Pulse contour cardiac output: an evaluation of the FloTrac method

被引:10
作者
Ostergaard, Morten [1 ]
Nielsen, Jonas [1 ]
Nygaard, Eigil [1 ]
机构
[1] Univ Copenhagen, Gentofte Hosp, Dept Anaesthesiol, DK-1168 Copenhagen, Denmark
关键词
cardiac output; haemodynamics; indicator dilution technique; pulmonary artery thermodilution; pulse contour cardiac output; Swan-Ganz; ARTERIAL-PRESSURE WAVE; FORM ANALYSIS; MONITORING-SYSTEM; THERMODILUTION; SURGERY;
D O I
10.1097/EJA.0b013e32831f343f
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background and objective The aim of this study was to determine the agreement between pulmonary artery thermodilution (PA-TD) and a new pulse contour method (PCM) FloTrac/Vigileo version 1.0, and to asses the ability of FloTrac to track sudden changes in cardiac output. Methods Cardiac output was determined twice after induction of anaesthesia, but before cardiac surgery, with both PA-TD and a PCM in order to determine the precision of both methods. The bias and agreement between the two methods were calculated using Bland-Altman analysis. Postoperatively, in patients with heart rates under 60 beats min(-1), atrial pacing was initiated and cardiac output was determined before and after with both methods. Results Twenty-five patients were investigated. The precisions of PA-TD and the PCM were 0.35 (95% confidence interval +/-0.12) and 0.6 l min(-1) (95% confidence interval +/-0.21%). The bias between PA-TD and the PCM was -0.51 l min(-1) and the limits of agreement were +/-1.87 l min(-1) (95% confidence interval +/-0.39 and +/-0.66). The percentage error was 48%. The changes in cardiac output with atrial pacing were in the same direction in all nine patients. Conclusion In this study, agreement between PA-TD and the PCM was poor, but the PCM was able to track the direction of pace-induced changes in cardiac output. Eur J Anaesthesiol 26:484-489 (C) 2009 European Society of Anaesthesiology.
引用
收藏
页码:484 / 489
页数:6
相关论文
共 25 条
  • [1] Thermodilution cardiac output - Cold vs. room temperature injectate and the importance of measuring the injectate temperature in the right atrium
    Berthelsen, PG
    Eldrup, N
    Nilsson, LB
    Rasmussen, JP
    [J]. ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2002, 46 (09) : 1103 - 1110
  • [2] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [3] Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery
    Breukers, Rose-Marieke B. G. E.
    Sepehrkhouy, Shahrzad
    Spiegelenberg, Stefan R.
    Groeneveld, A. B. Johan
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2007, 21 (05) : 632 - 635
  • [4] Buhre G, 1999, J CARDIOTHOR VASC AN, V13, P437
  • [5] Clinical evaluation of the FloTrac/Nigileo™ system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery
    Button, D.
    Weibel, L.
    Reuthebuch, O.
    Genoni, M.
    Zollinger, A.
    Hofer, C. K.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2007, 99 (03) : 329 - 336
  • [6] Comparison of FloTrac™ cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements
    Cannesson, M.
    Attof, Y.
    Rosamel, P.
    Joseph, P.
    Bastien, O.
    Lehot, J.-J.
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2007, 24 (10) : 832 - 839
  • [7] Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac™/Vigileo™) in haemodynamically unstable patients
    Compton, F. D.
    Zukunft, B.
    Hoffmann, C.
    Zidek, W.
    Schaefer, J. -H.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) : 451 - 456
  • [8] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [9] Cardiac index measurements during rapid preload changes: a comparison of pulmonary artery thermodilution with arterial pulse contour analysis
    Felbinger, TW
    Reuter, DA
    Eltzschig, HK
    Bayerlein, J
    Goetz, AE
    [J]. JOURNAL OF CLINICAL ANESTHESIA, 2005, 17 (04) : 241 - 248
  • [10] Continuous, less invasive, hemodynamic monitoring in intensive care after cardiac surgery
    Godje, O
    Hoke, K
    Lamm, P
    Schmitz, C
    Thiel, C
    Weinert, M
    Reichart, B
    [J]. THORACIC AND CARDIOVASCULAR SURGEON, 1998, 46 (04) : 242 - 249