High-sensitivity instrumentation for spectrally-resolved optically detected X-ray absorption spectroscopy

被引:12
作者
Poolton, NRJ [1 ]
Botter-Jensen, L
Denby, PM
Nakamura, T
Hamilton, B
Pantos, E
机构
[1] Daresbury Lab, Synchrotron Radiat Dept, Warrington WA4 4AD, Cheshire, England
[2] Riso Natl Lab, Radiat Res Dept, DK-4000 Roskilde, Denmark
[3] Shizuoka Univ, Fac Engn, Dept Mat Sci & Technol, Hamamatsu, Shizuoka 4328561, Japan
[4] Univ Manchester, Inst Sci & Technol, Dept Phys, Manchester M60 1QD, Lancs, England
基金
英国工程与自然科学研究理事会;
关键词
synchrotron; luminescence; spectrometer; XAS; OD-XAS; XEOL;
D O I
10.1016/j.nimb.2004.05.003
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The design use and implementation of simple but highly efficient instrumentation is described, that allows for the possibility of wavelength-resolving optically detected X-ray absorption spectra (OD-XAS) of even the weakest luminescent systems emitting in the range 380-1020 nm (1.2-3.3 eV). The instrument, based on linearly-variable interference filters, provides significant improvement in light throughput as compared with standard grating monochromators, and is used to demonstrate the power of wavelength-resolved OD-XAS in unravelling luminescence processes that are typically encountered in two common types of systems: (i) single-phase materials that possess a combination of radically different recombination pathways and (ii) materials where the recombination mechanisms are essentially invariant throughout the system, but which are chemically phase-separated on a sub-microscopic scale. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:590 / 598
页数:9
相关论文
共 24 条
  • [1] X-RAY-INDUCED LUMINESCENCE IN CRYSTALLINE SIO2
    ALONSO, PJ
    HALLIBURTON, LE
    KOHNKE, EE
    BOSSOLI, RB
    [J]. JOURNAL OF APPLIED PHYSICS, 1983, 54 (09) : 5369 - 5375
  • [2] BAILIFF IK, 1997, J PHYS E, V10, P1156
  • [3] BANGERT U, 1995, PHYS STATUS SOLIDI B, V190, P69, DOI 10.1002/pssb.2221900111
  • [4] INTRINSIC LUMINESCENCE EXCITATION SPECTRUM AND EXTENDED X-RAY ABSORPTION FINE-STRUCTURE ABOVE K EDGE IN CAF2
    BIANCONI, A
    JACKSON, D
    MONAHAN, K
    [J]. PHYSICAL REVIEW B, 1978, 17 (04): : 2021 - 2024
  • [5] A COMPACT DESIGN FOR MONOCHROMATIC OSL MEASUREMENTS IN THE WAVELENGTH RANGE 380-1020-NM
    BOTTERJENSEN, L
    POOLTON, NRJ
    WILLUMSEN, F
    CHRISTIANSEN, H
    [J]. RADIATION MEASUREMENTS, 1994, 23 (2-3) : 519 - 522
  • [6] A new XUV beamline on a multi-pole wiggler in the SRS
    Bowler, M
    West, JB
    Quinn, FM
    Holland, DMP
    Fell, B
    Hatherly, PA
    Humphrey, I
    Flavell, WR
    Hamilton, B
    [J]. SURFACE REVIEW AND LETTERS, 2002, 9 (01) : 577 - 581
  • [7] Photoluminescence at 1.9 eV in synthetic wet silica
    Cannas, M
    Leone, M
    [J]. JOURNAL OF NON-CRYSTALLINE SOLIDS, 2001, 280 (1-3) : 183 - 187
  • [8] Evidence of x-ray absorption-edge shift as a function of luminescence wavelength in porous silicon
    Dalba, G
    Daldosso, N
    Fornasini, P
    Grimaldi, M
    Grisenti, R
    Rocca, F
    [J]. PHYSICAL REVIEW B, 2000, 62 (15): : 9911 - 9914
  • [9] Deer W. A., 1992, INTRO ROCK FORMING M
  • [10] Efficient technique for producing high-brightness, size-selected cluster beams
    Denby, PM
    Eastham, DA
    [J]. APPLIED PHYSICS LETTERS, 2001, 79 (15) : 2477 - 2479