Automated quality assurance applied to mammographic imaging

被引:2
作者
Blot, L [1 ]
Davis, A
Holubinka, M
Marti, R
Zwiggelaar, R
机构
[1] Univ E Anglia, Sch Informat Syst, Norwich NR4 7TJ, Norfolk, England
[2] Portsmouth Hosp NHS Trust, Portsmouth PO3 6AD, Hants, England
关键词
automatic quality control; mammographic images; grey-level co-occurrence matrices; image segmentation;
D O I
10.1155/S1110865702203029
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Quality control in mammography is based upon subjective interpretation of the image quality of a test phantom. In order to suppress subjectivity due to the human observer, automated computer analysis of the Leeds TOR(MAM) test phantom is investigated. Texture analysis via grey-level co-occurrence matrices is used to detect structures in the test object. Scoring of the substructures in the phantom is based on grey-level differences between regions and information from grey-level co-occurrence matrices. The results from scoring groups of particles within the phantom are presented.
引用
收藏
页码:736 / 745
页数:10
相关论文
共 11 条
[1]  
Anderson T.W., 1986, STAT ANAL DATA, V2nd
[2]  
BROOKS JA, 1993, MICROSTR SC, V20, P3
[3]   A SURVEY OF IMAGE REGISTRATION TECHNIQUES [J].
BROWN, LG .
COMPUTING SURVEYS, 1992, 24 (04) :325-376
[4]   A PRELIMINARY INVESTIGATION OF THE IMAGING PERFORMANCE OF PHOTOSTIMULABLE PHOSPHOR COMPUTED RADIOGRAPHY USING A NEW DESIGN OF MAMMOGRAPHIC QUALITY-CONTROL TEST OBJECT [J].
COWEN, AR ;
BRETTLE, DS ;
COLEMAN, NJ ;
PARKIN, GJS .
BRITISH JOURNAL OF RADIOLOGY, 1992, 65 (774) :528-535
[5]   A COMPARISON OF MAMMOGRAPHIC PHANTOMS [J].
FAULKNER, K ;
LAW, J .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (794) :174-180
[6]  
Forrest P., 1986, Breast cancer screening: Report to the health minister of england, wales, Scotland and northern Ireland
[7]   TEXTURAL FEATURES FOR IMAGE CLASSIFICATION [J].
HARALICK, RM ;
SHANMUGAM, K ;
DINSTEIN, I .
IEEE TRANSACTIONS ON SYSTEMS MAN AND CYBERNETICS, 1973, SMC3 (06) :610-621
[8]  
KARSSEMEIJER N, 1995, EUR RADIOLOGY S, V5
[9]  
MCLACHAN JG, 1992, DISCRIMINANT ANAL ST
[10]   Objective assessment of phantom image quality in mammography: a feasibility study [J].
Smith, ADC ;
Smith, IAC ;
Dance, DR .
BRITISH JOURNAL OF RADIOLOGY, 1998, 71 (841) :48-58