Coincidence summing corrections for point and volume 152Eu sources

被引:12
作者
Novkovic, Dusan [1 ]
Durasevic, Mirjana [1 ]
Kandic, Aleksandar [1 ]
Vukanac, Ivana [1 ]
Seslak, Bojan [1 ]
Milosevic, Zoran [1 ]
机构
[1] Univ Belgrade, Vinca Inst Nucl Sci, Lab Nucl & Plasma Phys, Belgrade, Serbia
关键词
Gamma-ray spectrometry; Eu-152; Coincidence summing corrections; GAMMA-RAY-SPECTROMETRY;
D O I
10.1016/j.apradiso.2015.10.015
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this article, the X-ray and gamma-ray coincidence summing effect in Eu-152 is studied. Coincidence summing corrections and peak and total efficiencies of point and volume sources were determined using the direct matrices multiplication (DMM) method. The theoretically evaluated peak count rates were found to be in good agreement with the experimentally obtained values. Validation was performed by comparing the calculated efficiency curves and the corresponding correction factors with the results obtained using GESPECOR 4.2 software. (C) 2015 Elsevier Ltd. All rights reserved.
引用
收藏
页码:138 / 144
页数:7
相关论文
共 28 条
[1]   True coincidence summing corrections in point and extended sources [J].
Agarwal, Chhavi ;
Chaudhury, Sanhita ;
Goswami, A. ;
Gathibandhe, M. .
JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2011, 289 (03) :773-780
[2]  
[Anonymous], 2002, IAEA-TECDOC- 1275
[3]   Application of GESPECOR software for the calculation of coincidence summing effects in special cases [J].
Arnold, D ;
Sima, O .
APPLIED RADIATION AND ISOTOPES, 2004, 60 (2-4) :167-172
[4]   Cascade summing in gamma-ray spectrometry in marinelli-beaker geometries: the third efficiency curve [J].
Blaauw, M ;
Gelsema, SJ .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 505 (1-2) :311-315
[5]   Fast procedures for coincidence-summing correction in γ-ray spectrometry [J].
De Felice, P ;
Angelini, P ;
Fazio, A ;
Biagini, R .
APPLIED RADIATION AND ISOTOPES, 2000, 52 (03) :745-752
[6]   COINCIDENCE SUMMING CORRECTIONS IN GE(LI)-SPECTROMETRY AT LOW SOURCE-TO-DETECTOR DISTANCES [J].
DEBERTIN, K ;
SCHOTZIG, U .
NUCLEAR INSTRUMENTS & METHODS, 1979, 158 (2-3) :471-477
[7]  
Delgado J.U., 2006, P 18 IMEKO WORLD C 2, V1, P231
[8]   Testing the equivalence of several algorithms for calculation of coincidence summing corrections [J].
Kanisch, G. ;
Vidmar, J. ;
Sima, O. .
APPLIED RADIATION AND ISOTOPES, 2009, 67 (10) :1952-1956
[9]   Coincidence summing effects in gamma-ray spectrometry using a Marinelli beaker [J].
Lee, Mosung ;
Park, Tae Soon ;
Woo, Jong-Kwan .
APPLIED RADIATION AND ISOTOPES, 2008, 66 (6-7) :799-803
[10]   Intercomparison of methods for coincidence summing corrections in gamma-ray spectrometry-part II (volume sources) [J].
Lepy, M. -C. ;
Altzitzoglou, T. ;
Anagnostakis, M. J. ;
Capogni, M. ;
Ceccatelli, A. ;
De Felice, P. ;
Djurasevic, M. ;
Dryak, P. ;
Fazio, A. ;
Ferreux, L. ;
Giampaoli, A. ;
Han, J. B. ;
Hurtado, S. ;
Kabdic, A. ;
Kanisch, G. ;
Karfopoulos, K. L. ;
Klemola, S. ;
Kovar, P. ;
Laubenstein, M. ;
Lee, J. H. ;
Lee, J. M. ;
Lee, K. B. ;
Pierre, S. ;
Carvalhal, G. ;
Sima, O. ;
Chau Van Tao ;
Tran Thien Thanh ;
Vidmar, T. ;
Vukanac, I. ;
Yang, M. J. .
APPLIED RADIATION AND ISOTOPES, 2012, 70 (09) :2112-2118