime; isotope ratio accuracy and precision better than 5% error and 0.2% RSD, respectively; and a linear dynamic range of at least 6 orders of magnitude. A current weakness of the FPC is its pixel size, which limits both sensitivity and baseline resolution (to R =. 130). The minimum data acquisition time for multiple images at present is 1 ms/image, with a dead time of 3.2 ms between images, which will limit the ability of the FPC to monitor extremely short transient signals.
引用
收藏
页码:5327 / 5332
页数:6
相关论文
共 27 条
[21]   The LEDA512 integrated circuit anode array for the analog recording of mass spectra [J].
Nevejans, D ;
Neefs, E ;
Kavadias, S ;
Merken, P ;
Van Hoof, C .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 215 (1-3) :77-87
[22]   Evaluation of a microsphere plate detector for an orthogonal acceleration matrix-assisted laser desorption/ionisation time-of-flight mass spectrometer [J].
Selby, DS ;
Mlynski, V ;
Guilhaus, M .
INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2002, 215 (1-3) :31-43
[23]   Advancing the capabilities of a glow-discharge sector-field mass spectrometer [J].
Solyom, DA ;
Hieftje, GM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2002, 17 (04) :329-333
[24]   Analytical capabilities of an inductively coupled plasma Mattauch-Herzog mass spectrometer [J].
Solyom, DA ;
Gron, OA ;
Barnes, JH ;
Hieftje, GM .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2001, 56 (09) :1717-1729
[25]   Plasma-source sector mass spectrometry with array detection [J].
Solyom, DA ;
Burgoyne, TW ;
Hieftje, GM .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1999, 14 (08) :1101-1110
[26]  
SOLYOM DA, 2001, THESIS INDIANA U BLO
[27]   TIME-OF-FLIGHT MASS SPECTROMETER WITH IMPROVED RESOLUTION [J].
WILEY, WC ;
MCLAREN, IH .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1955, 26 (12) :1150-1157