摘要
Background and purpose- The accelerated wear of poorly functioning metal-on-metal (MoM) hip implants may cause elevated whole-blood cobalt (Co) and chromium (Cr) levels. Hematological and endocrinological changes have been described as the most sensitive adverse effects due to Co exposure. We studied whether there is an association between whole-blood Co/Cr levels and leukocyte, hemoglobin, or platelet levels. Patients and methods- We analyzed whole-blood Co and Cr values and complete blood counts (including leukocytes, hemoglobin, platelets) from 1,900 patients with MoM hips. The mean age at the time of whole-blood metal ion measurements was 67 years (SD 10). The mean time from primary surgery to whole-blood metal ion measurement was 8.2 years (SD 3.0). The mean interval between postoperative blood counts and metal ion measurements was 0.2 months (SD 2.7). Results- The median Co value was 1.9 mu g/L (0.2-225), Cr 1.6 mu g/L (0.2-125), mean leukocyte count 6.7 x 10(9)/L (SD 1.9), hemoglobin value 143 g/L (SD 13), and platelet count 277 x 10(9)/L (SD 70). We did not observe clinically significant correlations between whole-blood Co/Cr and leukocyte, hemoglobin, or platelet counts. Interpretation- Elevated whole-blood Co and Cr values are unlikely to explain abnormal blood counts in patients with MoM hips and the reason for possible abnormal blood counts should be sought elsewhere.
引用
收藏
页码:711 / 716
页数:6
相关论文
共 24 条
  • [1] Differential distribution of cobalt, chromium, and nickel between whole blood, plasma and urine in patients after metal-on-metal (MoM) hip arthroplasty
    Newton, Ashley W.
    Ranganath, Lakshminarayan
    Armstrong, Catherine
    Peter, Viju
    Roberts, Norman B.
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2012, 30 (10) : 1640 - 1646
  • [2] Blood levels of cobalt and chromium are inversely correlated to head size after metal-on-metal resurfacing arthroplasty
    Parry, Michael C.
    Eastaugh-Waring, Steve
    Bannister, Gordon C.
    Learmonth, Ian D.
    Case, Charles Patrick
    Blom, Ashley W.
    HIP INTERNATIONAL, 2013, 23 (06) : 529 - 534
  • [3] Transplacental transfer of cobalt and chromium in patients with metal-on-metal hip arthroplasty - A controlled study
    Ziaee, H.
    Daniel, J.
    Datta, A. K.
    Blunt, S.
    McMinn, D. J. W.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (03): : 301 - 305
  • [4] Mild or moderate renal insufficiency does not increase circulating levels of cobalt and chromium in patients with metal-on-metal hip arthroplasty
    Lainiala, O.
    Reito, A.
    Jamsa, P.
    Eskelinen, A.
    BONE & JOINT JOURNAL, 2017, 99B (09) : 1147 - 1152
  • [5] Chromium speciation in the blood of metal-on-metal hip implant patients
    Finley, Brent
    Scott, Paul K.
    Glynn, Meghan E.
    Paustenbach, Dennis
    Donovan, Ellen
    Thuett, Kerry A.
    TOXICOLOGICAL AND ENVIRONMENTAL CHEMISTRY, 2017, 99 (01) : 48 - 64
  • [6] Changes in Cobalt and Chromium Levels After Metal-on-Metal Hip Resurfacing in Young, Active Chinese Patients
    Yang, Jing
    Shen, Bin
    Zhou, ZongKe
    Pei, Fuxing
    Kang, PengDe
    JOURNAL OF ARTHROPLASTY, 2011, 26 (01) : 65 - 70
  • [7] Metal release from hip prostheses: biological monitoring of cobalt and chromium in patients with metal-on-metal hip arthroplasty
    Torra, Merce
    Sandalinas, Silvia
    Martin, Mayte
    Nogue, Santiago
    TRACE ELEMENTS AND ELECTROLYTES, 2017, 34 (01) : 1 - 4
  • [8] The value of serial metal ion levels in following up patients with metal-on-metal hip arthroplasty
    McHugh, Gavin
    Merchant, Rajiv
    Kelly, Gabrielle E.
    Bergin, Karen M.
    McCoy, Gerald F.
    Wozniak, Andrew P.
    Quinlan, John F.
    HIP INTERNATIONAL, 2017, 27 (01) : 14 - 20
  • [9] Case Report: High Chromium and Cobalt Levels in a Pregnant Patient with Bilateral Metal-on-Metal Hip Arthroplasties
    Fritzsche, Juliane
    Borisch, Cornelia
    Schaefer, Christof
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2012, 470 (08) : 2325 - 2331
  • [10] Serum cobalt concentrations remain at low levels at a minimum of 20 years following metal-on-metal total hip arthroplasty
    Waldstein, W.
    Koller, U.
    Springer, B.
    Kolbitsch, P.
    Brodner, W.
    Windhager, R.
    Lass, R.
    BONE & JOINT RESEARCH, 2020, 9 (03): : 146 - 151