Flexible active electrode arrays with ASICs that fit inside the rat's spinal canal

被引:16
作者
Giagka, Vasiliki [1 ,2 ]
Demosthenous, Andreas [3 ]
Donaldson, Nick [1 ]
机构
[1] UCL, Dept Med Phys & Biomed Engn, London WC1E 6BT, England
[2] Delft Univ Technol, Dept Microelect, NL-2628 CD Delft, Netherlands
[3] UCL, Dept Elect & Elect Engn, London WC1E 7JE, England
关键词
Active flexible electrode arrays; Application specific integrated circuits; Epidural stimulation; Laser patterned microelectrodes; Neurorehabilitation; Rivet bonding; ELECTRICAL-STIMULATION; IN-VITRO; CORD; FABRICATION; DESIGN; VIVO;
D O I
10.1007/s10544-015-0011-5
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Epidural spinal cord electrical stimulation (ESCS) has been used as a means to facilitate locomotor recovery in spinal cord injured humans. Electrode arrays, instead of conventional pairs of electrodes, are necessary to investigate the effect of ESCS at different sites. These usually require a large number of implanted wires, which could lead to infections. This paper presents the design, fabrication and evaluation of a novel flexible active array for ESCS in rats. Three small (1.7 mm(2)) and thin (100 mu m) application specific integrated circuits (ASICs) are embedded in the polydimethylsiloxane-based implant. This arrangement limits the number of communication tracks to three, while ensuring maximum testing versatility by providing independent access to all 12 electrodes in any configuration. Laser-patterned platinum-iridium foil forms the implant's conductive tracks and electrodes. Double rivet bonds were employed for the dice microassembly. The active electrode array can deliver current pulses (up to 1 mA, 100 pulses per second) and supports interleaved stimulation with independent control of the stimulus parameters for each pulse. The stimulation timing and pulse duration are very versatile. The array was electrically characterized through impedance spectroscopy and voltage transient recordings. A prototype was tested for long term mechanical reliability when subjected to continuous bending. The results revealed no track or bond failure. To the best of the authors' knowledge, this is the first time that flexible active electrode arrays with embedded electronics suitable for implantation inside the rat's spinal canal have been proposed, developed and tested in vitro.
引用
收藏
页数:13
相关论文
共 40 条
[1]   Floating light-activated microelectrical stimulators tested in the rat spinal cord [J].
Abdo, Ammar ;
Sahin, Mesut ;
Freedman, David S. ;
Cevik, Elif ;
Spuhler, Philipp S. ;
Unlu, M. Selim .
JOURNAL OF NEURAL ENGINEERING, 2011, 8 (05)
[2]   Flip chip assembly of thinned silicon die on flex substrates [J].
Banda, Charles ;
Johnson, R. Wayne ;
Zhang, Tan ;
Hou, Zhenwei ;
Charles, Harry K. .
IEEE TRANSACTIONS ON ELECTRONICS PACKAGING MANUFACTURING, 2008, 31 (01) :1-8
[3]   Ultra-thin chip technology and applications, a new paradigm in silicon technology [J].
Burghartz, Joachim N. ;
Appel, Wolfgang ;
Harendt, Christine ;
Rempp, Horst ;
Richter, Harald ;
Zimmermann, Martin .
SOLID-STATE ELECTRONICS, 2010, 54 (09) :818-829
[4]   A Computational Model for Epidural Electrical Stimulation of Spinal Sensorimotor Circuits [J].
Capogrosso, Marco ;
Wenger, Nikolaus ;
Raspopovic, Stanisa ;
Musienko, Pavel ;
Beauparlant, Janine ;
Luciani, Lorenzo Bassi ;
Courtine, Gregoire ;
Micera, Silvestro .
JOURNAL OF NEUROSCIENCE, 2013, 33 (49) :19326-19340
[5]   Ultra-thin electronic device package [J].
Chen, KY ;
Zenner, RLD ;
Arneson, M ;
Mountain, D .
IEEE TRANSACTIONS ON ADVANCED PACKAGING, 2000, 23 (01) :22-26
[6]   In Vivo and In Vitro Analysis of Rat Lumbar Spine Mechanics [J].
Cunningham, Matthew E. ;
Beach, Jocelyn M. ;
Bilgic, Serkan ;
Boachie-Adjei, Oheneba ;
van der Meulen, Marjolein C. H. ;
Hidaka, Chisa .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2010, 468 (10) :2695-2703
[7]   Adaptation of tissue to a chronic heat load [J].
Davies, C.R. ;
Fukumura, F. ;
Fukamachi, K. ;
Muramoto, K. ;
Himley, S.C. ;
Massiello, A. ;
Chen, J.-F. ;
Harasaki, H. .
ASAIO Journal, 1994, 40 (03)
[8]  
Detemple P., 2012, NEUWALK PROJECT REP, P25
[9]  
DONALDSON ND, 1986, MED BIOL ENG COMPUT, V24, P41, DOI 10.1007/BF02441604
[10]   Sub-threshold spinal cord stimulation facilitates spontaneous motor activity in spinal rats [J].
Gad, Parag ;
Choe, Jaehoon ;
Shah, Prithvi ;
Garcia-Alias, Guillermo ;
Rath, Mrinal ;
Gerasimenko, Yury ;
Zhong, Hui ;
Roy, Roland R. ;
Edgerton, Victor Reggie .
JOURNAL OF NEUROENGINEERING AND REHABILITATION, 2013, 10