Fracture of titanium nitride-coated femoral component after total knee arthroplasty

被引:11
|
作者
Park, Se-Wook [1 ]
Kim, Hyungsuk [1 ]
In, Yong [1 ]
机构
[1] Catholic Univ Korea, Seoul St Marys Hosp, Dept Orthopaed Surg, Seoul 137701, South Korea
来源
KNEE | 2014年 / 21卷 / 04期
关键词
Total knee arthroplasty; Fracture; Femoral component; Titanium nitride coating; REPLACEMENT; DESIGN;
D O I
10.1016/j.knee.2014.04.002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
We report a case of fracture of a titanium nitride-coated femoral component 3 years after primary total knee arthroplasty (TKA). The fracture was at the medial condylar area just posterior to the medial peg of the femoral component. The backside of the broken medial condylar portion of the femoral component was devoid of cement. Debonding of the component is a possible cause of the stress fracture. To our knowledge, this is the first case report of the fracture of the femoral component manufactured from titanium alloy. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:871 / 874
页数:4
相关论文
共 50 条
  • [1] Fracture of the oxidized zirconium femoral component after total knee arthroplasty
    Ichimura, Ryuji
    Minamikawa, Tomohiko
    Nakagawa, Hiroshi
    Mori, Akira
    Midorikawa, Koji
    Sakuragi, Koji
    Minamikawa, Hiromichi
    KNEE, 2022, 36 : 27 - 32
  • [2] Fracture of femoral component in a resurfacing total knee arthroplasty
    Michos, John
    Rallis, John
    Fassoulas, Apostolos
    JOURNAL OF ARTHROPLASTY, 2006, 21 (07): : 1068 - 1071
  • [3] FRACTURE OF THE FEMORAL COMPONENT IN CEMENTLESS TOTAL KNEE ARTHROPLASTY
    WHITESIDE, LA
    FOSCO, DR
    BROOKS, JG
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1993, (286) : 71 - 77
  • [4] Fracture of the femoral component associated with polyethylene wear and osteolysis after total knee arthroplasty
    Huang, CH
    Yang, CY
    Cheng, CK
    JOURNAL OF ARTHROPLASTY, 1999, 14 (03): : 375 - 379
  • [5] Atraumatic Fracture of Cemented Vanguard Total Knee Arthroplasty Femoral Component
    Vovos, Tyler J.
    Watters, Tyler S.
    Hughes, Erin M.
    McNabb, David C.
    ORTHOPEDICS, 2020, 43 (05) : E476 - E479
  • [6] Component fracture in total knee arthroplasty
    Clarke, HD
    Trousdale, RT
    KNEE, 1999, 6 (04): : 261 - 267
  • [7] Hip resurfacing with a highly cross-linked polyethylene acetabular liner and a titanium nitride-coated femoral component
    Pritchett, James W.
    HIP INTERNATIONAL, 2018, 28 (04) : 422 - 428
  • [8] Femoral Component Fracture in a Total Knee Arthroplasty Patient With a Persistent Flexion Contracture
    Tate, Jackson P.
    Schneider, Andrew M.
    Brown, Nicholas M.
    ARTHROPLASTY TODAY, 2023, 22
  • [9] FEMORAL COMPONENT LOOSENING IN TOTAL KNEE ARTHROPLASTY
    KING, TV
    SCOTT, RD
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1985, (194) : 285 - 290
  • [10] Femoral component rotation and Laurin angle after total knee arthroplasty
    Classen, Tim
    Wegner, Alexander
    Muller, Ralf-Dietrich
    Von Knoch, Marius
    ACTA ORTHOPAEDICA BELGICA, 2010, 76 (01): : 69 - 73