Enhancement of electron-beam surface dose with an electron multi-leaf collimator (eMLC): a feasibility study

被引:8
作者
Vatanen, T. [1 ]
Traneus, E. [2 ]
Lahtinen, T. [1 ,3 ]
机构
[1] Kuopio Univ Hosp, Dept Oncol, FIN-70211 Kuopio, Finland
[2] Nucletron Scandinavia AB, SE-75147 Uppsala, Sweden
[3] Univ Kuopio, Dept Phys, FIN-70211 Kuopio, Finland
关键词
TREATMENT-PLANNING SYSTEM; RADIATION-THERAPY; BOLUS; DESIGN; VERIFICATION; IRRADIATION; MODEL;
D O I
10.1088/0031-9155/54/8/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Use of a water-equivalent bolus in electron-beam radiotherapy is sometimes impractical and non-hygienic. Therefore, the feasibility of applying adjacent narrow beams for producing high surface dose electron beams without a bolus was investigated. Depth dose curves and profiles in water were calculated and measured for 6 and 9 MeV electron-beam segments (width 0.3-1.5 cm, length 10 cm) for source-to-surface distances (SSD) 102 and 105 cm. Segment shaping was performed with an add-on electron multi-leaf collimator prototype attached to the Varian 2100 C/D linac. Dose calculations were performed with the Voxel Monte Carlo++ algorithm. Resulting dose distributions in typical clinical cases were compared with the bolus technique. With a composite segmental field with 1.0 cm wide segments the surface dose was over 90% of the depth dose maximum for both energies. The build-up area practically disappeared with a 0.5 cm wide single beam. This led to decrease in the therapeutic range for composite fields with segment widths smaller than 1.0 cm. The new technique yielded similar surface doses as the bolus technique. The photon contamination was 4% with a 9 x 10 cm(2) field (1.0 cm wide segments) compared to 1% for the respective open field with 9 MeV with a bolus. The calculated dose agreed within 2 mm and 3% of the measured dose in 93.7% and 85.2% of the voxels. Adjacent narrow eMLC beams with a 1.0 cm width are suitable to produce electron fields with high surface dose. Despite a slight nonuniformity in the surface profiles in the lateral part of the field at SSD 102 cm, surface dose and target coverage are comparable with the bolus technique.
引用
收藏
页码:2407 / 2419
页数:13
相关论文
共 25 条
  • [1] Energy modulated electron therapy using a few leaf electron collimator in combination with IMRT and 3D-CRT: Monte Carlo-based planning and dosimetric evaluation
    Al-Yahya, K
    Schwartz, M
    Shenouda, G
    Verhaegen, F
    Freeman, C
    Seuntjens, J
    [J]. MEDICAL PHYSICS, 2005, 32 (09) : 2976 - 2986
  • [2] Design and dosimetry of a few leaf electron collimator for energy modulated electron therapy
    Al-Yahya, Khalid
    Verhaegen, Frank
    Seuntjens, Jan
    [J]. MEDICAL PHYSICS, 2007, 34 (12) : 4782 - 4791
  • [3] DEPTH-DOSE FLATTENING OF ELECTRON-BEAMS USING A WIRE MESH BOLUS
    ALASTI, H
    GALBRAITH, DM
    [J]. MEDICAL PHYSICS, 1995, 22 (10) : 1675 - 1683
  • [4] [Anonymous], 2004, J ICRU, V4, P2, DOI 10.1093/jicru/ndh018
  • [5] [Anonymous], 1976, ICRU Report 24
  • [6] USE OF VARIABLE THICKNESS BOLUS TO CONTROL ELECTRON-BEAM PENETRATION IN CHEST WALL IRRADIATION
    ARCHAMBEAU, JO
    FORELL, B
    DORIA, R
    FINDLEY, DO
    JURISCH, R
    JACKSON, R
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1981, 7 (06): : 835 - 842
  • [7] Cederbaum M, 2001, Med Dosim, V26, P275, DOI 10.1016/S0958-3947(01)00075-9
  • [8] Clinical use of a commercial Monte Carlo treatment planning system for electron beams
    Cygler, JE
    Lochrin, C
    Daskalov, GM
    Howard, M
    Zohr, R
    Esche, B
    Eapen, L
    Grimard, L
    Caudrelier, JM
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (05) : 1029 - 1034
  • [9] Evaluation of the first commercial Monte Carlo dose calculation engine for electron beam treatment planning
    Cygler, JE
    Daskalov, GM
    Chan, GH
    Ding, GX
    [J]. MEDICAL PHYSICS, 2004, 31 (01) : 142 - 153
  • [10] Gunhan Basri, 2003, Med Dosim, V28, P193, DOI 10.1016/S0958-3947(03)00072-4