Surface Characterization of Retrieved Metal-on-Metal Total Hip Implants from Patients with Adverse Reaction to Metal Debris

被引:4
作者
Burbano, Maria [1 ]
Russell, Robert [2 ]
Huo, Michael [2 ]
Welch, Robert [2 ]
Roy, Diana [1 ]
Rodrigues, Danieli C. [1 ]
机构
[1] Univ Texas Dallas, Dept Bioengn, Richardson, TX 75080 USA
[2] Univ Texas SW Med Ctr Dallas, Dept Orthoped Surg, Dallas, TX 75235 USA
来源
MATERIALS | 2014年 / 7卷 / 03期
关键词
metal-on-metal; adverse local tissue reaction; retrieval; pseudotumor; RESURFACING ARTHROPLASTY; RISK-FACTORS; ION LEVELS; CORROSION; FAILURE; REPLACEMENTS; INCLINATION; COMPONENT; CHROMIUM; COBALT;
D O I
10.3390/ma7031866
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The use of metal-on-metal (MoM) total hip implants has decreased recently due to reports of high failure rates and adverse local tissue reaction (ALTR). It has been hypothesized that wear metal debris released from CoCr bearing surfaces may provoke delayed hypersensitivity reactions. The goal of this study is to evaluate the microscopic bearing surface characteristics of implants revised due to evidence of ALTR. The bearing surface of each head and cup was analyzed using multiple microscopy techniques for characterization of the surface features. The presence of severe mechanical scratching was a common characteristic found in all of the implants evaluated. Mechanical factors seemed to be the prevalent failure mode related to the appearance of ALTR with this particular set of retrieved implants.
引用
收藏
页码:1866 / 1879
页数:14
相关论文
共 30 条
  • [1] Corrosion fatigue of biomedical metallic alloys: Mechanisms and mitigation
    Antunes, Renato Altobelli
    Lopes de Oliveira, Mara Cristina
    [J]. ACTA BIOMATERIALIA, 2012, 8 (03) : 937 - 962
  • [2] Cup inclination and serum concentration of cobalt and chromium after metal-on-metal total hip arthroplasty
    Brodner, W
    Grübl, A
    Jankovsky, R
    Meisinger, V
    Lehr, S
    Gottsauner-Wolf, F
    [J]. JOURNAL OF ARTHROPLASTY, 2004, 19 (08) : 66 - 70
  • [3] Failure analysis of cemented metal-on-metal total hip replacements from a single centre cohort
    Bryant, Michael
    Ward, Michael
    Farrar, Richard
    Freeman, Robert
    Brummitt, Ken
    Nolan, John
    Neville, Anne
    [J]. WEAR, 2013, 301 (1-2) : 226 - 233
  • [4] Inflammatory pseudotumor associated with femoral nerve palsy following metal-on-metal resurfacing of the hip
    Clayton, Robert A. E.
    Beggs, Ian
    Salter, Donald M.
    Grant, M. Helen
    Patton, James T.
    Porter, Daniel E.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (09) : 1988 - 1993
  • [5] Corrosion at the Head-Neck Taper as a Cause for Adverse Local Tissue Reactions After Total Hip Arthroplasty
    Cooper, H. John
    Della Valle, Craig J.
    Berger, Richard A.
    Tetreault, Matthew
    Paprosky, Wayne G.
    Sporer, Scott M.
    Jacobs, Joshua J.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2012, 94A (18) : 1655 - 1661
  • [6] Correlation between inclination of the acetabular component and metal ion levels in metal-on-metal hip resurfacing replacement
    De Haan, R.
    Pattyn, C.
    Gill, H. S.
    Murray, D. W.
    Campbell, P. A.
    De Smet, K.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (10): : 1291 - 1297
  • [7] Failure Modes of 433 Metal-on-Metal Hip Implants: How, Why, and Wear
    Ebramzadeh, Edward
    Campbell, Patricia A.
    Takamura, Karren M.
    Lu, Zhen
    Sangiorgio, Sophia N.
    Kalma, Jeremy J.
    De Smet, Koen A.
    Amstutz, Harlan C.
    [J]. ORTHOPEDIC CLINICS OF NORTH AMERICA, 2011, 42 (02) : 241 - +
  • [8] The Effect of Acetabular Inclination on Metal Ion Levels Following Metal-on-Metal Hip Arthroplasty
    Emmanuel, Andrew R.
    Bergin, Karen M.
    Kelly, Gabrielle E.
    McCoy, Gerald F.
    Wozniak, Andrew P.
    Quinlan, John F.
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (01) : 186 - 191
  • [9] Risk factors for inflammatory pseudotumour formation following hip resurfacing
    Glyn-Jones, S.
    Pandit, H.
    Kwon, Y. -M.
    Doll, H.
    Gill, H. S.
    Murray, D. W.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2009, 91B (12): : 1566 - 1574
  • [10] Incidence of adverse wear reactions in hip resurfacing arthroplasty: a single surgeon series of 2,600 cases
    Gross, Thomas P.
    Liu, Fei
    [J]. HIP INTERNATIONAL, 2013, 23 (03) : 250 - 258