Dissolution for the analysis of alloys containing boron by inductively coupled plasma optical emission spectrometry
被引:1
|作者:
Danzaki, Y
[1
]
Wagatsuma, K
[1
]
机构:
[1] Tohoku Univ, Inst Mat Res, Aoba Ku, Sendai, Miyagi 9808577, Japan
关键词:
B-containing alloys;
dissolution method;
ICP-OES;
D O I:
10.2116/bunsekikagaku.53.757
中图分类号:
O65 [分析化学];
学科分类号:
070302 ;
081704 ;
摘要:
This paper presents dissolution methods of several B-containing alloys to determine the amounts of B accurately by ICP-OES. It should be principally noted that boron is easily volatilized when the samples are dissolved in acid solutions, which may cause a serious error of the analytical result. Fe-B, Mn-Al-B, and Al-Pd-Mn-B alloy samples (30 mg) were dissolved in TEFE beakers by heating in 25 ml of an acid mixture: H3PO4 : HNO3 : HCl : H2O = 1 : 2 : 2 : 5. A Nd-Fe-B- Si alloy sample (100 mg) was dissolved in a TEFE beaker by heating in 30 ml of an acid mixture: H3PO4 : HNO3 : HCl : H2O = 1 : 2 : 2 : 5 and 4 ml of NaOH solution (8 mg/ml). The NaOH addition can prevent Si in the solution from precipitating because silicic acid can be dissolved in alkaline solutions. An Fe-Co-Nd-B alloy sample (30 mg) was dissolved in a TEFE beaker by heating in 30 ml of an acid mixture: HNO3 : guliceline : H2O = 15 : 0.3 : 14.7. A Nb-B alloy sample (100 mg) was dissolved in a polypropylene flask in 30 ml of HNO3 (1 + 1) and 2 ml of HF at room temperature. An Fe-Zr-Co-Ni-B sample (50 mg) was dis-HNO3 solved in a polypropylene flask in 35 ml of an acid mixture: HNO3 : HF : guliceline : H2O = 15 : 2 : 0.1 : 17.9 at room temperature. Ni-Al-Hf-B and Ni-Al-Ta-B alloy samples (40 mg) were dissolved in a TEFE beaker by heating in 35 ml of an acid mixture: HNO3 : HF : H2O 15 : 2 : 18 and 5 ml of mannitol (1 w/v%).
引用
收藏
页码:757 / 761
页数:5
相关论文