Objectives: Evaluation of the performance of the EMIT 2000 Tacrolimus assay on the Abbott Architect c8000 analyzer. Design and Methods: Imprecision studies were performed and patient samples were assayed by EMIT assay and by LC-MS/MS. Results: Limit of quantification was established at 2.8 mu g/L. A positive bias of 17.5% between results measured on EMIT and on LC-MS/MS was detected. Conclusions: EMIT 2000 Tacrolimus assay is suitable for automated analyses of Tacrolimus on the Architect c8000. (c) 2006 The Canadian Society of Clinical Chemists. All rights reserved.
引用
收藏
页码:1041 / 1043
页数:3
相关论文
共 9 条
[1]  
BEHRING D, 2004, EMIT 2000 TACROLIMUS
[2]   Low hematocrit and serum albumin concentrations underlie the overestimation of tacrolimus concentrations by microparticle enzyme immunoassay versus liquid chromatography-tandem mass spectrometry [J].
Brown, NW ;
Gonde, CE ;
Adams, JE ;
Tredger, JM .
CLINICAL CHEMISTRY, 2005, 51 (03) :586-592
[3]  
IWASAKI K, 1995, DRUG METAB DISPOS, V23, P28
[4]   Simultaneous determination of four immunosuppressants by means of high speed and robust on-line solid phase extraction-high performance liquid chromatography-tandem mass spectrometry [J].
Koal, T ;
Deters, M ;
Casetta, B ;
Kaever, V .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2004, 805 (02) :215-222
[5]   The EMIT 2000 tacrolimus assay: an application protocol for the Beckman Synchron LX20 PRO analyzer [J].
LeGatt, DF ;
Shalapay, CE ;
Cheng, SB .
CLINICAL BIOCHEMISTRY, 2004, 37 (11) :1022-1030
[6]   Evaluation of the Dade Behring Syva® EMIT 2000 tacrolimus assay on the Bayer Advia 1650 [J].
Stephen, DWS ;
Rooke, P ;
Clark, S ;
Coutts, M ;
Crowe, L ;
Docherty, D .
ANNALS OF CLINICAL BIOCHEMISTRY, 2003, 40 :697-700
[7]   Therapeutic drug monitoring of immunosuppressant drugs by high-performance liquid chromatography-mass spectrometry [J].
Taylor, PJ .
THERAPEUTIC DRUG MONITORING, 2004, 26 (02) :215-219
[8]   Therapeutic monitoring of immunosuppressant drugs. Where are we? [J].
Wallemacq, PE .
CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2004, 42 (11) :1204-1211
[9]   Comparative clinical pharmacokinetics of tacrolimus in paediatric and adult patients [J].
Wallemacq, PE ;
Verbeeck, RK .
CLINICAL PHARMACOKINETICS, 2001, 40 (04) :283-295